Skip to content

Medi-Trace Cadence 22661RRecalls & reported problems

Cardinal Health, Inc. · Adult Multi-Function Defibrillation Electrodes, Radiotransparent · Class 2

All models of Medi-Trace Cadence (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20884527022370), received between Jun 11, 2021 and Jun 11, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE CUSTOMER REPORTED THAT THEY HAD AN INCIDENT WHERE A DEFIB PATCH WAS NOT STICKING TO THE PATIENT DURING THE PROCEDURE. THE CUSTOMER STATED THAT THE EP TEAM WAS ROLLING THE PATIENT BACK ON HIS BED WHEN THEY NOTICED THAT THE R2 PATCH WAS NOT STUCK TO THE PATIENT. THE PATIENT WAS FREE OF HAIR AND MOISTURE. ADDITIONAL INFORMATION RECEIVED ON 4-JUN-2021 STATED THAT THE PROCEDURE BEING PERFORMED WAS A EP ABLATION AND CARDIAC CATH. THE PADS HAD BEEN PLACED IN CASE FOR THE PROCEDURE, THEY DID NOT NEED TO SHOCK THE PATIENT. THE PROCEDURE WAS OVER WHEN THE ISSUE WAS DISCOVERED. THERE WAS NO DELAY IN TREATMENT AND NO INJURY OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11988044 ↗

Other Pacemaker, Cardiac, External Transcutaneous (Non-Invasive) devices

All Pacemaker, Cardiac, External Transcutaneous (Non-Invasive) devices →

Sources and updates