LIGACLIP LX207Recalls & reported problems
Ethicon Endo-Surgery, LLC · Ligaclip Clip Applier Medium 19cm · Class 1
Quick summary
This Ethicon Ligaclip LX207 is a surgical clip applier, a handheld tool that places and closes metal clips on vessels or tissue during surgery. The 34 reports mostly describe the applier failing to form the clip properly, jaws that would not close or lock, misalignment, or clips breaking off; most mention no effect on the patient, and there are no deaths and 1 injury. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 20705036012891), received between Apr 16, 2018 and Sep 16, 2026 — classified as 1 injury, 34 malfunctions.
The most frequently reported specific device problem is “Failure to Form Staple” (24 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Form Staple24
- Failure to Align5
- Mechanical Problem5
- Difficult to Open or Close4
- Appropriate Device Problem Term/Code Not Available3
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Corroded1
Patient problems
- No Consequences Or Impact To Patient19
- No Clinical Signs, Symptoms or Conditions14
- Cardiac Arrest1
- Hemorrhage/Blood Loss/Bleeding1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Failure to Form Staple; Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788303 ↗MalfunctionFeb 10, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, FUNCTIONAL ISSUES WITH ALL FIVE CLIP APPLIERS IN THE LOT. SPECIFICALLY, THE LAPRA-TY DEVICES DID NOT SEAT PROPERLY ON THE JAWS AND WOULD NOT CLOSE OR LOCK AS INTENDED. IN ADDITION, BOTH THE HEM-O-LOCK BLUE CLIP APPLIERS AND THE WHITE CLIP APPLIERS WERE UNABLE TO HOLD THE CLIPS SECURELY AND WERE NOTED TO BE MISALIGNED. NO PATIENT CONSEQUENCE WERE REPORTED IN ASSOCIATION WITH THESE FAILURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24305508 ↗MalfunctionFeb 10, 2026Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNK PROCEDURE, FUNCTIONAL ISSUES WITH ALL FIVE CLIP APPLIERS IN THE LOT. SPECIFICALLY, THE LAPRA-TY DEVICES DID NOT SEAT PROPERLY ON THE JAWS AND WOULD NOT CLOSE OR LOCK AS INTENDED. IN ADDITION, BOTH THE HEM-O-LOCK BLUE CLIP APPLIERS AND THE WHITE CLIP APPLIERS WERE UNABLE TO HOLD THE CLIPS SECURELY AND WERE NOTED TO BE MISALIGNED. NO PATIENT CONSEQUENCE WERE REPORTED IN ASSOCIATION WITH THESE FAILURES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24305420 ↗MalfunctionOct 30, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT THE CLIP BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23423527 ↗MalfunctionOct 30, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT THE CLIP BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23423525 ↗MalfunctionApr 24, 2025Difficult to Open or Close; Difficult to Open or Close
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE GETTING STUCK IN BETWEEN WHILE LODGING THE CLIPS, IT WAS NOT OPENING BACK ON ITS OWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21901020 ↗InjuryNov 19, 2024Mechanical Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Failure to Form Staple; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, DEVICE HAS ASYMMETRICAL JAWS AND THE CLIPS ARE NOT FORMED PROPERLY.
Patient problems: Cardiac Arrest; Hemorrhage/Blood Loss/Bleeding
MAUDE report 20725214 ↗MalfunctionNov 29, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A GASTRECTOMY, CLIPS WERE OPENING UP. 15 PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18233766 ↗MalfunctionNov 29, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A GASTRECTOMY, CLIPS WERE OPENING UP. 15 PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18233755 ↗MalfunctionOct 26, 2023Failure to Form Staple; Material Integrity Problem; Failure to Form Staple; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, CLIP APPLIER WILL NO LONGER HOLD THE CLIPS AND THE COATING IS CRACKED ON THE HANDLE AND A FEW THE TIPS ARE CROSSED. THERE WAS NO ADVERSE CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18011513 ↗MalfunctionSep 21, 2023Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS DO NOT BITE WELL. THERE WAS NO ADVERSE CONSEQUENCE TO THE PATIENT. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17790844 ↗MalfunctionAug 7, 2023Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WERE MISALIGNED/BENT. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477168 ↗MalfunctionAug 7, 2023Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WERE MISALIGNED/BENT. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477151 ↗MalfunctionAug 7, 2023Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS WERE MISALIGNED/BENT. NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477135 ↗MalfunctionAug 5, 2021Mechanical Problem; Failure to Form Staple; Mechanical Problem; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE HAD LACK OF BITE AND FAULTY CLIP FORMATION. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. PATIENT CONSEQUENCE WAS NOT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12285574 ↗MalfunctionNov 13, 2020Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING A THYROIDECTOMY PROCEDURE, THE FORMATION OF THE CLIPS WAS NOT OPTIMAL. THE LEGS DIDN'T CLOSE SYMMETRICALLY, WHICH LEAVES A SMALL PORTION OF THE LEG WITH NO COMPRESSION. SEVERAL CLIPS HAD TO BE USED TO OBTAIN HEMOSTASIS. THE CLIP FORMATION IN CLIP APPLIER WAS NOT OPTIMAL, THE LEGS DON¿T CLOSE SYMMETRICALLY, WHICH LEAVES A PORTION OF THE TISSUE WITH NO COMPRESSION. THE PORTION WITH NO COMPRESSION WAS MILLIMETRIC. THERE WAS NO DELAY TO PROCEDURE, SURGERY WAS SUCCESSFULLY COMPLETED, AND THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10836312 ↗MalfunctionAug 5, 2020Failure to Form Staple; Failure to Form Staple
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE APPLICATOR CONTROLLER DOES NOT TIGHTEN THE STAPLE COMPLETELY. IT IS UNKNOWN HOW PROCEDURE WAS COMPLETED. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10370001 ↗MalfunctionMay 27, 2020Mechanical Problem; Failure to Form Staple; Mechanical Problem; Failure to Form Staple
IT WAS REPORTED BY SENIOR OPERATING NURSE THAT THERE WAS A DEFECT IN THE WORKING PART OF THE CLIP APPLIER (DIFFERENT JAW LENGTHS). ACCORDING TO THE NURSE, THE CLIP APPLIER DELIVERED TO THE CLINIC HAS AN OBVIOUS DEFECT. THE DIFFERENT LENGTHS OF THE "CLIPS" OF CLIP APPLIER, WHICH PREVENTS THE CLIP FROM CLOSING CORRECTLY (THE CLIP IS DEFORMED AND DOESN'T CLOSE COMPLETELY). THE START OF WORK OF THE CLIP APPLIER IN THE "OPERBLOCK" WAS SUPPOSED TO BE ON (B)(6) 2020, BUT AFTER (IN THAT DATE) PROBLEM WAS IDENTIFIED. SENIOR OPERATING NURSE, WHO HAD BEEN EXTENSIVE EXPERIENCE WORKING WITH A SIMILAR MEDICAL PRODUCT EARLIER, POINTED THE ABOVE DEFECT AND PERFORMED A CONTROL CHECK OF THE QUALITY OF THIS MEDICAL DEVICE (QUALITY OF CLOSURE THE CLIPS) BEFORE GIVING THE INSTRUMENT TO THE SURGEON. THERE WAS NO OPERATION, NO PATIENT INVOLVEMENT, AND NO ADVERSE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10094559 ↗MalfunctionDec 11, 2019Failure to Form Staple
IT WAS REPORTED THAT DURING A PEDIATRIC THORACIC PROCEDURE, THE SURGEON DID AN ECHOCARDIOGRAM AND FOUND THAT THE CLIP WAS SCISSORED WHERE HE HAD APPLIED IT. ACCORDING TO THE CUSTOMER THE CLIP DID WHAT IS WAS SUPPOSED TO DO AND THERE WERE NO PATIENT CONSEQUENCES, BUT THE SURGEON NOTICED THAT IS WAS SCISSORED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9454823 ↗MalfunctionJun 13, 2019Corroded; Failure to Align; Failure to Form Staple
IT WAS REPORTED THAT THE CLIP APPLIER PRESENTED WITH OXIDIZED RINGS AND BAD CLIP FORMATION DUE TO THE MISALIGNED JAWS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8696796 ↗MalfunctionJun 13, 2019Failure to Align; Failure to Form Staple
IT WAS REPORTED THAT THE CLIP APPLIER PRESENTED WITH OXIDIZED RINGS AND BAD CLIP FORMATION DUE TO THE MISALIGNED JAWS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8696739 ↗MalfunctionMay 6, 2019Failure to Align
IT WAS REPORTED THAT DURING A CABG, THE QUALITY ISSUE PRESENTED WAS MISALIGNMENT IN THE JAWS OF THE DEVICE, CAUSING CLIPS MALFORMATION, DETACHING FROM THE PATIENT¿S TISSUE. THERE WERE NO CONSEQUENCES TO THE PATIENT. IT WAS NECESSARY TO USE ANOTHER DEVICE TO CONTINUE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8581971 ↗MalfunctionApr 10, 2019Failure to Align; Device Fell
IT WAS REPORTED THAT THE JAWS OF THE CLIP APPLIERS ARE MISALIGNED, RESULTING IN THE CLIPS FALLING OUT OF THE JAWS OF THE INSTRUMENT DURING USE OR NOT LOCKING ON TISSUE DURING USE. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8500193 ↗MalfunctionJan 18, 2019Failure to Form Staple
IT WAS REPORTED THAT THE CLIP APPLIER IS NOT HOLDING THE CLIPS. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8262014 ↗MalfunctionJan 18, 2019Failure to Form Staple
IT WAS REPORTED THAT THE CLIP APPLIER IS NOT HOLDING THE CLIPS. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8261995 ↗
Other Applier, surgical, clip devices
- LIGACLIP MCS20Ethicon Endo-Surgery, LLC · Applier, surgical, clip1 death337 reports
- Endo Clip III 176630BCovidien LP · Applier, surgical, clip312 reports
- LIGACLIP MCM20Ethicon Endo-Surgery, LLC · Applier, surgical, clip308 reports
- LIGACLIP MCM30Ethicon Endo-Surgery, LLC · Applier, surgical, clip1 death236 reports
- WECK IPN915189Teleflex Incorporated · Applier, surgical, clip221 reports
- LIGACLIP MSM20Ethicon Endo-Surgery, LLC · Applier, surgical, clip152 reports
- LIGACLIP LX107Ethicon Endo-Surgery, LLC · Applier, surgical, clip105 reports
- WECK IPN028280Teleflex Incorporated · Applier, surgical, clip93 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.