Skip to content

ELITECROSS Support Catheter MGC39140XRecalls & reported problems

Cordis US Corp. · Support Catheter · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032067536), received between Jul 8, 2024 and Nov 27, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Stretched1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 27, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, DURING DEEP VEIN LAST WEEK, WHILE USING ONE OF THE 132CM ELITECROSS EXTRA SUPPORT CATHETER, THE WHITE END PIECE ENDING UP BREAKING OFF AFTER ONE USE. THE DEVICE WILL BE RETURNED FOR EVALUATION. THE USER WAS TRAINED IN USING THE DEVICE. THE DEVICE WAS STORED AND OPENED IN A STERILE FIELD. THE DEVICE WAS USED IN THE PATIENT. ADDITIONAL INFORMATION AND DEVICE RETURN WAS REQUESTED; HOWEVER, NEITHER WERE NOT OBTAINED AFTER MULTIPLE ATTEMPTS MADE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20797316 ↗
  • MalfunctionJul 8, 2024Stretched; Unraveled Material; Detachment of Device or Device Component; Stretched; Unraveled Material; Detachment of Device or Device Component

    AS REPORTED, THE PHYSICIAN USED TWO (2) ELITECROSS 132CM EXTRA SUPPORT CATHETERS ON A PATIENT ALONG WITH AN UNKNOWN HYDROPHILIC WIRE TO ADVANCE THE ELITECROSS PROXIMAL TO A STENTED PORTION IN THE PATIENT¿S SUPERFICIAL FEMORAL ARTERY (SFA). UPON ADVANCEMENT OF THE FIRST ELITECROSS OVER THE WIRE, IT WAS NOTICED THAT THE CATHETER WAS UNRAVELING. THE PHYSICIAN THEN PROCEEDED TO PULL A SECOND ELITECROSS CATHETER AND ADVANCED IT OVER A HYDROPHILIC WIRE DISTAL TO THE STENTED PORTION IN THE PATIENT¿S SFA AND NOTICED THAT THIS ELITECROSS WAS ALSO UNRAVELING, AND A PORTION OF THE CATHETER HAD BROKEN OFF AND WAS LEFT IN THE ARTERY. THE PHYSICIAN REPORTED USING A .035 NON-CORDIS SUPPORT CATHETER TO RECAPTURE THE PIECE OF THE ELITECROSS THAT WAS BROKEN OFF IN THE ARTERY. THE RECOVERED PIECE IS INCLUDED IN THE COMPLAINT BAG. THE HOSPITAL IS UNABLE TO CONFIRM THE LOT NUMBER FOR EITHER ONE OF THESE ELITECROSS CATHETERS. MULTIPLE ATTEMPTS TO GATHER ADDITIONAL INFORMATION HAVE BEEN MADE AND HAVE BEEN UNSUCCESSFUL. THE DEVICES WERE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19697625 ↗

Other Catheter for crossing total occlusions devices

All Catheter for crossing total occlusions devices →