Anchor Tissue Retrieval System TRS-ROBO-8LRecalls & reported problems
Conmed Corporation · ANCHOR TISSUE RETRIEVAL SYSTEM, BAG SIZE 175 ML, PORT SIZE 8 MM · Class 2
All models of ANCHOR TISSUE RETRIEVAL SYSTEM (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20653405988523), received between Sep 18, 2024 and Oct 24, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn2
- Difficult to Remove1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2025Difficult to Remove; Material Split, Cut or Torn; Difficult to Remove; Material Split, Cut or Torn
A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE TRS-ROBO-8L, ANCHOR TISSUE RETRIEVAL SYSTEM DEVICE WAS USED IN A ROBOTIC LAPAROSCOPIC APPENDECTOMY PROCEDURE ON (B)(6) 2025, AND ¿SPECIMEN BAG RIPPED HORIZONTALLY 1 INCH FROM THE BASE OF THE BAG WHILE THE SPECIMEN BAG WAS BEING REMOVED FROM THE PATIENT THROUGH THE INCISION. SPECIMEN GOT STUCK IN BETWEEN PATIENT'S SKIN AND FASCIA. BAG WAS NOT PUNCTURED DURING PROCEDURE.¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE AND A DELAY OF 15 MINUTES. FURTHER ASSESSMENT FOUND THE SPECIMEN WAS RETRIEVED WITH A CARMALT INSTRUMENT AND THE SURGICAL FIELD WAS NOT CONTAMINATED. THE BAG DID NOT FRAGMENT, BUT "RIPPED IN ONE PLACE. AND THE SPECIMEN GOT STUCK BEHIND IT" AND "SURGICAL RETRIEVAL OF THE SPECIMEN WAS NEEDED.THE SPECIMEN WAS LARGER THAN THE INCISION CREATED TO REMOVE IT. SO, THE BAG RIPPED WHILE TRYING TO PULL IT OUT THROUGH AN INCISION THAT WAS MUCH SMALLER THAN THE SPECIMEN INSIDE OF THE BAG. THEY HAD TO MAKE A LARGER INCISION TO GET THE ACTUAL SPECIMEN OUT.". THE PATIENT'S STATUS IS "DOING WELL.". THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23379349 ↗MalfunctionSep 18, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE TRS-ROBO-8L, ANCHOR TISSUE RETRIEVAL SYSTEM DEVICE WAS BEING USED DURING A ROBOTIC CHOLE PROCEDURE ON (B)(6)2024, AND ¿WHILE THE PA WAS REMOVING THE BAG WITH SPECIMEN INSIDE THE BOTTOM OF THE BAG RIPPED ALONG THE BOTTOM SEAM LEAVING THE SPECIMEN IN THE PATIENT BETWEEN THE SKIN AND FASCIA. THE PA WAS ABLE TO REMOVE THE SPECIMEN WITH LITTLE TROUBLE AND INTACT.¿. FOLLOW-UP ASSESSMENT FOUND THERE HAD BEEN NO FRAGMENTATION OF THE BAG. "THE BAG RIPPED ALONG THE LOWER SEAM, ALONG THE CURVE OF THE BAG"; "THE BAG WAS IN ONE PIECE AND ALL PIECES WERE COLLECTED FOR THE REP.¿; "THE GALLBLADDER THAT WAS BEING REMOVED WAS INTACT WHEN REMOVED"; AND "THE GALLSTONE WAS 3CMX2.5CM IN DIAMETER". THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE. "IT TOOK THE PA ABOUT 10-15 MINS TO REMOVED THE WHOLE GALLBLADDER FROM THE PATIENT. IT WAS THE END OF THE CASE; REMOVAL OF THE SPECIMEN WAS ALL THAT WAS LEFT.". THERE WAS NO INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION REQUIRED FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20245370 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page