Bausch + Lomb BL5110Recalls & reported problems
Bausch & Lomb Incorporated · Premium Vacuum Phaco Pack · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 20757770057010), received between Dec 21, 2016 and Oct 22, 2025 — classified as 3 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Particulates” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Particulates7
- Difficult to Insert4
- Device Contamination with Chemical or Other Material1
- Leak/Splash1
- Product Quality Problem1
Patient problems
- No Known Impact Or Consequence To Patient4
- Insufficient Information3
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions2
- Capsular Bag Tear1
- Corneal Edema1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
InjuryOct 22, 2025Product Quality Problem; Product Quality Problem
A USER FACILITY IN THE UNITED KINGDOM REPORTED THAT DURING CATARACT SURGERY BUBBLES OCCURRED AND OBSCURED THE SURGEONS VIEW. THIS CAUSED THE PHACO TIP TO TOUCH THE CAPSULE BAG AND RESULTED IN A RUPTURE. AN ANTERIOR VITRECTOMY WAS PERFORMED AND SURGERY WAS INCREASED BY 15-30 MINUTES. NO ADDITIONAL ANESTHESIA WAS REQUIRED.
Patient problems: Capsular Bag Tear
MAUDE report 23359928 ↗InjuryJul 10, 2024Leak/Splash; Leak/Splash
IT WAS REPORTED BY A USER FACILITY IN TURKEY THAT DURING THE CASE, WATER LEAKED FROM THE CASSETTE CHAMBER OF THE DEVICE. THE CASE WAS PAUSED UNTIL THE PUMP WAS DRIED AND A NEW CASSETTE WAS INSTALLED. DUE TO THE DRYING TIME OF THE CASSETTE CHAMBER THE CASE WAS PROLONGED BY APPROXIMATELY ONE HOUR. SEDATION WAS GIVEN TO THE PATIENT. IN ADDITION, WHILE THE PATIENT WAS UNDER THE EFFECT OF LOCAL ANESTHESIA, THEY MOVED CAUSING AN EDEMA IN THE CORNEA. THE PATIENT IS NOW IN GOOD CONDITION.
Patient problems: Corneal Edema
MAUDE report 19709103 ↗MalfunctionNov 1, 2023Particulates; Particulates
A USER FACILITY IN FRANCE REPORTED THAT THREE SMALL METAL WIRES CAME OUT OF THE ULTRASONIC HANDPIECE AND ENTERED THE EYE AS SOON AS THE INFUSION WAS ACTIVATED. THE FOREIGN MATERIAL WAS REMOVED WITH TROUTMAN FORCEPS.
Patient problems: Insufficient Information
MAUDE report 18048626 ↗MalfunctionNov 1, 2023Particulates; Particulates
A USER FACILITY IN FRANCE REPORTED THAT THREE SMALL METAL WIRES CAME OUT OF THE ULTRASONIC HANDPIECE.
Patient problems: Insufficient Information
MAUDE report 18048509 ↗MalfunctionNov 1, 2023Particulates; Particulates
A USER FACILITY IN FRANCE REPORTED THAT THREE SMALL METAL WIRES CAME OUT OF THE ULTRASONIC HANDPIECE.
Patient problems: Insufficient Information
MAUDE report 18048422 ↗MalfunctionApr 7, 2023Particulates; Particulates
A USER FACILITY IN THE UNITED KINGDOM REPORTED THAT THE SURGEON NOTICED A SMALL FOREIGN BODY FLUSH INTO ANTERIOR CHAMBER AT START OF PHACO. THE OBJECT WAS REMOVED. IT WAS DESCRIBED AS ¿WHITE PLASTIC¿, AND TRANSLUCENT MATERIAL. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16702841 ↗MalfunctionMar 28, 2023Particulates; Particulates
A USER FACILITY IN THE UNITED KINGDOM REPORTED A PIECE OF PLASTIC ENTERED THE EYE OUT OF THE PHACO TIP NEEDLE. THE PIECE OF PLASTIC WAS RETRIEVED FROM THE EYE AND THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16628402 ↗InjuryMay 23, 2018Difficult to Insert; Difficult to Insert
THE CUSTOMER REPORTED THAT A BLUE PARTICLE WAS OBSERVED IN THE EYE. THE PARTICLE WAS REMOVED FROM THE EYE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7538628 ↗MalfunctionDec 21, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
LIGHT BLUE RESIDUE FOUND INSIDE THE EYE OF THE PATIENT. IT WAS IMMEDIATELY REMOVED BY THE SURGEON. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7140305 ↗MalfunctionDec 19, 2017Difficult to Insert; Difficult to Insert
THE SLEEVE LEFT ONE SMALL LIGHT BLUE PARTICLE IN THE PATIENT'S EYE. THE SURGEON WAS ABLE TO REMOVE IT IMMEDIATELY. NO PATIENT INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7130567 ↗MalfunctionMay 8, 2017Difficult to Insert; Difficult to Insert
THE SURGEON REPORTED THE SLEEVE DID NOT ENTER THE INCISION SMOOTHLY AND WAS DIFFICULT TO INSERT. THERE WAS NO REPORTED IMPACT TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6552689 ↗MalfunctionMay 8, 2017Difficult to Insert; Difficult to Insert
THE SLEEVE COLLAPSED DURING INSERTION AND WOULD NOT ENTER THE INCISION SMOOTHLY. ADDITIONAL FORCE WAS NEEDED TO INSERT THE SLEEVE. THERE WAS NO REPORTED IMPACT TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6552675 ↗MalfunctionApr 25, 2017Particulates; Particulates
A REPORT RECEIVED FROM A USER FACILITY IN THE (B)(6) STATED THAT DURING A CATARACT PROCEDURE, A PARTICLE WAS SEEN COMING OUT OF THE IRRIGATION PORT AT THE BEGINNING OF A CASE. THERE WAS NO REPORT OF INJURY OR CONSEQUENCES TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6519874 ↗MalfunctionDec 21, 2016Particulates; Particulates
THE USER FACILITY REPORTS OBSERVING WHAT APPEARS TO BE GLITTER IN THE EYE WOUNDS OF SEVERAL PATIENTS. THE GLITTER HAS BEEN OBSERVED DURING POST-OP EXAMINATION WITH A SLIT LAMP OVER THE PAST SEVERAL MONTHS. THE USERS REPORTED NOTICING THE SLEEVES THAT COVER THE PHACO TIP AND I/A TIPS APPEAR TO HAVE GLITTER OR A SHIMMER ON THE SURFACE. TO DATE, THERE HAVE BEEN NO PATIENT COMPLICATIONS DUE TO THIS ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6195360 ↗
Other Unit, Phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page