Skip to content

OUTBACK OTB42120Recalls & reported problems

Cordis Corporation · OUTBACK LTD 120cm Re-Entry · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20705032035016), received between Jul 10, 2017 and May 10, 2022 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Fracture2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Dislodged or Dislocated1
  • Difficult or Delayed Positioning1

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • Perforation1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 10, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AS REPORTED, THE ¿L¿ MARKER OF AN OUTBACK LTD RE-ENTRY CATHETER 120CM, DETACHED. IT DETACHED DURING USE IN THE VESSEL AND IS IN THE VESSEL. THE USER IS TRAINED WITH THE DEVICE. THE DEVICE WILL BE RETURNED FOR EVALUATION. ADDENDUM FROM ADDITIONAL INFO: THE "L" MARKER WAS NOTED BE DISLODGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14349413 ↗
  • InjuryOct 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AFTER USING THE DEVICE AND ADVANCING THE NON CORDIS GUIDEWIRE INTO THE VESSEL LUMEN, THE OUTBACK LTD KINKED/BENT WHILE GOING OVER THE BIFURCATION AND WOULD NOT RETRACT WITHOUT PULLING IT OUT TOGETHER WITH THE WIRE AS ONE UNIT. THE WIRE COULD NOT BE REMOVED FROM THE DEVICE UNTIL THE WIRE WAS OUT OF THE BODY. THEREFORE, THE PROCEDURE COULD NOT BE SUCCESSFULLY COMPLETED DUE TO LOSS OF WIRE PLACEMENT AND IT CAUSED A PERFORATION OF THE SUPERFICIAL FEMORAL ARTERY (SFA) THAT WAS IMMEDIATELY TREATED USING MANUAL PRESSURE FROM A BLOOD PRESSURE CUFF SINCE WIRE ACCESS HAD BEEN LOST. THE TARGET VESSEL WAS THE SFA. THE ACCESS SITE WAS FEMORAL. THE DEVICE WAS PREPPED AS PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DIFFICULTY EXPERIENCED IN PREPPING THE DEVICE. THE SFA HAS MODERATE CALCIFICATION, NO TORTUOSITY, 100% STENOSIS WITH A CHRONIC TOTAL OCCLUSION (CTO), NO ACUTE ANGLE, AND NO BIFURCATION. THERE WAS NO DAMAGE TO THE OUTBACK DEVICE NOTICED PRIOR TO OPENING THE PACKAGE AND NO APPARENT DAMAGE TO THE DEVICE NOTICED PRIOR TO USE. THE PORTS WERE FLUSHED, FOLLOWED BY A 30 SECOND PAUSE, AND THEN FLUSHED AGAIN. HEPARINIZED SALINE WAS NOT USED FOR PREPARATION AND FLUSHING OF ALL PORTS. THE CATH

    Patient problems: Perforation

    MAUDE report 10755464 ↗
  • MalfunctionOct 19, 2020Fracture; Fracture

    AS REPORTED, THE CANNULA TIP OF A 120CM OUTBACK RE-ENTRY CATHETER COULD NOT BE RETRACTED WHEN CHECKING THE FUNCTIONALITY OF THE DEVICE. WHILE CHECKING THE FUNCTIONALITY OF THE OUTBACK, THE CANNULA TIP COULD NOT BE RETRACTED. THE PHYSICIAN TRIED SEVERAL TIMES, BUT IT FELL OFF AND WAS BROKEN. THERE WAS NO REPORTED PATIENT INJURY. THE CANNULA TIP COVER WAS REMOVED AND THE DEVICE WAS FLUSHED. THE DEVICE WAS OPENED IN STERILE FILED. THE DEVICE WAS STORED IN THE CATH LAB IN A COOL DRY PLACE, FOR APPROXIMATELY 9 MONTHS. THE DEVICE WAS STORED, HANDLED, AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO DAMAGE NOTICED TO THE DEVICE PRIOR TO OPENING THE PACKAGE. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE STERILE PACKAGING. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10698179 ↗
  • MalfunctionJul 10, 2017Difficult or Delayed Positioning; Fracture; Difficult or Delayed Positioning; Fracture

    IT WAS REPORTED THAT THE OUTBACK RE-ENTRY CATHETER WAS SEPARATED AND THE NEEDLE DID NOT DEPLOY. THERE WERE NO PATIENT INJURY. THE PHYSICIAN TRIED TO RE-ENTER THE ARTERY BY USING THE OUTBACK. HE PUSHED THE BUTTON ON THE HANDLE TO DEPLOY THE NEEDLE BUT THE NEEDLE DIDN'T DEPLOY. HE NOTICED THAT THE HANDLE AND THE BLACK SLEEVE WAS DETACHED. SO HE GRABBED THE HANDLE AND THE BLACK SLEEVE, THEN THE NEEDLE WAS DEPLOYED. THE PATIENT WAS CURED WELL. HE SAID THAT THE HANDLE AND THE BLACK SLEEVE WAS DETACHED FROM THE FIRST.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6697253 ↗

FDA clearance and classification

The FDA regulates product code PDU (catheter for crossing total occlusions) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the OUTBACK OTB42120

Has the OUTBACK OTB42120 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20705032035016), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the OUTBACK OTB42120?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 10, 2017 and May 10, 2022. In those reports, the most common specific device problems are “Fracture” (2), “Device Dislodged or Dislocated” (1) and “Difficult or Delayed Positioning” (1); the most common patient problem is “Perforation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the OUTBACK OTB42120?

None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the OUTBACK OTB42120, and is it still sold?

The FDA's device database lists Cordis Corporation as the company responsible for it (brand OUTBACK, model OTB42120). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter for crossing total occlusions devices

All Catheter for crossing total occlusions devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page