RELIEVA SPINPLUS NAV RSP0616MFSNZRecalls & reported problems
Acclarent, Inc. · RELIEVA SPINPLUS NAV BALLOON SINUPLASTY SYSTEM, 6X16MM, 3 GUIDE KIT - 1PK · Class 2
All models of RELIEVA SPINPLUS NAV (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705031245430), received between Aug 4, 2021 and Jan 19, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Incorrect Interpretation of Signal1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 19, 2026Deformation Due to Compressive Stress
IT WAS REPORTED THAT DURING AN IN-OFFICE BALLOON SINUPLASTY PROCEDURE, THE TIP OF THE NAVIGATED WIRE OF THE SPINPLUS NAV 6X16MM 3 GUIDE (RSP0616MFSNZ) BECAME TOO KINKED FOR THE SURGEON TO COMPLETE THE PROCEDURE WITH THE DEVICE. "IT WAS PERMANENTLY KINKED IN THREE DIFFERENT LOCATIONS." IT WAS REPORTED THAT THERE WAS A 30 MINUTE DELAY DURING THE PROCEDURE. NO PATIENT ADVERSE EVENT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24112340 ↗MalfunctionAug 4, 2021Incorrect Interpretation of Signal; Incorrect Interpretation of Signal
THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN IN-OFFICE PRIMARY HYBRID FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE THAT WAS AIMED TO TARGET ALL SIX SINUSES; IT WAS REPORTED THAT ACCURACY OF THE 6MM X 16MM RELIEVA SPINPLUS NAVIGATION BALLOON SINUPLASTY SYSTEM (RSP0616MFSNZ / 210525A-PC) WAS ¿OFF¿ ON THE TRUDI NAVIGATION SYSTEM. THE POSITIONING WAS VERIFIED AND ¿EVERYTHING WAS GREEN.¿ THE PROBE ACCURACY WAS REPORTED TO ¿LOOKED GOOD.¿ THE SPINPLUS NAVIGATION BALLOON WAS REPLACED, BUT THERE WAS NO RESOLUTION. THE PROCEDURE WAS CONTINUED AND COMPLETED WITH A REPLACEMENT PRODUCT. THE REPORTED ISSUE OCCURRED AFTER THE BALLOON HAD BEEN INFLATED TWICE. THE GUIDE / BALLOON WAS NOT FLUSHED PRIOR TO USE. THE BALLOON WAS APPROPRIATELY FOLDED PRIOR TO INSERTION. THERE WAS NO REPORT OF ANY PATIENT ADVERSE EVENT OR COMPLICATION. ADDITIONAL INFORMATION WAS RECEIVED ON 20 JULY 2021. THE INFORMATION INDICATED THAT THE POSITION WAS VERIFIED INTERNALLY USING ANATOMICAL LANDMARKS AND EXTERNALLY USING ALONG THE BRIDGE OF THE NOSE. ALL STATUS SYMBOLS ON THE WORKSTATION INDICATED PROPERLY CONNECTED INSTRUMENTS AND PROPERLY PLACED WITHIN THE TRUDI ZONE WITHOUT METAL INTERFERENCE. THERE WERE N
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12277822 ↗
Other Ear, Nose, and Throat Stereotaxic Instrument devices
- TruDi NAV Suction TDNS000ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall61 reports
- Scopis 8000-021-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument1 recall19 reports
- FUSION Compact ™ 9735602Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument64 reports
- TruDi™ Navigation System 1Biosense Webster (Israel) Ltd. · Ear, Nose, and Throat Stereotaxic Instrument1 recall14 reports
- StealthStation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- StealthStation Cranial Software 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Medtronic Navigation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Scopis ENT 8000-020-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument33 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page