Skip to content

Shiley 8UN85HRecalls & reported problems

Covidien LP · Adult Flexible Tracheostomy Tube Cuffless Disposable Inner Cannula · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 20884521176277), received between Jan 18, 2023 and Oct 22, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (6 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2023Open, Classified

    A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

    Z-1357-2023 · Root cause: Component change control · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn6
  • Appropriate Device Problem Term/Code Not Available1
  • Connection Problem1
  • Gas/Air Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Hemoptysis1
  • Hemorrhage/Blood Loss/Bleeding1
  • Unintended Extubation1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356980 ↗
  • MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356930 ↗
  • MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356790 ↗
  • MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23356707 ↗
  • MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23339354 ↗
  • MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23339137 ↗
  • MalfunctionApr 16, 2024Gas/Air Leak; Gas/Air Leak

    ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY CUFF HAD A SLOW LEAK AND WAS BEING REFILLED WITH ABOUT 3 CC OF AIR ONCE PER DAY. A CALL WAS RECEIVED THAT THE VENTILATOR WAS SOUNDING A LOW AIR ALARM, WENT TO REFILL THE TRACHEOSTOMY WITH MORE AIR AND COULD NOT RE-INFLATE THE CUFF. THE TUBE HAS BEEN REPLACED WITH A NEW ONE.

    Patient problems: Unintended Extubation

    MAUDE report 19117559 ↗
  • InjuryJun 12, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    ACCORDING TO THE REPORTER, THE PATIENT WAS ON AIRWAY ANTIBIOTICS AND HAD A TUBE REPLACEMENT. AFTER THE REPLACEMENT, THE PATIENT WAS "COUGHING OUT A LITTLE BLOOD AND HAD BLEEDING".

    Patient problems: Hemoptysis; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17106025 ↗
  • MalfunctionJan 18, 2023Connection Problem; Connection Problem

    ACCORDING TO THE REPORTER, DURING USE, THE DEVICE PLUG WOULD NOT FIT IN THE TRACH. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16194184 ↗

Other Tube tracheostomy and tube cuff devices

All Tube tracheostomy and tube cuff devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.