Enlite ® MMT-7008ARecalls & reported problems
Medtronic Minimed, Inc. · SENSOR MMT-7008A ENLT PLUS 5PK 2L 1CL PR · Class 3
Quick summary
This Medtronic MiniMed Enlite glucose sensor is worn under the skin to measure blood sugar for an insulin pump. The 994 reports often describe sensor readings that do not match the blood sugar test, the pump suspending insulin after inaccurate readings, and cannulas going missing or electrodes breaking. There are 212 injury reports and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 994
- Deaths reported
- 0
- Injuries reported
- 212
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 994 adverse event reports that list this exact device (DI 20643169786725), received between Jan 8, 2018 and Nov 3, 2023 — classified as 212 injuries, 782 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (620 reports).
26 other recalls cover devices cleared under the same FDA submission (P120010, P150001, P150019, P150029, P160017); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings620
- Unexpected Therapeutic Results213
- Device Displays Incorrect Message156
- Calibration Problem124
- Incorrect Or Inadequate Test Results108
- Adverse Event Without Identified Device or Use Problem41
- No Apparent Adverse Event29
- Break23
Patient problems
- No Consequences Or Impact To Patient633
- Hyperglycemia218
- Hypoglycemia114
- Blood Loss58
- Foreign Body In Patient7
- Pain4
- Skin Irritation4
- Bruise/Contusion3
Recent adverse event reports
The 25 most recent of 994 reports, in the reporter's own words as published by the FDA.
InjuryNov 3, 2023Unexpected Therapeutic Results; Use of Device Problem; Calibration Problem; Unexpected Therapeutic Results; Use of Device Problem; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA WITH A BLOOD GLUCOSE VALUE OF 500 MG/DL AND 45 MG/DL AT THE TIME OF THE EVENT BUT THE CUSTOMER WAS NOT FEELING HIGH/LOW BLOOD GLUCOSE. THE CUSTOMER ALSO REPORTED A SENSOR GLUCOSE VS BLOOD GLUCOSE READING MISMATCH. TROUBLESHOOTING WAS PERFORMED AND FOUND THAT THE BLOOD GLUCOSE VALUE WAS 326 MG/DL AND THE SENSOR GLUCOSE VALUE WAS 52 MG/DL AT THE TIME OF THE EVENT. THE INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE SENSOR AND WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 18062773 ↗MalfunctionJul 23, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Communication or Transmission Problem; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Communication or Transmission Problem
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 150 MG/DL AND SENSOR GLUCOSE WAS UNKNOWN AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10312802 ↗MalfunctionJul 22, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 70 MG/DL AND SENSOR GLUCOSE WAS 40 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308061 ↗MalfunctionJul 10, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 90 MG/DL AND SENSOR GLUCOSE WAS 58 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10256175 ↗MalfunctionJul 8, 2020Break; Patient Device Interaction Problem; Break; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CUSTOMER STATED CANNULA WAS NOT THERE AT ALL, UNSURE IF IT WAS BAD OUT OF BOX OR WAS STUCK INSIDE HIM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER ALSO CONFIRMED IT WAS NOT CRIMPED, CANNULA WAS JUST COMPLETELY GONE. CUSTOMER LOOKED IN NEEDLE HOUSING AND INSIDE SENSOR AREA AND SEES NOTHING. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10243953 ↗MalfunctionJul 2, 2020Break; Break
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE SENSOR ELECTRODE WAS BROKE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10223148 ↗MalfunctionJun 29, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 180 MG/DL AND SENSOR GLUCOSE WAS 70 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. THEY ALSO RECEIVED CALIBRATION NOT ACCEPTED ALARM AND CHANGE SENSOR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10205457 ↗MalfunctionJun 18, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 205 MG/DL AND SENSOR GLUCOSE WAS 50 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10166474 ↗MalfunctionJun 18, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 205 MG/DL AND SENSOR GLUCOSE WAS 50 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10166447 ↗MalfunctionJun 10, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 170 MG/DL AND SENSOR GLUCOSE WAS 60 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10136943 ↗InjuryMay 29, 2020Incorrect, Inadequate or Imprecise Result or Readings; No Apparent Adverse Event; Incorrect, Inadequate or Imprecise Result or Readings; No Apparent Adverse Event
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE LEVEL. CUSTOMER'S BLOOD GLUCOSE VALUE WAS 49 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER ALSO STATED THAT SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 161 MG/DL AND THE SENSOR GLUCOSE WAS 54 MG/DL AT THE TIME OF THE INCIDENT. LOW LIMIT SETTING WAS 80 MG/DL. THE CUSTOMER¿S OTHER BLOOD GLUCOSE WAS 179 MG/DL, 173 MG/DL AND THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DECLINED FOR LOW BLOOD GLUCOSE. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hypoglycemia
MAUDE report 10100380 ↗MalfunctionMay 27, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 194 MG/DL AND SENSOR GLUCOSE WAS 55 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. CUSTOMER STATED DIFFERENCE WERE 100 POINTS OFF. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10092239 ↗MalfunctionMay 26, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED VIA PHONE THAT THEY EXPERIENCING LOW BLOOD GLUCOSE LEVEL FOR 51 MG/DL. SENSOR WAS NOTIFYING GLUCOSE LEVEL 70MG/DL, IT WAS MORE THAN ACTUAL BLOOD GLUCOSE. CUSTOMER TREATED LOW BLOOD GLUCOSE LEVEL WITH CARBOHYDRATE INTAKE. CUSTOMER DECLINED TROUBLESHOOTING FOR LOW BLOOD GLUCOSE LEVEL. INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hypoglycemia
MAUDE report 10090351 ↗MalfunctionMay 4, 2020Calibration Problem; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 148MG/DL AND SENSOR GLUCOSE WAS 60MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 60 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10021169 ↗MalfunctionMay 4, 2020Calibration Problem; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 148MG/DL AND SENSOR GLUCOSE WAS 60MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 60 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10021164 ↗MalfunctionMay 4, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 148MG/DL AND SENSOR GLUCOSE WAS 60MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 60 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10021160 ↗MalfunctionApr 20, 2020Incorrect, Inadequate or Imprecise Result or Readings; Adverse Event Without Identified Device or Use Problem; Incorrect, Inadequate or Imprecise Result or Readings; Adverse Event Without Identified Device or Use Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTED SENSOR GLUCOSE VERSES BLOOD GLUCOSE DIFFERENCES THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 198 MG/DL AND 195 MG/DL AND SENSOR GLUCOSE WAS 58 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE WAS NOT IN THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 60 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9976378 ↗MalfunctionApr 13, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Material Twisted/Bent; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Material Twisted/Bent
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 223 MG/DL AND SENSOR GLUCOSE WAS 40 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 80 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE SENSOR GLUCOSE VALUE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9952068 ↗MalfunctionApr 11, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 229 MG/DL AND SENSOR GLUCOSE WAS 52 MG/DL AND 58 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9951593 ↗MalfunctionApr 4, 2020Fracture; Packaging Problem; Fracture; Packaging Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE SENSOR NEEDLE WAS BROKEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9925740 ↗MalfunctionApr 2, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 187 MG/DL AND SENSOR GLUCOSE WAS 44 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 78 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9916955 ↗MalfunctionMar 31, 2020Fracture; Fracture
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE NEEDLE WAS BROKEN OF SENSOR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9902413 ↗MalfunctionMar 30, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results
INFORMATION RECEIVED BY MEDTRONIC THE USER REPORTED INACCURATE SENSOR READINGS THAT TRIGGERED A THRESHOLD SUSPEND. THE CUSTOMER¿S BLOOD GLUCOSE WAS 150 MG/DL AND SENSOR GLUCOSE WAS 78 MG/DL AT THE TIME OF THE EVENT, THE DIFFERENCE IS OUTSIDE THE ACCEPTABLE RANGE. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 80 MG/DL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9896746 ↗MalfunctionMar 27, 2020Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem; Incorrect, Inadequate or Imprecise Result or Readings; Unexpected Therapeutic Results; Calibration Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 99 MG/DL AND THE SENSOR GLUCOSE WAS 65 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR GLUCOSE VALUES. LOW LIMIT IN SENSOR SETTING WAS 70 MG/DL. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9889714 ↗MalfunctionMar 26, 2020Unexpected Therapeutic Results; Unexpected Therapeutic Results
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE LEVEL OF OVER 130 MG/DL. THE CUSTOMER DID NOT MENTION ANY TREATMENT. THE CUSTOMER ALSO REPORTED GETTING INACCURATE READINGS WITH THE SENSOR. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 130 MG/DL AND THE SENSOR GLUCOSE READING WAS 65 MG/DL THAT TRIGGERED THE INSULIN PUMP TO SUSPEND INSULIN DELIVERY. SUSPEND ON LOW LIMIT IN SENSOR SETTINGS WAS AT 70 MG/DL. THE SENSOR WAS NOT RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9886314 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010, P150001, P150019, P150029, P160017. One submission often covers many models, so check the product and lot details in each record.
- Z-2981-2026 · Jul 13, 2026 · Open, ClassifiedMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-1742-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-1740-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1739-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-1266-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 780G (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)
- Z-1261-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 670G (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)
- Z-1259-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 630G (MMT-1714, MMT-1715, MMT-1754, MMT-1755)
- Z-1265-2025 · Jan 31, 2025 · Open, ClassifiedMiniMed insulin pump, REF: 770G (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)
- Z-0003-2025 · Jul 31, 2024 · Open, ClassifiedMiniMed Insulin Pump Device. MiniMed 670G, REF: MMT-1580, MMT-1581, MMT-1582, MMT-1780, MMT-1781, MMT-1782, MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762; MiniMed 770G
- Z-0001-2025 · Jul 31, 2024 · Open, ClassifiedMiniMed 630G Insulin Pump, REF: MMT-1515, MMT-1714, MMT-1715, MMT-1754, MMT-1755; MiniMed 700G Insulin Pump, REF: MMT-1800, MMT-1801, MMT-1805, MMT-1850, MMT-1851
- Z-0193-2023 · Sep 20, 2022 · Open, ClassifiedInsulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
- Z-0855-2022 · Jan 27, 2022 · Open, ClassifiedMiniMed 630G: MMT-1715, MMT-1754, MMT-1755; MiniMed 670G: MMT-1740, MMT-1741, MMT-1742, MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782; MiniMed 770G: MMT-1880, MMT-18
- Z-1771-2021 · Mar 15, 2021 · Open, ClassifiedMiniMed 780G Insulin Pump: Models: PUMP MMT-1885 MM780G 6.5V BLE MMOL; PUMP MMT-1886 MM780G 6.5W BLE MG; KIT MMT-1895WW MM780G 6.5V BLE SF MMOL; KIT MMT-1896WW MM780G 6.5W BLE
…and 11 more.
Other Automated insulin dosing , threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.