Level 1 ES400-18Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Level 1 (52)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 30695085000321), received between May 12, 2017 and May 12, 2017 — classified as 1 injury.
The most frequently reported specific device problem is “Use of Device Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Use of Device Problem1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 12, 2017Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT A LEVEL 1® ESOPHAGEAL STETHOSCOPE TEMPERATURE SENSOR MIGRATED INTO THE PATIENT'S STOMACH. THE PATIENT WAS UNDERGOING A LAPAROSCOPIC GASTRIC SLEEVE SURGERY AT THE TIME OF THE EVENT. WHILE THE DOCTOR WAS STAPLING THE STOMACH (AS A PART OF THE GASTRIC SLEEVE SURGERY), THE DOCTOR STAPLED OVER THE ESOPHAGEAL STETHOSCOPE TEMPERATURE SENSOR; STAPLING THE DEVICE INTO THE PATIENT'S STOMACH. TO CORRECT THIS, THE PATIENT'S SURGERY HAD TO BE CONVERTED TO AN "OPEN" PROCEDURE, REQUIRING A FULL OPENING OF THE PATIENT'S ABDOMEN. THE EVENT ITSELF WAS DETERMINED TO BE AN OPERATOR ERROR. IT WAS REPORTED THAT THERE WERE NO FURTHER CLINICAL CONSEQUENCES TO THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 6563648 ↗
FDA clearance and classification
Cleared through 510(k) K854212 on Jan 3, 1986 (decision: Substantially Equivalent). Submitted by Respiratory Support Products, Inc. as “Acoustascope Esophageal Stethoscope W/Temp Sensor”. The FDA review took 78 days. FDA record ↗
The FDA regulates product code BZT (stethoscope, esophageal, with electrical conductors) under 21 CFR 868.1920 as a Class II device. 21 CFR 868.1920 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Level 1 ES400-18
Has the Level 1 ES400-18 been recalled?
No FDA recall record in our data names this exact device identifier (DI 30695085000321), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Level 1 ES400-18?
The FDA's MAUDE database lists 1 adverse event report for this device, received on May 12, 2017. That report lists device problem “Use of Device Problem” (1); patient problem “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Level 1 ES400-18?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Level 1 ES400-18 FDA approved?
It was cleared through 510(k) K854212 on Jan 3, 1986. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Level 1 ES400-18, and is it still sold?
The FDA's device database lists Icu Medical, Inc. as the company responsible for it (brand Level 1, model ES400-18). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS devices
- Mon-a-Therm 90042Covidien LP · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 recall0 reports
- DeRoyal 81-040418Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS3 reports
- DeRoyal 81-040412Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS3 reports
- Medline DYNJAER4009Medline Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS2 reports
- Mon-a-Therm 90041Covidien LP · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- DeRoyal 81-040409Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- DeRoyal 81-050409Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
- DeRoyal 81-050418Deroyal Industries, Inc. · STETHOSCOPE, ESOPHAGEAL, WITH ELECTRICAL CONDUCTORS1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page