Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20845854083526), received between Jun 12, 2023 and Aug 28, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2024Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE DEVICE, KRR090, INFINITY RETRO-REAMER, 9MM, WAS BEING USED DURING AN ACL PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED, ¿DURING THE SURGERY, WHILE DRILLING, THE TIP OF THE REAMER BROKE AND THE BROKEN PIECE WAS FOUND INSIDE THE KNEE JOINT, WHICH WAS IDENTIFIED USING C-ARM AND REMOVED. THE PATIENT'S CONDITION IS GOOD.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS COMPLETED USING AN ALTERNATE SAME DEVICE WHICH CAUSED A 20-MINUTE DELAY. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20084551 ↗
  • MalfunctionJun 12, 2023Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE KRR090, INFINITY RETRO-REAMER, 9MM, WAS BEING USED DURING AN ACL RECONSTRUCTION PROCEDURE ON 31JAN23 WHEN IT WAS REPORTED, ¿THE CUTTING TIP IS BROKEN.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. FURTHER ASSESSMENT QUESTIONS FOUND THAT THE DEVICE BROKE, AND A FRAGMENT FELL INTO PATIENT. FRAGMENT WAS REMOVED USING A GRASPER OR KELLY STRAIGHT CLAMP. THE PROCEDURE WAS COMPLETED WITH A 10-DELAY USING AN ALTERNATE DEVICE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17105214 ↗

Other BIT, SURGICAL devices

All BIT, SURGICAL devices →