Step S110000Recalls & reported problems
Covidien LP · Insufflation Needle · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 20884521090849), received between Nov 17, 2017 and Dec 25, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K961940); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
- Adverse Event Without Identified Device or Use Problem1
- Bent1
- Difficult to Insert1
- Noise, Audible1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient2
- Hematoma1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 25, 2025Break; Break
IT WAS REPORTED THAT PREOPERATIVELY OF LAPAROSCOPIC GASTROPEXY, THE JAWS OF THE TWO CLIP APPLIERS WOULD NOT OPEN ALL THE WAY AND WOULD NOT FIRE. ADDITIONALLY, THE CANNULA BROKE FOR BOTH THE PORTS. THE DEVICE WAS NOT USED ON THE PATIENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23892194 ↗MalfunctionDec 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE FOR ACUTE APPENDICITIS, WHILE INSERTING THE LAPAROSCOPIC PORTS, THE SHORT VERESS NEEDLE HAD BEEN REPLACED WITH A SUBSTITUTION NEEDLE THAT WAS LONGER THAN WHAT WAS NORMALLY USED. THE NEEDLE HIT THE PATIENT'S INNER ABDOMEN AND CAUSED AN INTRAMURAL HEMATOMA OF THE SIGMOID COLON. AFTER REPAIRING, THE COLON WAS INSPECTED DIRECTLY; THERE WAS NO FULL-THICKNESS INJURY. IT WAS NOTED THAT THE SURGICAL TIME WAS EXTENDED FOR 30 MINUTES.
Patient problems: Hematoma
MAUDE report 20919450 ↗MalfunctionDec 18, 2023Break; Break
ACCORDING TO THE REPORTER, PREOPERATIVELY, WHEN PACKAGED OPENED, NEEDLE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18346278 ↗MalfunctionJun 29, 2021Break; Break
ACCORDING TO THE REPORTER, THE CANNULA BROKE WHEN THE DEVICE WAS OPENED. THIS HAPPENED PRIOR TO PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12085405 ↗MalfunctionNov 13, 2019Difficult to Insert; Noise, Audible; Difficult to Insert; Noise, Audible
ACCORDING TO THE REPORTER DURING A PEDIATRIC LAPAROSCOPIC ABDOMINAL PROCEDURE, BOTH DEVICE MADE A STRANGE NOISE, AND THERE IS SOME PROBLEM INSERTING THE OBTURATOR INTO THE SLEEVE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9315318 ↗MalfunctionJul 23, 2019Break; Break
ACCORDING TO THE REPORTER, WHILE INSERTING THE TROCAR THROUGH THE ABDOMINAL WALL DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE SURGEON WOULD INSERT THE NEEDLE, BUT COULD NOT REMOVE IT BECAUSE IT WAS TOO TIGHT. IT WAS ALSO REPORTED THAT THE NEEDLE WAS TOO BIG FOR THE SHEATH. TWO OTHER SEPARATE SHEATHS WERE BROUGHT, BUT BOTH WERE RIPPED/BUCKLED IN WHEN IT WAS ATTEMPTING TO REMOVE THE NEEDLE. THE SURGEON PERSISTED WITH THE TROCARS TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8819741 ↗MalfunctionNov 17, 2017Bent; Break; Bent; Break
ACCORDING TO THE REPORTER: OCCURRED DURING A LAPAROSCOPIC GONADECTOMY PROCEDURE. UPON INSERTION INTO THE BODY, THE NEEDLE BENT. THE CANNULA BROKE. IN ORDER TO RESOLVE THE ISSUE AND COMPLETE THE CASE, THE SURGEON USED ANOTHER NEEDLE WITH THE SAME ANGLE, FORCE, AND TECHNIQUE AND IT WORKED FINE. THERE WAS NO PATIENT HARM. THE PATIENT OUTCOME IS ALIVE, NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7040506 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K961940. One submission often covers many models, so check the product and lot details in each record.
- Z-1387-2023 · Mar 1, 2023 · Open, ClassifiedStep Bladeless Trocars, 10 mm cannula and dilator with radially expandable sleeve and 5 - 10 mm reducer, Product Number S101010
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page