NA 02.111.500.10Recalls & reported problems
Synthes GmbH · 1.25MM PLATE REDUCTION WIRE THRDD TIP W/SM STOP/150MM · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20886982031676), received between Feb 25, 2021 and Feb 25, 2021 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 25, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, THREE DEPTH GAUGES WERE BROKEN IN THE PROCESS OF MEASURING, THE TIPS COMPLETELY BROKE OFF. A TOTAL OF SEVEN WIRES BROKE OFF THE OLIVE TIPS, THE CANNULATED SCREWDRIVER WAS BROKEN, AND TWO T8'S WERE STRIPPED. THE PARALLEL GUIDE AND COUNTERSINK HAD RUST RESIDUE ON THEM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: 2.0/2.4MM DEPTH GAUGE 50MM (PART NUMBER SD319.006, LOT UNKNOWN, QUANTITY 1) CANNULATED COUNTERSINK FOR 4.5MM CANNULATED SCREWS (PART NUMBER 310.85, LOT UNKNOWN, QUANTITY 1). ADJUSTABLE PARALLEL WIRE GUIDE (PART NUMBER 312.73, LOT UNKNOWN, QUANTITY 1) CANNULATED 4.0MM HEXAGONAL SCREWDRIVER (PART NUMBER 314.05, LOT UNKNOWN, QUANTITY 1) STARDRIVE SCREWDRIVER SHAFT T8 105MM (PART NUMBER 314.467, LOT UNKNOWN, QUANTITY 1) 1.25MM PLATE REDUCTION WIRE THRDD TIP W/SM STOP/150MM (PART NUMBER 02.111.500.10, LOT UNKNOWN, QUANTITY 1). 1.6MM COMPRESSION WIRE 15MM THREAD/150MM LENGTH (PART NUMBER 03.211.415, LOT UNKNOWN, QUANTITY 1). 2.0/2.4MM DEPTH GAUGE 50MM (PART NUMBER SD319.006, LOT UNKNOWN, QUANTITY 2). THIS REPORT INVOLVES ONE (1) 1.25MM PLATE R
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11378747 ↗MalfunctionFeb 25, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, DURING AN UNKNOWN PROCEDURE, THREE DEPTH GAUGES WERE BROKEN IN THE PROCESS OF MEASURING, THE TIPS COMPLETELY BROKE OFF. A TOTAL OF SEVEN WIRES BROKE OFF THE OLIVE TIPS, THE CANNULATED SCREWDRIVER WAS BROKEN, AND TWO T8'S WERE STRIPPED. THE PARALLEL GUIDE AND COUNTERSINK HAD RUST RESIDUE ON THEM. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. NO PATIENT CONSEQUENCE. CONCOMITANT DEVICES REPORTED: 2.0/2.4MM DEPTH GAUGE 50MM (PART NUMBER SD319.006, LOT UNKNOWN, QUANTITY 1) CANNULATED COUNTERSINK FOR 4.5MM CANNULATED SCREWS (PART NUMBER 310.85, LOT UNKNOWN, QUANTITY 1). ADJUSTABLE PARALLEL WIRE GUIDE (PART NUMBER 312.73, LOT UNKNOWN, QUANTITY 1) CANNULATED 4.0MM HEXAGONAL SCREWDRIVER (PART NUMBER 314.05, LOT UNKNOWN, QUANTITY 1) STARDRIVE SCREWDRIVER SHAFT T8 105MM (PART NUMBER 314.467, LOT UNKNOWN, QUANTITY 1) 1.25MM PLATE REDUCTION WIRE THRDD TIP W/LRG STOP/150MM (PART NUMBER 02.111.501.10, LOT UNKNOWN, QUANTITY 1). 1.6MM COMPRESSION WIRE 15MM THREAD/150MM LENGTH (PART NUMBER 03.211.415, LOT UNKNOWN, QUANTITY 1). 2.0/2.4MM DEPTH GAUGE 50MM (PART NUMBER SD319.006, LOT UNKNOWN, QUANTITY 2). THIS REPORT INVOLVES ONE (1) 1.25MM PLATE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11378536 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page