SafeSheath® II 667464-300Recalls & reported problems
Oscor Inc. · Hemostatic Tear-away Introducer System with Infusion Side Port, 9.5F · Class 2
All models of SafeSheath II (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20802526532522), received between Feb 27, 2018 and Feb 27, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Air Leak” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K122084); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Air Leak2
- Leak/Splash2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 27, 2018Air Leak; Leak/Splash; Air Leak; Leak/Splash
THE COMPLAINT WAS THAT THE INTRODUCERS THEY WERE USING DURING A COUPLE OF IMPLANT CASES APPEARED TO BE LEAKING. HERE IS A SUMMARY FROM THE COMPLAINT: THE PHYSICIAN COMPLAINT THAT THE SHEATHS WE USE SEEM TO BE LEAKING. AFTER GAINING ACCESS AND PUTTING A SHEATH THROUGH THE PUNCTURE SITE, WHEN ATTEMPTING TO FLUSH THE SHEATH, THERE ARE A LOT OF AIR BUBBLES THAT IS ASPIRATED AS HE DRAWS BACK THE SHEATH[?]..AS A RESULT, THERE WAS AIR INTRODUCED INTO THE PATIENT'S VESSELS AS HE ADVANCES THE LEAD INTO THE BODY. HE IS CONCERNED AS THIS PUTS PATIENT AT RISK OF EMBOLISM AND HYPOTENSION. THE FOLLOWING INFORMATION RECEIVED WAS RECEIVED ON FEBRUARY 8, 2018: EVENT DATE: (B)(6) 2017. REPORTED REGION: (B)(6). DID THE EVENT INVOLVE A PATIENT: YES. AGE AND SEX OF PATIENT: MALE (B)(6) YEARS OLD. WHAT IS THE OUTCOME OF THE PATIENT: NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO IMPACT TO THE PATIENT/STABLE: THERE WAS NO REPORTED IMPACT TO THE PATIENT. IF PATIENT IMPACT, PLEASE PROVIDE DETAILS: ANY REPORTED ADVERSE PATIENT EFFECTS: NO ADVERSE PATIENT EFFECTS WERE REPORTED. ANY EXTENDED HOSPITALIZATION: NONE REPORTED. IT WAS REPORTED THE SHEATH WAS LEAKING, HOW WAS THE PRODUCT ISSUE RESOLVED: WAS ANOTHER M
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7301640 ↗MalfunctionFeb 27, 2018Air Leak; Leak/Splash; Air Leak; Leak/Splash
THE COMPLAINT WAS THAT THE INTRODUCERS THEY WERE USING DURING A COUPLE OF IMPLANT CASES APPEARED TO BE LEAKING. HERE IS A SUMMARY FROM THE COMPLAINT: THE PHYSICIAN COMPLAINT THAT THE SHEATHS WE USE SEEM TO BE LEAKING. AFTER GAINING ACCESS AND PUTTING A SHEATH THROUGH THE PUNCTURE SITE, WHEN ATTEMPTING TO FLUSH THE SHEATH, THERE ARE A LOT OF AIR BUBBLES THAT IS ASPIRATED AS HE DRAWS BACK THE SHEATH[?]..AS A RESULT, THERE WAS AIR INTRODUCED INTO THE PATIENT'S VESSELS AS HE ADVANCES THE LEAD INTO THE BODY. HE IS CONCERNED AS THIS PUTS PATIENT AT RISK OF EMBOLISM AND HYPOTENSION. THE FOLLOWING INFORMATION RECEIVED WAS RECEIVED ON FEBRUARY 8, 2018: EVENT DATE: (B)(6) 2017. REPORTED REGION: (B)(6). DID THE EVENT INVOLVE A PATIENT: YES. AGE AND SEX OF PATIENT: MALE (B)(6) YEARS OLD. WHAT IS THE OUTCOME OF THE PATIENT: NO ADVERSE PATIENT EFFECTS WERE REPORTED. NO IMPACT TO THE PATIENT/STABLE: THERE WAS NO REPORTED IMPACT TO THE PATIENT. IF PATIENT IMPACT, PLEASE PROVIDE DETAILS: ANY REPORTED ADVERSE PATIENT EFFECTS: NO ADVERSE PATIENT EFFECTS WERE REPORTED. ANY EXTENDED HOSPITALIZATION: NONE REPORTED. IT WAS REPORTED THE SHEATH WAS LEAKING, HOW WAS THE PRODUCT ISSUE RESOLVED: WAS ANOTHER M
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7301605 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122084. One submission often covers many models, so check the product and lot details in each record.
- Z-2680-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
- Z-2679-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
- Z-2678-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
- Z-1467-2021 · Feb 22, 2021 · TerminatedAngioDynamics Safe Sheath Ultralite - Vascular Introducer/Catheter Introducer Catalogue #: VI10 - Product Usage: For the introduction of various types of pacing leads and catheter
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page