Bausch + Lomb BL3124SRecalls & reported problems
Bausch & Lomb Incorporated · 2.4 MM Sleeve for BL3318 and BL3420 · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 20757770059618), received between Feb 12, 2024 and Mar 13, 2026 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Particulates” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Particulates12
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Insufficient Information2
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 13, 2026Particulates
THE USER FACILITY IN AUSTRALIA REPORTED THAT AFTER THE NUCLEUS HAD BEEN EXTRACTED DURING CATARACT SURGERY WITH A 2.0 TIP AND 2.4 SLEEVE, WHEN THE NURSE REMOVED THE PHACO HANDPIECE FROM THE EYE TO SWAP TO IRRIGATION/ASPIRATION, PART OF THE SLEEVE TORE OFF AND REMAINED IN THE EYE. THE PARTICLE WAS ABLE TO REMOVED AND THE PROCEDURE WAS SUCCESSFUL WITH NO IMPACT TO THE PATIENT. STAFF NOTED THE SLEEVE MAY HAVE HAD A MICRO-TEAR PRIOR TO COMMENCING SURGERY THAT WAS UNDETECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24591490 ↗MalfunctionNov 20, 2025Particulates; Particulates
A USER FACILITY IN SPAIN REPORTED THAT DURING THE IRRIGATION STAGE OF EYE SURGERY, THE HOLES FROM THE SLEEVE DETACHED IN THE EYE. THEY WERE REMOVED WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23603906 ↗MalfunctionNov 20, 2025Particulates; Particulates
A USER FACILITY IN SPAIN REPORTED THAT DURING THE IRRIGATION STAGE OF EYE SURGERY, THE HOLES FROM THE SLEEVE DETACHED IN THE EYE. THEY WERE REMOVED WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23603905 ↗MalfunctionMay 14, 2025Particulates; Particulates
A USER FACILITY REPORTED THAT DURING TWO SEPARATE CATARACT SURGERIES, PERFORMED ON THE SAME DAY, THE SURGEON FOUND A CUT FROM ONE OF THE PORTS FROM THE YELLOW SLEEVE IN THE EYE. THE SURGEON EXTRACTED PARTICLES IMMEDIATELY WHEN NOTICED. THERE WERE NO COMPLICATIONS OR DELAYS IN THE SURGERIES. ONLY ONE PARTICLE WAS RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22020351 ↗MalfunctionMay 14, 2025Particulates; Particulates
A USER FACILITY REPORTED THAT DURING TWO SEPARATE CATARACT SURGERIES, PERFORMED ON THE SAME DAY, THE SURGEON FOUND A CUT FROM ONE OF THE PORTS FROM THE YELLOW SLEEVE IN THE EYE. THE SURGEON EXTRACTED PARTICLES IMMEDIATELY WHEN NOTICED. THERE WERE NO COMPLICATIONS OR DELAYS IN THE SURGERIES. ONLY ONE PARTICLE WAS RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22020314 ↗MalfunctionDec 9, 2024Particulates; Particulates
THE USER FACILITY REPORTED THAT DURING THE PROCEDURE, A PIECE OF THE SLEEVE WAS IN THE PATIENT'S EYE. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20883721 ↗MalfunctionDec 6, 2024Particulates; Particulates
A USER FACILITY IN FRANCE REPORTED THAT DURING A PROCEDURE, A PIECE OF THE SLEEVE WAS IN THE PATIENT'S EYE. IT WAS REMOVED INTRAOPERATIVELY WITH NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20860523 ↗MalfunctionDec 4, 2024Particulates; Particulates
THE USER FACILITY IN FRANCE REPORTED THAT DURING THE PROCEDURE, A PIECE OF THE SLEEVE WAS IN THE PATIENT 'S EYE. IT WAS REMOVED INTRAOPERATIVELY WITH NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20847710 ↗MalfunctionNov 5, 2024Particulates; Particulates
THE USER FACILITY IN SWITZERLAND REPORTED THAT DURING THE PROCEDURE, A ROUND SECTION OF THE SIDE PORT ON THE SLEEVE GOT INTO THE PATIENT'S EYE, ALTHOUGH THE SLEEVE HAD BEEN RINSED OUT BEFOREHAND. THE SECTION OF THE SLEEVE WAS REMOVED IMMEDIATELY FROM THE EYE USING THE SUCTION SYSTEM. THE PROCEDURE WAS EXTENDED BY APPROXIMATELY FIVE MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20613142 ↗MalfunctionNov 5, 2024Particulates; Particulates
THE USER FACILITY IN SWITZERLAND REPORTED THAT DURING THE PROCEDURE, A ROUND SECTION OF THE SIDE PORT ON THE SLEEVE GOT INTO THE PATIENT'S EYE, ALTHOUGH THE SLEEVE HAD BEEN RINSED OUT BEFOREHAND. THE SECTION OF THE SLEEVE WAS REMOVED IMMEDIATELY FROM THE EYE USING THE SUCTION SYSTEM. THE PROCEDURE WAS EXTENDED BY APPROXIMATELY FIVE MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20612276 ↗MalfunctionFeb 28, 2024Particulates; Particulates
THE USER FACILITY REPORTED THAT A SMALL CIRCULAR CUTOUT PIECE FROM THE SLEEVE'S' IRRIGATION PORT ENTERED THE PATIENT'S EYE AND THE DOCTOR WAS ABLE TO GET IT OUT. NO PATIENT IMPACT.
Patient problems: Insufficient Information
MAUDE report 18799406 ↗MalfunctionFeb 12, 2024Particulates; Particulates
THERE WERE NO PATIENT COMPLICATIONS.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 18684505 ↗
Other Unit, Phacofragmentation devices
- Centurion 8065751763Alcon Laboratories, Inc. · Unit, phacofragmentation728 reports
- Centurion 8065753057Alcon Laboratories, Inc. · Unit, phacofragmentation639 reports
- CENTURION® 8065752914Alcon Laboratories, Inc. · Unit, phacofragmentation571 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751617Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls398 reports
- Alcon 8065752201Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall386 reports
- Alcon 8065752200Alcon Laboratories, Inc. · Unit, phacofragmentation1 recall366 reports
- Constellation, Ultravit, 25+, TotalPlus, EdgePlus 8065751767Alcon Laboratories, Inc. · Unit, phacofragmentation2 recalls302 reports
- Constellation, ULTRAVIT, Engauge, 25+ 8065752415Alcon Laboratories, Inc. · Unit, phacofragmentation369 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page