Shiley 4CN65HRecalls & reported problems
Covidien LP · Adult Flexible Tracheostomy Tube with TaperGuard Cuff Disposable Inner Cannula · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Shiley 4CN65H is an adult flexible tracheostomy tube with a TaperGuard cuff and disposable inner cannula. Reports mostly describe the cuff leaking or blowing holes, the tube becoming blocked or coming out unexpectedly, and loose connections. There are 31 reports, including 1 death and 15 injuries. One open recall from March 2023 covers a manufacturing error that left connectors undersized, so 15mm caps and circuit parts fit loosely and can lead to breathing failure or delayed treatment.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 31
- Deaths reported
- 1
- Injuries reported
- 15
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 20884521176307), received between Dec 15, 2017 and May 22, 2026 — classified as 1 death, 15 injuries, 15 malfunctions.
The most frequently reported specific device problem is “Inflation Problem” (7 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 24, 2023Open, Classified
A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.
Z-1357-2023 · Root cause: Component change control · FDA record ↗
Most reported problems
Device problems
- Inflation Problem7
- Use of Device Problem6
- Gas/Air Leak4
- Material Split, Cut or Torn4
- Connection Problem3
- Obstruction of Flow3
- Appropriate Device Problem Term/Code Not Available2
- Defective Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Unintended Extubation8
- Extubate6
- Dyspnea4
- Hypoventilation3
- Insufficient Information3
- Low Oxygen Saturation3
- Tachycardia2
Recent adverse event reports
The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.
InjuryMay 22, 2026Obstruction of Flow
ACCORDING TO THE REPORTER, THE PATIENT WAS USING A TRACHEOSTOMY TUBE. THE EMERGENCY MEDICAL TECHNICIANS WERE CALLED TO THE PATIENT'S HOME FOR "BREATHING PROBLEMS". "TRACH BLOCKAGE," "FOREIGN OBJECT IN TRACH," AND "TRACH NOT OPERATING PROPERLY" WERE NOTED. THE PATIENT WAS TRACHEOSTOMY-DEPENDENT FOR RESPIRATION.
Patient problems: Dyspnea
MAUDE report 25260257 ↗InjuryMay 22, 2026Obstruction of Flow
ACCORDING TO THE REPORTER, THE PATIENT WAS USING A TRACHEOSTOMY TUBE. THE EMERGENCY MEDICAL TECHNICIANS WERE CALLED TO THE PATIENT'S HOME FOR "BREATHING PROBLEMS". "TRACH BLOCKAGE," "FOREIGN OBJECT IN TRACH," AND "TRACH NOT OPERATING PROPERLY" WERE NOTED. THE PATIENT WAS TRACHEOSTOMY-DEPENDENT FOR RESPIRATION AND WAS HOSPITALIZED FOR BREATHING ISSUES RELATED TO THE TRACH BLOCKAGE.
Patient problems: Dyspnea
MAUDE report 25260169 ↗InjuryNov 12, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING USE, THE TRACH TUBE'S CUFF HAD BLOWN HOLES FROM A PATIENT RECEIVING COUGH ASSIST TREATMENT THREE TIMES A DAY. THERE WAS REPORTED COMPLICATION/HARM TO THE PATIENT.
Patient problems: Unspecified Tissue Injury
MAUDE report 23532362 ↗MalfunctionNov 12, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING USE, THE TRACH TUBE'S CUFF HAD BLOWN HOLES FROM A PATIENT RECEIVING COUGH ASSIST TREATMENT THREE TIMES A DAY. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 23532342 ↗MalfunctionNov 12, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING USE, THE TRACH TUBE'S CUFF HAD BLOWN HOLES FROM A PATIENT RECEIVING COUGH ASSIST TREATMENT THREE TIMES A DAY. THERE WAS NO REPORTED PATIENT HARM.
Patient problems: Insufficient Information
MAUDE report 23532278 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23339757 ↗MalfunctionOct 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT PRIOR TO USAGE, THE TRACH TUBE'S CUFF HAD ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23335591 ↗MalfunctionJul 10, 2025Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT THE 6.5 MM TRACHEOSTOMY CANNULA WAS ASSOCIATED WITH A CUFF THAT WAS NOT HOLDING PRESSURE, RESULTING IN CUFF LEAKS. THE EVENT OCCURRED IN A HOSPITAL SETTING. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22472124 ↗MalfunctionMay 16, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S BALLOON DID NOT INFLATE DURING USE. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19337202 ↗InjuryApr 16, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE WAS NOTED WITH LEAK AND LOW VOLUMES FROM THE PATIENT FOR A FEW SECONDS. AN EMERGENCY UPSIZE TRACH CHANGE WAS DONE. PATIENT WAS PLACED BACK ON MECHANICAL VENTILATOR WITH SAME SETTINGS AND VITAL SIGN WERE STABLE.
Patient problems: Hypoventilation; Unintended Extubation
MAUDE report 19119950 ↗MalfunctionApr 16, 2024Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE HAD FITMENT ERRORS TO THE FILTER DURING USE. A TAPE WAS USED TO FILL THE GAP BETWEEN THE TRACH AND FILTER ATTACHMENT TO AVOID LEAKS. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19119594 ↗MalfunctionApr 16, 2024Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE HAD FITMENT ERRORS TO THE FILTER DURING USE. A TAPE WAS USED TO FILL THE GAP BETWEEN THE TUBE AND THE FILTER. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19119354 ↗MalfunctionApr 3, 2024Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE'S CUFF "BURST/BLOWN" DURING PATIENT USE. THE TUBE WAS REMOVED AND REPLACED, NO RESPIRATORY DISTRESS WAS NOTED.
Patient problems: Unintended Extubation
MAUDE report 19032923 ↗InjuryApr 3, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, THE TRACHEOSTOMY TUBE HAD A LEAK AND LOW VOLUMES DURING PATIENT USE. THE DOCTOR PERFORMED AN EMERGENCY TRACH CHANGE WITH UPSIZED TRACH. PATIENT WAS PLACED BACK ON MECHANICAL VENTILATOR WITH SAME SETTINGS, VITAL SIGNS WERE STABLE.
Patient problems: Hypoventilation; Unintended Extubation
MAUDE report 19032521 ↗InjuryFeb 29, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, THE PATIENT WAS NOT MAINTAINING ADEQUATE TIDAL VOLUMES AND THE RESPIRATORY CARE PRACTITIONER PERFORMED AN EMERGENT TRACH TUBE CHANGE WITH THE SAME SIZE. THE PATIENT WAS MONITORED THROUGHOUT WITH NO COMPLICATIONS.
Patient problems: Low Oxygen Saturation; Unintended Extubation
MAUDE report 18808962 ↗DeathAug 13, 2023Obstruction of Flow; Obstruction of Flow
ACCORDING TO THE REPORTER, THE PATIENT WAS A HOME CARE PATIENT, THAT WAS USING AN ADULT TRACHEOSTOMY TUBE, AND WAS CONNECTED TO A VENTILATOR. THE REPORTER STATED THE PATIENT ¿WAS UNABLE TO BREATHE¿ AS AIR WAS UNABLE TO GET IN PAST THE TRACHEOSTOMY TUBE. THE VENTILATOR WAS CHECKED AND SHOWED NORMAL OPERATION. IT WAS ADDITIONALLY REPORTED THAT EMERGENCY RESPONSE (911) WAS CALLED, AND UPON ARRIVAL; MEDICATION, HEART PUMP AND MANUAL VENTILATION VIA AMBU BAG WERE APPLIED BY THE MEDICAL RESPONDERS WITH NEGATIVE RESULTS. THE PATIENT WAS TRANSPORTED TO THE HOSPITAL BUT DIED. THE REPORTER STATED THE MEDICAL EXAMINER¿S REPORT ATTRIBUTED THE DEATH TO CARDIAC ARREST THAT RESULTED FROM MUCOUS IN THE TRACH WHICH PREVENTED AIR FROM GETTING INTO THE PATIENT¿S SYSTEM.
Patient problems: Airway Obstruction; Cardiac Arrest; Dyspnea; Hypoventilation; Loss of consciousness; Respiratory Failure
MAUDE report 17538030 ↗InjuryMay 4, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER, THE PATIENT "PULLED THE CUFF DOWN" TO PLACE THE SPEAKING VALVE AND IT "REFLATED" WITH HEALTHCARE PROVIDERS AT THE BEDSIDE IN SHORT DURATION. THE PATIENT HAD AN UNKNOWN INJURY.
Patient problems: Respiratory Insufficiency; Tachycardia; Unintended Extubation; Unspecified Tissue Injury
MAUDE report 16868170 ↗InjuryApr 11, 2023Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE WAS PLACED ON THE PATIENT IN ACCORDANCE WITH HOSPITAL POLICY, AND THE RESPIRATORY CARE PROFESSIONAL (RCP) OBSERVED THAT THE PATIENT HAD INCREASED WORK OF BREATHING (WOB), TACHYCARDIA, AND DESATURATION. RCP ATTEMPTED TROUBLESHOOTING BY INFLATING THE PILOT BALLOON WITH A 10CC SYRINGE, BUT THE PILOT BALLOON WAS UNABLE TO MAINTAIN AIR PRESSURE AND WAS ACTIVELY DEFLATING. A REPLACEMENT OF THE DEVICE WAS PERFORMED AS AN EMERGENCY.
Patient problems: Dyspnea; Low Oxygen Saturation; Tachycardia; Unintended Extubation
MAUDE report 16713171 ↗MalfunctionMar 3, 2023Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, THE SPEAKING VALVE POPPED OFF REGULARLY WHEN SPEECH WAS INITIATED WITH THE NEW TRACHEA. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16479852 ↗MalfunctionFeb 6, 2023Labelling, Instructions for Use or Training Problem; Labelling, Instructions for Use or Training Problem
ACCORDING TO THE REPORTER, DURING USE, THE WRITING ON THE FLANGE WORE OFF WHEN CLEANED SO THE HOSPITAL WAS NOT ABLE TO DETERMINE WHAT SIZE DEVICE THE PATIENT HAD. THERE WAS A REQUIRED TUBE REPLACEMENT.
Patient problems: Unintended Extubation
MAUDE report 16309512 ↗MalfunctionJul 20, 2022Defective Component; Defective Component
ACCORDING TO THE REPORTER, THERE WAS AN ISSUE WITH CAPPING A PATIENT. THE CAP DID NOT STAY ON THE HUB. TROUBLESHOOTED WITH MULTIPLE CAPS. SPEAKING VALVE, BECAUSE IT'S TAPERED WORKED FINE. MANUAL RESUSCITATION BAG ALSO WORKED BUT WOULD NEED TO BE HELD IN PLACE. WHICH NOT TYPICALLY MUCH OF THE TIME. TRIED MULTIPLE CAPS. THEY ALL SPIN AROUND AND SLID OR DROPPED OFF.
Patient problems: Insufficient Information
MAUDE report 15065939 ↗MalfunctionMar 16, 2022Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE CUFF OF THE DEVICE WAS RUPTURED. THERE WAS NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13781206 ↗InjuryAug 27, 2021Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE PATIENT WENT BACK TO THE CLINIC AND HAD TO GO TO ER AS PATIENT WAS UNABLE TO GET OXYGEN. SATURATION WAS 93%-96% ON 10 LITERS. THE PATIENT WAS ADMITTED AND O2 WAS OKAY. SEEN BY THE PHYSICIAN AND WAS FOUND TO BE HYPERCAPNIC. SIZE 4 CUFFED WAS CHANGED TO SIZE 6, "6DIC" WHICH IMPROVED THE ISSUE WITH HIGH CO2 LEVELS. EVENTUALLY HAD AN EVENT WITH CHANGE TO A NEW VENT AND PASSED OUT AND WAS ADMITTED TO ICU. SHORTLY AFTERWARDS THE PATIENT WAS DISCHARGED FROM THE HOSPITAL TO HOME. THE PATIENT HAD A MEDICAL HISTORY OF ASTHMA AND SCOLIOSIS. REPORTED USING THE DCT MOSTLY WITH THE CUFF DEFLATED.
Patient problems: Low Oxygen Saturation; Unintended Extubation
MAUDE report 12380940 ↗MalfunctionJul 5, 2021Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, THE DEVICE HAD AN AIR LEAKED. THE PATIENT HAD A CYANOSIS AND NEEDED ADDITIONAL OXYGEN FROM 2 LITERS TO 3 LITERS WITH EXERTION. IT WAS INDICATED THAT THE PATIENT FELT MORE AIRFLOW AROUND THE FLANGE AND ALSO STATED THAT NOW BREATHING IN AIR, THE AIR FEELS COLD. PATIENT WAS A VENTILATOR DEPENDENT AND DID NOT USE A SPEAKING VALVE WITH MEDICAL HISTORY OF ASTHMA AND SCOLIOSIS. PATIENT DID NOT MONITOR CUFF PRESSURE BUT INFLATED USING A SYRINGE TO DESIRED COMFORT OR DESIRED EFFECT. THE PATIENT HAD A CONTINUED SYMPTOMS DESCRIBED ON THE ORIGINAL REPORT.
Patient problems: Cyanosis
MAUDE report 12119307 ↗InjuryJan 3, 2018Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, DURING USE OF THE DEVICE, IT HAD AN ISSUE WITH CUFF INFLATION/DEFLATION. THE CUFF WAS INFLATED TO 23 CMH2O AND WAS REDUCED TO 5-10 CMH20. THE TUBE WAS REMOVED AND REPLACED. THERE WAS NO PATIENT INJURY.
Patient problems: Extubate
MAUDE report 7160455 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.