WECK IPN914937Recalls & reported problems
Teleflex Incorporated · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 24026704709940), received between Nov 18, 2021 and Feb 28, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
- Dent in Material1
- Detachment of Device or Device Component1
- Separation Problem1
Patient problems
- Hemorrhage/Blood Loss/Bleeding2
- Insufficient Information1
- Nerve Damage1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryFeb 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
REPORTED ISSUE: REPORTED ISSUE: MULTIPLE CARDIAC POST-OP BLEEDS AS A RESULT OF LIGATING CLIPS COMING OFF ARTERIAL BRANCHES. THE PATIENT'S CONDITION IS UNKNOWN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 16453143 ↗MalfunctionApr 1, 2022Break; Break
SURGEON REPORTED THAT BIG ISSUES THESE APPLIERS. THEY ARE USED VERY LITTLE AND STILL BREAK QUICKLY.
Patient problems: Insufficient Information
MAUDE report 13978042 ↗MalfunctionFeb 3, 2022Dent in Material; Dent in Material
THE PART OF THE PIVOT PIN WAS FOUND DENTED UPON OPENING THE PACKAGE. THEREFORE, A NEW UNIT WAS USED INSTEAD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13444671 ↗InjuryNov 18, 2021Separation Problem; Separation Problem
INCIDENT HAPPENED ON THE (B)(6) 2021 IN SURGERY ROOM. DURING A THYROID SURGERY THE DOCTOR HAD TO USE A BLUE TITANIUM CLIP REF (B)(4) 0 LOT 73J2000654. THE CLIP DID NOT HOLD ON THE THYROID ARTERY. THIS CLIP GOT STUCK ON ONE OF THE JAWS THE HORIZON MEDIUM CLIP PLIERS 237081 LOT 06K1757772. THAT INJURED THE ARTERY WHEN THE SURGEON REMOVED THE APPLIER CAUSING SIGNIFICANT BLEEDING. FACING THIS URGENCY SITUATION, THE SURGEON USED A SMALLER HEMOSTASIS FORCEPS CAT # 137081 + CLIP 001200 LOT# 73J2000649, WHICH CAUSED A LOSS OF SIGNAL FROM THE RECURRENT NERVE WHICH CONTROLS VOICE FUNCTION. THE SURGEON STATED THAT THE SAME INCIDENT HAPPENED WITH THE SMALLER APPLIER LIGATION FAILURE. CONSEQUENCE: PER-OPERATIVE BLEEDING AND POSSIBLE RECURRENT NERVE DAMAGE. THIS PROBLEM HAS OCCURRED SEVERAL TIMES WITHOUT CONSEQUENCE FOR PATIENTS. EACH TIME WE CHANGED THE APPLIER AND FOR THIS INCIDENT THE APPLIER HAD BEEN CHANGED 15 DAYS BEFORE. SINCE THIS INCIDENT SALES REPRESENTATIVE WENT ON SITE ON MONDAY 08 NOVEMBER AND CHANGED APPLIERS. THE PATIENT'S CONDITION IS FINE. THE RECURRENT NERVE IS NON-FUNCTIONAL POSTOPERATIVELY. THERE WERE NO UNDERLYING MEDICAL CONDITIONS. THE REPORTED CLIPS AND APPLIER WERE CONFI
Patient problems: Hemorrhage/Blood Loss/Bleeding; Nerve Damage
MAUDE report 12836657 ↗
Other Applier, surgical, clip devices
- LIGACLIP MCS20Ethicon Endo-Surgery, LLC · Applier, surgical, clip1 death337 reports
- Endo Clip III 176630BCovidien LP · Applier, surgical, clip312 reports
- LIGACLIP MCM20Ethicon Endo-Surgery, LLC · Applier, surgical, clip308 reports
- LIGACLIP MCM30Ethicon Endo-Surgery, LLC · Applier, surgical, clip1 death236 reports
- WECK IPN915189Teleflex Incorporated · Applier, surgical, clip221 reports
- LIGACLIP MSM20Ethicon Endo-Surgery, LLC · Applier, surgical, clip152 reports
- LIGACLIP LX107Ethicon Endo-Surgery, LLC · Applier, surgical, clip105 reports
- WECK IPN028280Teleflex Incorporated · Applier, surgical, clip93 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page