Alexis Contained Extraction System with Kii Balloon Blunt Tip System GTK14Recalls & reported problems

Applied Medical Resources Corporation · Kit Includes: GTB14, ALEXIS CONTAINED EXTRACTION SYSTEM C0R47, 12x100 Kii BALLOON BLUNT TIP TROCAR · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20607915137070), received between Mar 14, 2023 and Jan 29, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Device Damaged by Another Device1
  • Incomplete or Missing Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Laceration(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: VAGINAL ASSISTED TOTAL LAPAROSCOPIC HYSTERECTOMY. EVENT DESCRIPTION: HOSPITAL NAME: DR. [NAME]. "THE SURGEON SOUGHT TO UTILIZE THE ALEXIS CES FOR SPECIMEN EXTRACTION VIA THE VAGINAL CANAL. FOLLOWING THE RESECTION OF THE UTERUS, THE ALEXIS CES WAS INSERTED THROUGH THE VAGINA INTO THE ABDOMINAL CAVITY. THE VAGINA WAS THEN CLOSED WITH A PAD TO FACILITATE INSUFFLATION AND TO SECURE THE SPECIMEN WITHIN THE BAG. THE SURGEON PROCEEDED TO PULL ON THE ALEXIS CES THREAD IN AN ATTEMPT TO EXTRACT THE SPECIMEN. HOWEVER, DUE TO THE SIZE OF THE SPECIMEN, MANUAL MORCELLATION WAS INITIATED TO REDUCE ITS SIZE." "DESPITE HIS EFFORTS, THE SPECIMEN REMAINED SOMEWHAT LARGE, PROMPTING THE SURGEON TO APPLY INCREASED FORCE TO REMOVE IT. AFTER SEVERAL MINUTES OF STRONG PULLING, THE BAG BECAME TORN ALONG THE SEAM. THE SURGEON WAS ABLE TO RETRIEVE THE SPECIMEN BY SECURING IT WITH AN OPEN INSTRUMENT AND CONTINUED PULLING UNTIL IT WAS SUCCESSFULLY REMOVED. THE SURGEON ACKNOWLEDGED THAT THE FORCE APPLIED IN AN ATTEMPT TO EXTRACT THE SPECIMEN CONTRIBUTED TO THE RUPTURE OF THE BAG, BUT I CHOSE TO SHARE THIS INCIDENT FOR YOUR INFORMATION AND REVIEW." INTERVENTION: THE SURGEON WAS ABLE TO RETRIE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21270024 ↗
  • MalfunctionMay 2, 2023Device Damaged by Another Device; Material Split, Cut or Torn; Device Damaged by Another Device; Material Split, Cut or Torn

    PROCEDURE PERFORMED: HYSTERECTOMY EVENT DESCRIPTION: ACCORDING TO THE SURGEON, POTENTIALLY DURING THE MANUAL MORCELLATION STAGE OF THE PROCEDURE. APPLIED MEDICAL REP NOT PRESENT DURING CASE. DURING A ROBOTIC CASE, THE BAG WAS DEPLOYED INSIDE THE PATIENT, THE BAG OPENED UP, SURGEON COMMENTED THAT VISION OF ROBOTIC INSTRUMENTS MAY HAVE BEEN OBSCURED AT THE STAGE BY THE BAG THE UTERUS AND TUBES WERE PLACED INSIDE THE BAG, A PHOTO WAS CAPTURED AT THIS TIME AND ALL APPEARED NORMAL. MANUAL MORCELLATION TOOK PLACE AND THE SURGEON NOTICED FATTY TISSUE AND BRIGHT RED BLOOD. SURGEON THEN TOOK THE HASSAN PORT ROBOT ARMS OFF, TOOK OUT SPECIMEN, RE-ESTABLISHED PNEUMO, WENT BACK TO THE CONSOLE AND CALLED FOR THE ASSISTANCE OF A GENERAL SURGEON TO REPAIR THE MESENTERIC VESSEL WHICH HAD BEEN CUT. SURGEON NOTICED THE BAG HAD BEEN CUT. MESENTERIC VESSEL HAD BEEN CUT REQUIRING REPAIR AND SUBSEQUENT 2 UNITS OF BLOOD GIVEN POST SURGERY. PATIENT ALSO COMPLAINING OF UPSET STOMACH AND IS BEING MONITORED BY GENERAL SURGEON WHO WAS CALLED IN TO ASSIST. DEVICE REMOVED AND SURGICAL INTERVENTION TO REPAIR TISSUE DAMAGE. OTHER INSTRUMENTS USED WHEN THE COMPLAINT EVENT OCCURRED: [NAME] ROBOT, ROBOTIC PORTS AND S

    Patient problems: Laceration(s)

    MAUDE report 16857250 ↗
  • MalfunctionMar 14, 2023Incomplete or Missing Packaging; Incomplete or Missing Packaging

    NAME OF PROCEDURE BEING PERFORMED. NA (BEFORE USE). DETAILED DESCRIPTION OF EVENT. "ISSUE: SURGEON ASKED FOR A STERILE DISPOSABLE SUPPLY. ORIENTEE RN OPENED PRODUCT. PRODUCT WAS NOT FACTORY SEALED ON ONE EDGE. SURG TECH ACCEPTED PRODUCT WITHOUT KNOWLEDGE OF UNSEALED. SHE CHANGED GLOVES WHEN ISSUE DISCOVERED, PRODUCT NOT PUT ON STERILE FIELD." PRODUCT IS AVAILABLE FOR RETURN. TYPE OF INTERVENTION. NA (BEFORE USE). PATIENT STATUS. NA (BEFORE USE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16540124 ↗

Other Laparoscope, General & Plastic Surgery devices

All Laparoscope, General & Plastic Surgery devices →