POLARCUP 75017231Recalls & reported problems
Smith & Nephew, Inc. · POLARCUP ANCHORING PEG (2 EACH) · Class 2 · Implant
All models of POLARCUP (61)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 27611996096570), received between Jun 25, 2021 and May 7, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Remove” (1 report).
1 other recall cover devices cleared under the same FDA submission (K070278, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Remove1
- Fitting Problem1
- Mechanical Jam1
- Physical Resistance/Sticking1
- Positioning Failure1
- Positioning Problem1
Patient problems
- Failure of Implant1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2026Positioning Failure; Difficult to Remove
IT WAS REPORTED THAT, DURING AN ACETABULAR REVISION SURGERY (UNKNOWN IF FROM S+N IMPLANTS), ON ATTEMPTING TO UNSCREW ONE (1) POLARCUP ANCHORING PEG 2 EACH OF THE PEG HOLE COVERS FROM ONE (1) POLARCUP SHELL TI-PLASMA/HA 51 NON-CEM, THE THREADS CROSS THREADED AND COULD NOT BE REMOVED. THE SURGEON ATTEMPTED TO SEAT A PEG IN THE SECOND, AVAILABLE PEG HOLE AFTER DRILLING THE HOLE AS PER THE TECHNIQUE, BUT THE PEG FAILED TO SEAT FLUSH WITH THE CUP AND HAD TO BE REMOVED. THE PEG HOLE WAS LEFT OPEN. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25112264 ↗InjuryJun 25, 2021Fitting Problem; Mechanical Jam; Positioning Problem; Physical Resistance/Sticking; Fitting Problem; Mechanical Jam; Positioning Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT A REVISION OF A 40MM PINNACLE CUP AND EXETER STEM WAS PERFORMED DUE TO INSTABILITY. A 45 MM POLAR CUP WITH SCREWS AND FLANGES WAS CHOSEN. AFTER TRIALING, IT WAS DECIDED PEGS WERE NECESSARY. FIRST PEG HOLE WAS REMOVED UNEVENTFULLY BUT THE SECOND BECAME STUCK AFTER INITIALLY TURNING. ATTEMPTS TO KEEP REMOVING ROUNDED THE HEX IN COVER. EVENTUALLY, HOLE COVER WAS REMOVED USING MIDAS REX CAUSING A DELAY OF 30 MINUTES OR MORE. CUP WAS IMPLANTED, PEG HOLES DRILLED AND PEGS IMPACTED. UNFORTUNATELY THE PEGS DIDN¿T IMPACT ALL THE WAY IN AND WERE LEFT PROUD. IT WAS DECIDED THIS WAS AN UNACCEPTABLE RISK TO THE POLY AND CUP WAS REMOVED. SURGERY WAS FINISHED USING A COMPETITOR'S DEVICE. NO PATIENT HARM REPORTED.
Patient problems: Failure of Implant
MAUDE report 12068003 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070278, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page