Skip to content

POLARCUP 75017231Recalls & reported problems

Smith & Nephew, Inc. · POLARCUP ANCHORING PEG (2 EACH) · Class 2 · Implant

All models of POLARCUP (61)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 27611996096570), received between Jun 25, 2021 and May 7, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Remove” (1 report).

1 other recall cover devices cleared under the same FDA submission (K070278, K211176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Remove1
  • Fitting Problem1
  • Mechanical Jam1
  • Physical Resistance/Sticking1
  • Positioning Failure1
  • Positioning Problem1

Patient problems

  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Positioning Failure; Difficult to Remove

    IT WAS REPORTED THAT, DURING AN ACETABULAR REVISION SURGERY (UNKNOWN IF FROM S+N IMPLANTS), ON ATTEMPTING TO UNSCREW ONE (1) POLARCUP ANCHORING PEG 2 EACH OF THE PEG HOLE COVERS FROM ONE (1) POLARCUP SHELL TI-PLASMA/HA 51 NON-CEM, THE THREADS CROSS THREADED AND COULD NOT BE REMOVED. THE SURGEON ATTEMPTED TO SEAT A PEG IN THE SECOND, AVAILABLE PEG HOLE AFTER DRILLING THE HOLE AS PER THE TECHNIQUE, BUT THE PEG FAILED TO SEAT FLUSH WITH THE CUP AND HAD TO BE REMOVED. THE PEG HOLE WAS LEFT OPEN. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25112264 ↗
  • InjuryJun 25, 2021Fitting Problem; Mechanical Jam; Positioning Problem; Physical Resistance/Sticking; Fitting Problem; Mechanical Jam; Positioning Problem; Physical Resistance/Sticking

    IT WAS REPORTED THAT A REVISION OF A 40MM PINNACLE CUP AND EXETER STEM WAS PERFORMED DUE TO INSTABILITY. A 45 MM POLAR CUP WITH SCREWS AND FLANGES WAS CHOSEN. AFTER TRIALING, IT WAS DECIDED PEGS WERE NECESSARY. FIRST PEG HOLE WAS REMOVED UNEVENTFULLY BUT THE SECOND BECAME STUCK AFTER INITIALLY TURNING. ATTEMPTS TO KEEP REMOVING ROUNDED THE HEX IN COVER. EVENTUALLY, HOLE COVER WAS REMOVED USING MIDAS REX CAUSING A DELAY OF 30 MINUTES OR MORE. CUP WAS IMPLANTED, PEG HOLES DRILLED AND PEGS IMPACTED. UNFORTUNATELY THE PEGS DIDN¿T IMPACT ALL THE WAY IN AND WERE LEFT PROUD. IT WAS DECIDED THIS WAS AN UNACCEPTABLE RISK TO THE POLY AND CUP WAS REMOVED. SURGERY WAS FINISHED USING A COMPETITOR'S DEVICE. NO PATIENT HARM REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12068003 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K070278, K211176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0326-2025 · Oct 15, 2024 · Open, Classified
    Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.

Other Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices

All Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented devices →

Sources and updates