Skip to content

BREEZE® 2 1465ARecalls & reported problems

Ascensia Diabetes Care US Inc. · Blood Glucose Test Strips. Sold in a carton of 50 Test Strips (5 Discs). · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20301931465505), received between Jun 9, 2017 and Jul 28, 2017 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Incorrect Measurement1
  • Low Readings1

Patient problems

  • Hyperglycemia1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2017Incorrect Measurement; Incorrect Measurement

    ADVOCATE STATED HIS WIFE'S BLOOD GLUCOSE READING ON THE BREEZE2 WAS 60-100 MG/DL DIFFERENT THAN THE READING ON THE PHARMACY METER. CUSTOMER HAD SYMPTOMS FOR HIGH BLOOD GLUCOSE. SPECIFIC INFORMATION WAS NOT PROVIDED. DEPENDING ON THE ACTUAL READINGS, THE DIFFERENCE BETWEEN THEM COULD FALL IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE ADVOCATE WAS ADVISED TO RETURN THE STRIPS FOR EVALUATION. NEW METER AND STRIPS WERE SENT TO THE CUSTOMER.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6749898 ↗
  • MalfunctionJun 9, 2017Low Readings

    THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 40MG/DL ON THE BREEZE2, RETESTED ON A DIFFERENT METER AND THE READING WAS OVER 200MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE CUSTOMER WAS ADVISED TO RETURN THE TEST STRIPS FOR EVALUATION. NEW METER AND TEST STRIPS WERE SENT TO HIM.

    Patient problems: Hyperglycemia

    MAUDE report 6629389 ↗

FDA clearance and classification

The FDA regulates product code CGA (glucose oxidase, glucose) under 21 CFR 862.1345 as a Class II device. 21 CFR 862.1345 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the BREEZE® 2 1465A

Has the BREEZE® 2 1465A been recalled?

No FDA recall record in our data names this exact device identifier (DI 20301931465505), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BREEZE® 2 1465A?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jun 9, 2017 and Jul 28, 2017. In those reports, the most common specific device problems are “Incorrect Measurement” (1) and “Low Readings” (1); the most common patient problem is “Hyperglycemia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BREEZE® 2 1465A?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the BREEZE® 2 1465A, and is it still sold?

The FDA's device database lists Ascensia Diabetes Care US Inc. as the company responsible for it (brand BREEZE® 2, model 1465A). Its record shows it as no longer in commercial distribution.