Vacuette 450161Recalls & reported problems
Greiner Bio-One GmbH · SAFETY Blood Collection Set + Holder 23G x 3/4" tubing length 12" (30 cm), single-packed, sterile, not made with natural rubber latex · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 19120049202413), received between Feb 21, 2020 and Feb 21, 2020 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K011786, K080235, K121908); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Material Separation1
- Mechanical Problem1
- Use of Device Problem1
Patient problems
- Needle Stick/Puncture2
- Exposure to Body Fluids1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 21, 2020Use of Device Problem
CUSTOMER IS REPORTING A NEEDLE STICK. PHLEBOTOMIST THOUGHT THE SAFETY DEVICE WAS ACTIVATED AFTER THE STICK. WHEN GATHERING SUPPLIES HE FELT THE SHARP. THE NEEDLE SLID HAD OUT OF THE PROTECTION SHEATH AND STUCK HIS HAND. CUSTOMER STATED THEY DO NOT BELIEVE THIS WAS A USER ERROR. CUSTOMER ADVISED THEY PLACE REFERENCE ITEMS IN STOCK BINS PRIOR TO USE AND THE BINS CAN CONTAIN MULTIPLE LOT NUMBERS. CUSTOMER STATES THEY HAVE SEVERAL LOTS IN USE AND CANNOT IDENTIFY WHICH LOT WAS ASSOCIATED WITH THE INJURY.
Patient problems: Needle Stick/Puncture
MAUDE report 9740862 ↗InjuryFeb 21, 2020Break; Mechanical Problem; Material Separation
CUSTOMER STATES THEY EXPERIENCED A NEEDLE STICK INJURY WHEN THE NEEDLE WAS BEING WITHDRAWN FROM PATIENT'S HAND. WHILE THE PHLEBOTOMIST WAS TERMINATING THE VENIPUNCTURE AND ACTIVATING THE SAFETY, THE PATIENT MOVED AND THE NEEDLE CAUGHT THE PHLEBOTOMIST NEAR THE THUMB. CUSTOMER ADVISED THAT WHILE ACTIVATING THE SAFETY, THE PHLEBOTOMIST WAS NOT ABLE TO ANCHOR THE DEVICE AT THE WING DUE TO THE DRAW SITE BEING ON THE HAND. NO ISSUES WITH THE DEVICE REPORTED. CUSTOMER ADVISED OF AN INCIDENT WHERE THE NEEDLE/WING SECTION SEPARATED FROM THE REMAINDER OF THE DEVICE WHILE THE PHLEBOTOMIST ACTIVATED THE SAFETY MECHANISM. THE PHLEBOTOMIST PUSHED THE BUTTONS ON THE HUB TO ACTIVATE THE SAFETY AND WHILE PULLING BACK THE SAFETY MECHANISM, IT KEPT GOING AND BROKE AWAY FROM NEEDLE. CUSTOMER ADVISED THAT THE DEVICE WAS SAVED. CUSTOMER ADVISED THAT THE CONNECTOR TO THE LUER NEEDLE SEPARATED FROM THE LUER/HOLDER RESULTING IN A BLOOD EXPOSURE. THE PHLEBOTOMIST PUSHED THE BLOOD COLLECTION TUBE ON, THE TUBE PUSHED BACK AND WHEN THE PHLEBOTOMIST PUSHED THE TUBE BACK ON, THE CONNECTOR DISCONNECTED FROM THE LUER/HOLDER. CUSTOMER ADVISED THAT THE DEVICE WAS NOT SAVED. CUSTOMER ADVISED OF ANOTHER BLOOD EXPOSUR
Patient problems: Exposure to Body Fluids; Needle Stick/Puncture
MAUDE report 9739213 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K011786, K080235, K121908. One submission often covers many models, so check the product and lot details in each record.
- Z-0835-2013 · Feb 1, 2013 · TerminatedVACUETTE Safety Blood Collection Set + Luer Adapter, 23x12" Safety Blood Collect LA Product, Item number 450096. Usage: The Safety Blood Collection Set is used in routine venipunc
- Z-0797-03 · Mar 10, 2003 · TerminatedVacuette, Safety Blood Collection Set with Luer Adapter
Other Needle, hypodermic, single lumen devices
- BD Nano 2nd Gen 320550Becton Dickinson Pen Limited · Needle, hypodermic, single lumen2,044 reports
- BD SafetyGlide 305916Becton, Dickinson And Company · Needle, hypodermic, single lumen1,767 reports
- BD Nano Ultra-Fine 320122Becton Dickinson Pen Limited · Needle, hypodermic, single lumen601 reports
- N/A 305106Becton, Dickinson And Company · Needle, hypodermic, single lumen1 recall467 reports
- N/A 305110Becton, Dickinson And Company · Needle, hypodermic, single lumen418 reports
- BD Safe-Clip 328235Becton, Dickinson and Company Inc · Needle, hypodermic, single lumen268 reports
- BD Ultra-Fine 320109Becton Dickinson Pen Limited · Needle, hypodermic, single lumen229 reports
- BD AutoShield Duo 329515Becton, Dickinson And Company · Needle, hypodermic, single lumen208 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.