Skip to content

SafeSheath® II 667464-300Recalls & reported problems

Oscor Inc. · Hemostatic Tear-away Introducer System with Infusion Side Port, 9.5F · Class 2

All models of SafeSheath II (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20802526532522), received between Feb 27, 2018 and Feb 27, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Air Leak” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K122084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Air Leak2
  • Leak/Splash2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 27, 2018Air Leak; Leak/Splash; Air Leak; Leak/Splash

    THE COMPLAINT WAS THAT THE INTRODUCERS THEY WERE USING DURING A COUPLE OF IMPLANT CASES APPEARED TO BE LEAKING. HERE IS A SUMMARY FROM THE COMPLAINT: THE PHYSICIAN COMPLAINT THAT THE SHEATHS WE USE SEEM TO BE LEAKING. AFTER GAINING ACCESS AND PUTTING A SHEATH THROUGH THE PUNCTURE SITE, WHEN ATTEMPTING TO FLUSH THE SHEATH, THERE ARE A LOT OF AIR BUBBLES THAT IS ASPIRATED AS HE DRAWS BACK THE SHEATH[?]..AS A RESULT, THERE WAS AIR INTRODUCED INTO THE PATIENT'S VESSELS AS HE ADVANCES THE LEAD INTO THE BODY. HE IS CONCERNED AS THIS PUTS PATIENT AT RISK OF EMBOLISM AND HYPOTENSION. THE FOLLOWING INFORMATION RECEIVED WAS RECEIVED ON FEBRUARY 8, 2018: EVENT DATE: (B)(6) 2017. REPORTED REGION: (B)(6). DID THE EVENT INVOLVE A PATIENT: YES. AGE AND SEX OF PATIENT: MALE (B)(6) YEARS OLD. WHAT IS THE OUTCOME OF THE PATIENT: NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO IMPACT TO THE PATIENT/STABLE: THERE WAS NO REPORTED IMPACT TO THE PATIENT. IF PATIENT IMPACT, PLEASE PROVIDE DETAILS: ANY REPORTED ADVERSE PATIENT EFFECTS: NO ADVERSE PATIENT EFFECTS WERE REPORTED. ANY EXTENDED HOSPITALIZATION: NONE REPORTED. IT WAS REPORTED THE SHEATH WAS LEAKING, HOW WAS THE PRODUCT ISSUE RESOLVED: WAS ANOTHER M

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7301640 ↗
  • MalfunctionFeb 27, 2018Air Leak; Leak/Splash; Air Leak; Leak/Splash

    THE COMPLAINT WAS THAT THE INTRODUCERS THEY WERE USING DURING A COUPLE OF IMPLANT CASES APPEARED TO BE LEAKING. HERE IS A SUMMARY FROM THE COMPLAINT: THE PHYSICIAN COMPLAINT THAT THE SHEATHS WE USE SEEM TO BE LEAKING. AFTER GAINING ACCESS AND PUTTING A SHEATH THROUGH THE PUNCTURE SITE, WHEN ATTEMPTING TO FLUSH THE SHEATH, THERE ARE A LOT OF AIR BUBBLES THAT IS ASPIRATED AS HE DRAWS BACK THE SHEATH[?]..AS A RESULT, THERE WAS AIR INTRODUCED INTO THE PATIENT'S VESSELS AS HE ADVANCES THE LEAD INTO THE BODY. HE IS CONCERNED AS THIS PUTS PATIENT AT RISK OF EMBOLISM AND HYPOTENSION. THE FOLLOWING INFORMATION RECEIVED WAS RECEIVED ON FEBRUARY 8, 2018: EVENT DATE: (B)(6) 2017. REPORTED REGION: (B)(6). DID THE EVENT INVOLVE A PATIENT: YES. AGE AND SEX OF PATIENT: MALE (B)(6) YEARS OLD. WHAT IS THE OUTCOME OF THE PATIENT: NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO IMPACT TO THE PATIENT/STABLE: THERE WAS NO REPORTED IMPACT TO THE PATIENT. IF PATIENT IMPACT, PLEASE PROVIDE DETAILS: ANY REPORTED ADVERSE PATIENT EFFECTS: NO ADVERSE PATIENT EFFECTS WERE REPORTED. ANY EXTENDED HOSPITALIZATION: NONE REPORTED. IT WAS REPORTED THE SHEATH WAS LEAKING, HOW WAS THE PRODUCT ISSUE RESOLVED: WAS ANOTHER M

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7301605 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122084. One submission often covers many models, so check the product and lot details in each record.

  • Z-2680-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
  • Z-2679-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
  • Z-2678-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
  • Z-1467-2021 · Feb 22, 2021 · Terminated
    AngioDynamics Safe Sheath Ultralite - Vascular Introducer/Catheter Introducer Catalogue #: VI10 - Product Usage: For the introduction of various types of pacing leads and catheter

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page