Shiley 8UN85HRecalls & reported problems
Covidien LP · Adult Flexible Tracheostomy Tube Cuffless Disposable Inner Cannula · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 20884521176277), received between Jan 18, 2023 and Oct 22, 2025 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 24, 2023Open, Classified
A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.
Z-1357-2023 · Root cause: Component change control · FDA record ↗
Most reported problems
Device problems
- Material Split, Cut or Torn6
- Appropriate Device Problem Term/Code Not Available1
- Connection Problem1
- Gas/Air Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Hemoptysis1
- Hemorrhage/Blood Loss/Bleeding1
- Unintended Extubation1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356980 ↗MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356930 ↗MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356790 ↗MalfunctionOct 22, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356707 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23339354 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23339137 ↗MalfunctionApr 16, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, DURING USE, THE TRACHEOSTOMY CUFF HAD A SLOW LEAK AND WAS BEING REFILLED WITH ABOUT 3 CC OF AIR ONCE PER DAY. A CALL WAS RECEIVED THAT THE VENTILATOR WAS SOUNDING A LOW AIR ALARM, WENT TO REFILL THE TRACHEOSTOMY WITH MORE AIR AND COULD NOT RE-INFLATE THE CUFF. THE TUBE HAS BEEN REPLACED WITH A NEW ONE.
Patient problems: Unintended Extubation
MAUDE report 19117559 ↗InjuryJun 12, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, THE PATIENT WAS ON AIRWAY ANTIBIOTICS AND HAD A TUBE REPLACEMENT. AFTER THE REPLACEMENT, THE PATIENT WAS "COUGHING OUT A LITTLE BLOOD AND HAD BLEEDING".
Patient problems: Hemoptysis; Hemorrhage/Blood Loss/Bleeding
MAUDE report 17106025 ↗MalfunctionJan 18, 2023Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE PLUG WOULD NOT FIT IN THE TRACH. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16194184 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.