Microstream Advance MVIIHHRecalls & reported problems
Oridion Medical 1987 Ltd. · Neonatal-Infant Intubated CO2 Filter Line; High Humidity · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20884521762012), received between Sep 7, 2022 and Feb 20, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Filtration Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
10 other recalls cover devices cleared under the same FDA submission (K181467); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Feb 17, 2025Open, Classified
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Z-1385-2025 · Root cause: Labeling design · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Filtration Problem1
Patient problems
- Cardiac Arrest1
- No Clinical Signs, Symptoms or Conditions1
- Unintended Extubation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
ACCORDING TO THE REPORTER PRIOR TO USE, THE LINE FITS TOO TIGHTLY WITH THE NEONATE OR PEDIATRIC DEVICE CLOSED SUCTION SYSTEM REFERENCE CODE 195, CAUSING RESPIRATORY THERAPISTS TO STRUGGLE WHEN TRYING TO REMOVE IT FROM THE SUCTION SYSTEM. IT WAS NOT LOT SPECIFIC. THE CUSTOMER HAD BEEN USING THE SAME DEVICE SAMPLING LINES FOR ABOUT A YEAR, AND THE ISSUE HAD BEEN CONSISTENT. IT WAS DIFFICULT TO REMOVE THE ADAPTER FROM THE ENDOTRACHEAL TUBE (ETT). THE ISSUE DID NOT OCCUR WHEN USING THE OTHER BRAND LINES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21420724 ↗InjurySep 7, 2022Filtration Problem; Filtration Problem
ACCORDING TO THE REPORTER, DURING USE, THE FILTER LINE WAS CLOGGING AFTER FIVE HOURS OF USE. THE INTUBATION WAS FAILED. THE SIDE STREAM DID NOT WORK AND PATIENT WAS INTUBATED WITHOUT IT. THE PROBLEM CONTINUED BUT NOT DUE TO STICKY MEDS. THE NEO SIDE STREAM NEVER PICKED UP ANYTHING. THE MONITOR SHOWED THAT IT WAS CONNECTED AND READY TO USE. HOWEVER, AFTER 10 BREATHS IT JUST WAS NOT PICKING ANYTHING UP. THE MD CONTINUED THE INTUBATION AS THE PATIENT WAS IN FULL CARDIAC ARREST.
Patient problems: Cardiac Arrest; Unintended Extubation
MAUDE report 15370247 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K181467. One submission often covers many models, so check the product and lot details in each record.
- Z-1722-2025 · Apr 7, 2025 · Open, ClassifiedMicrostream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity
- Z-1382-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Infant Intubated, PT00134354 - Microstream Advance Neonatal-Infant Intub
- Z-1380-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: OmniVentLine Set, 012808 - OmniVentLine Set Adult-Pediatric CO2 Sampling Line and Airway
- Z-1384-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Advance Intubated Adult-Pediatric, PT00156254 - Microstream Advance Adult-Ped
- Z-1377-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: FilterLine Sets & VitaLine Sets Adult/Pediatric, MQ04616 - MICROSTREAM FilterLine H Set L
- Z-1379-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Airway Adapter and CO2 Sampling Line, 01158 - Microstream Luer Adult-Pediatric Airway Ada
- Z-1376-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set I
- Z-1383-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Luer Intubated Adult Pediatric, PT00156355 - Microstream Luer Adult-Pediatric
- Z-1378-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant
- Z-1381-2025 · Feb 17, 2025 · Open, ClassifiedMicrostream Instructions for Use and Part Number used with - Product Description, REF: Microstream Adv. Adult-Pediatric Intubated, PT00134333 - Microstream Advance Adult-Pediat
Other ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE devices
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall2 deaths8,444 reports
- Alaris 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE1 recall607 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE485 reports
- BD Alaris™ ETCO2 Module, Model 8300Carefusion 303, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE471 reports
- NA GF-210RNihon Kohden Corporation · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE386 reports
- Trilogy IT2110X21BRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls30 reports
- Trilogy Trilogy Evo Ventilator, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls26 reports
- Trilogy Evo Trilogy Evo Ventilator with oxygen blending model, GermanyRespironics, Inc. · ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE5 recalls9 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.