Mon-a-Therm 90041Recalls & reported problems
Covidien LP · Esophageal Stethoscope with Temperature Sensor 400TM · Class 2
All models of Mon-a-Therm (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20884521031712), received between Sep 13, 2022 and Sep 13, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “High Readings” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- High Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 13, 2022High Readings; High Readings
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE DID NOT MEET THE EXPECTED RANGE REQUIREMENTS AS STATED IN THE PRODUCT INSTRUCTION FOR USE (IFU). THE EXPECTED READINGS WERE 35 TO 42 DEGREES CELSIUS. HOWEVER, THE PROBE READ HIGH BY 0.9C. THE TEST EQUIPMENT CALIBRATIONS WERE UP TO DATE AND REPEATED FOR SEVERAL TIMES WITH THE SAME RESULTS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15412237 ↗
FDA clearance and classification
Cleared through 510(k) K251025 on Dec 29, 2025 (decision: Substantially Equivalent). Submitted by Covidien, LLC as “Mon-a-Therm Esophageal Stethoscope with Temperature Sensor 400TM (90041, 90042, 90043, 90049)”. The FDA review took 271 days. FDA record ↗
Cleared through 510(k) K811862 on Jul 10, 1981 (decision: Substantially Equivalent). Submitted by Mallinckrodt Critical Care as “Hi-Lo Temp Esophageal Stethoscope”. The FDA review took 9 days. FDA record ↗
The FDA regulates product code BZT (stethoscope, esophageal, with electrical conductors) under 21 CFR 868.1920 as a Class II device. 21 CFR 868.1920 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Mon-a-Therm 90041
Has the Mon-a-Therm 90041 been recalled?
No FDA recall record in our data names this exact device identifier (DI 20884521031712), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Mon-a-Therm 90041?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 13, 2022. That report lists device problem “High Readings” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Mon-a-Therm 90041?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Mon-a-Therm 90041 FDA approved?
It was cleared through 510(k) K811862 on Jul 10, 1981; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Mon-a-Therm 90041 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Mon-a-Therm 90041, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Mon-a-Therm, model 90041). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS devices
- Mon-a-Therm 90042Covidien LP · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 recall0 reports
- DeRoyal 81-040418Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS3 reports
- DeRoyal 81-040412Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS3 reports
- Medline DYNJAER4009Medline Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS2 reports
- DeRoyal 81-040409Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- Level 1 ES400-18Icu Medical, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- DeRoyal 81-050409Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- DeRoyal 81-050418Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page