Skip to content

RELIEVA SPINPLUS NAV RSP0616MFSNZRecalls & reported problems

Acclarent, Inc. · RELIEVA SPINPLUS NAV BALLOON SINUPLASTY SYSTEM, 6X16MM, 3 GUIDE KIT - 1PK · Class 2

All models of RELIEVA SPINPLUS NAV (2)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705031245430), received between Aug 4, 2021 and Jan 19, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Incorrect Interpretation of Signal1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 19, 2026Deformation Due to Compressive Stress

    IT WAS REPORTED THAT DURING AN IN-OFFICE BALLOON SINUPLASTY PROCEDURE, THE TIP OF THE NAVIGATED WIRE OF THE SPINPLUS NAV 6X16MM 3 GUIDE (RSP0616MFSNZ) BECAME TOO KINKED FOR THE SURGEON TO COMPLETE THE PROCEDURE WITH THE DEVICE. "IT WAS PERMANENTLY KINKED IN THREE DIFFERENT LOCATIONS." IT WAS REPORTED THAT THERE WAS A 30 MINUTE DELAY DURING THE PROCEDURE. NO PATIENT ADVERSE EVENT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24112340 ↗
  • MalfunctionAug 4, 2021Incorrect Interpretation of Signal; Incorrect Interpretation of Signal

    THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN IN-OFFICE PRIMARY HYBRID FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE THAT WAS AIMED TO TARGET ALL SIX SINUSES; IT WAS REPORTED THAT ACCURACY OF THE 6MM X 16MM RELIEVA SPINPLUS NAVIGATION BALLOON SINUPLASTY SYSTEM (RSP0616MFSNZ / 210525A-PC) WAS ¿OFF¿ ON THE TRUDI NAVIGATION SYSTEM. THE POSITIONING WAS VERIFIED AND ¿EVERYTHING WAS GREEN.¿ THE PROBE ACCURACY WAS REPORTED TO ¿LOOKED GOOD.¿ THE SPINPLUS NAVIGATION BALLOON WAS REPLACED, BUT THERE WAS NO RESOLUTION. THE PROCEDURE WAS CONTINUED AND COMPLETED WITH A REPLACEMENT PRODUCT. THE REPORTED ISSUE OCCURRED AFTER THE BALLOON HAD BEEN INFLATED TWICE. THE GUIDE / BALLOON WAS NOT FLUSHED PRIOR TO USE. THE BALLOON WAS APPROPRIATELY FOLDED PRIOR TO INSERTION. THERE WAS NO REPORT OF ANY PATIENT ADVERSE EVENT OR COMPLICATION. ADDITIONAL INFORMATION WAS RECEIVED ON 20 JULY 2021. THE INFORMATION INDICATED THAT THE POSITION WAS VERIFIED INTERNALLY USING ANATOMICAL LANDMARKS AND EXTERNALLY USING ALONG THE BRIDGE OF THE NOSE. ALL STATUS SYMBOLS ON THE WORKSTATION INDICATED PROPERLY CONNECTED INSTRUMENTS AND PROPERLY PLACED WITHIN THE TRUDI ZONE WITHOUT METAL INTERFERENCE. THERE WERE N

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12277822 ↗

Other Ear, Nose, and Throat Stereotaxic Instrument devices

All Ear, Nose, and Throat Stereotaxic Instrument devices →