Skip to content

Medi-Trace Cadence 22550A (UDI 20884527022301)Recalls & reported problems

Cardinal Health, Inc. · Adult Multi-Function Defibrillation Electrodes · Class 2

All models of Medi-Trace Cadence (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20884527022301), received between Jul 31, 2018 and Jun 3, 2021 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Defibrillation/Stimulation Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Defibrillation/Stimulation Problem2
  • Detachment of Device or Device Component1
  • Material Integrity Problem1
  • Sparking1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 3, 2021Sparking; Sparking

    THE CUSTOMER REPORTED THE DEFIB ELECTRODES WAS SPARKING DURING TREATMENT OF A PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11931395 ↗
  • MalfunctionJun 2, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER REPORTED THAT THE DEVICE FAILED TO DO SYNCHRONIZED CARDIO-VERSION BECAUSE THE DEFIBRILLATOR ELECTRODES DID NOT RECOGNIZE THE T WAVE ON THE EKG AND DID NOT SEND A SIGNAL TO SHOCK. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10112318 ↗
  • MalfunctionJun 2, 2020Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem

    THE CUSTOMER REPORTED THAT THE DEVICE FAILED TO DO SYNCHRONIZED CARDIO-VERSION BECAUSE THE DEFIBRILLATOR ELECTRODES DID NOT RECOGNIZE THE T WAVE ON THE EKG AND DID NOT SEND A SIGNAL TO SHOCK. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10112315 ↗
  • MalfunctionJul 31, 2018Detachment of Device or Device Component; Material Integrity Problem; Material Integrity Problem

    THE CUSTOMER REPORTED THE GEL IS DETACHED FROM THE ELECTRODE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7736844 ↗

FDA clearance and classification

The FDA regulates product code DRO (pacemaker, cardiac, external transcutaneous (non-invasive)) under 21 CFR 870.5550 as a Class II device. 21 CFR 870.5550 ↗

Devices of this type normally need 510(k) clearance before sale.

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Medi-Trace Cadence 22550A

Has the Medi-Trace Cadence 22550A been recalled?

No FDA recall record in our data names this exact device identifier (DI 20884527022301), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Medi-Trace Cadence 22550A?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Jul 31, 2018 and Jun 3, 2021. In those reports, the most common specific device problems are “Defibrillation/Stimulation Problem” (2), “Detachment of Device or Device Component” (1) and “Material Integrity Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Medi-Trace Cadence 22550A?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the Medi-Trace Cadence 22550A, and is it still sold?

The FDA's device database lists Cardinal Health, Inc. as the company responsible for it (brand Medi-Trace Cadence, model 22550A). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Pacemaker, Cardiac, External Transcutaneous (Non-Invasive) devices

All Pacemaker, Cardiac, External Transcutaneous (Non-Invasive) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page