Skip to content

BD SafetyGlide 23x1-1/2 RB TW 304387Recalls & reported problems

Becton, Dickinson And Company · NEEDLE SAFETYGLIDE 23X1-1/2 RB TW · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 30382903043874), received between Jun 6, 2023 and Sep 21, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

4 other recalls cover devices cleared under the same FDA submission (K951254); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment of Device or Device Component2
  • Activation Failure1
  • Activation Problem1
  • Connection Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Needle Stick/Puncture1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Leak/Splash

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50849056 ↗
  • MalfunctionAug 18, 2026Detachment of Device or Device Component

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50304541 ↗
  • MalfunctionMar 16, 2026Break

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603743 ↗
  • MalfunctionMar 16, 2026Break

    NO ADDITIONAL INFORMATION WAS PROVIDED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603738 ↗
  • MalfunctionMar 12, 2026Break

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24581454 ↗
  • MalfunctionAug 13, 2025Break; Break

    IT WAS REPORTED THAT BD NEEDLE SAFETYGLIDE 23X1-1/2 RB TW NEEDLE PULLED OUT OF THE HUB. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. DESCRIBE EVENT: MA GIVING B12 INJECTION NEEDLE CAME APART OF SYRINGE AND MEDICATION SPILLED OUT. ADDITIONAL INFORMATION PROVIDED: 1. KINDLY PROVIDE THE DATE OF EVENT. - 07/21/2025. 2. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NONE TO PROVIDER/STAFF OR PATIENT; PATIENT JUST DID NOT RECEIVE ALL OF THE B12 MEDICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22796216 ↗
  • MalfunctionJul 3, 2025Connection Problem; Connection Problem

    IT WAS REPORTED THAT THE BD NEEDLE SAFETYGLIDE 23X1-1/2 RB TW HAD A SYRINGE NEEDLE CONNECTIVITY ISSUE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL # 304387 BATCH # UNKNOWN. IT WAS REPORTED BY THE CUSTOMER THAT THE SAFETY NEEDLES HAVE A FLIMSY ACTUAL NEEDLE AND ARE HARD TO ACTIVATE THE SAFETY ON. VERBATIM: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED I DON'T KNOW IF ANYONE HAS TOLD YOU BUT*** IS VERY CONCERNED ABOUT OUR NEW "SAFETY" NEEDLES. THE ACTUAL NEEDLE IS FLIMSY BUT IT IS HARD TO ACTIVATE THE SAFETY ON IT. THE NURSES FEEL LIKE THEY ARE GOING TO STICK THEMSELVES BECAUSE THEIR FINGER IS TRYING TO SLIDE IT UP. WHAT HAPPENS IS THEIR FINGER PUSHES TO ENGAGE THE SAFETY PIECE AND IT SLIPS OFF OF IT AND GOES TO THE TOP OF THE NEEDLE. IT DID ACTUALLY HAPPEN IN . ADDITIONAL INFORMATION PROVIDED: I THINK WE FIGURED OUT THE ISSUE. IT WAS HAPPENING WITH NON-LURE LOCK SYRINGES. 1. WHEN TRYING TO ACTIVATE THE SAFETY USING THE TABLE TOP METHOD IT WOULD FLY OFF THE SYRINGE WITHOUT ENGAGING. 2. IF USING THE FINGER TO SLIDE IT UP IT IS HARD TO ACTIVATE WITH "WET" GLOVED HANDS IT IS HARD TO ACTIVATE AND THE FINGER WOULD SLIDE UP THE DEVICE WITHOUT ENGAGING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22398640 ↗
  • MalfunctionJul 3, 2025Activation Problem; Activation Problem

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Needle Stick/Puncture

    MAUDE report 22398626 ↗
  • MalfunctionSep 24, 2024Break; Break

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20294104 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20232786 ↗
  • MalfunctionAug 12, 2024Break; Break

    MATERIAL#: 304387 BATCH NUMBER#: 3012230. IT WAS REPORTED THAT THE BD NEEDLE SAFETYGLIDE 23X1-1/2 RB TW NEEDLE BROKE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RCC RECEIVED A COMPLAINT VIA EMAIL. EMAIL(S) ATTACHED SURGEON INSERTED HYPO NEEDLE INTO PATIENT'S CHEST TO INJECT LOCAL FOR SURGICAL INCISION. WHEN HE INSERTED THE HYPO, THE NEEDLE BROKE OFF WHILE USING IT ON THE PATIENT. NEEDLE AND OTHER PIECES WERE REMOVED AND NONE LEFT INSIDE PATIENT. (BF304387) LOT NUMBER - 3012230. PLEASE REVIEW THE CASE INFORMATION BELOW FOR A CUSTOMER REPORTED COMPLAINT FOR ONE OF YOUR PRODUCTS. THERE IS A SAMPLE AVAILABLE. COMPLAINT NUMBER, CREATED DATE, EVENT DATE, MANUFACTURING SITE, PRODUCT MALFUNCTION/FAILURE MODE. PRODUCT DESCRIPTION SUMMARY: COMPLAINT QUANTITY SAMPLE AVAILABLE RESOURCE COMPONENT # MFG. SKU NUMBER COMPONENT NAME INVESTIGATED COMPONENT LOT NUMBER WAS THERE AN INJURY WAS THERE A DEATH CASE (B)(4). ON (B)(6) 2024 BROKE SURGEON INSERTED HYPO NEEDLE INTO PATIENT'S CHEST TO INJECT LOCAL FOR SURGICAL INCISION. WHEN HE INSERTED THE HYPO, THE NEEDLE BROKE OFF WHILE USING IT ON THE PATIENT. NEEDLE AND OTHER PIECES WERE REMOVED AND NONE LEFT INSIDE PATIENT. (BF304387)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19960407 ↗
  • MalfunctionJun 6, 2023Activation Failure; Activation Failure

    IT WAS REPORTED THAT THE BD SAFETYGLIDE¿ NEEDLE SAFETY MECHANISM WAS DIFFICULT TO ENGAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MAY 3. DR. WAS PREFORMING AN INJECTION ON A HIP JOINT AND THE NEEDLE BENT AND WAS DIFFICULT TO WITHDRAW FROM THE PATIENT. WHEN TEH SAFETY GLIDE WAS ATTEMPTED TO BE APPLIED THE NEEDLE BENT FURTHER AND IT WAS DIFFICULT TO CAP - RISK TO STAFF FOR NEEDLE PRICK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17075413 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K951254. One submission often covers many models, so check the product and lot details in each record.

  • Z-2889-2020 · Jul 28, 2020 · Terminated
    BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack
  • Z-0315-2018 · Jul 5, 2017 · Terminated
    SafetyGlide Needle, 22GA 1-1/2in, Catalog n. 301701, manufactured by BD, sold in Centurion Convenience Kits. Centurion component code BD17O1NS.
  • Z-3116-2017 · Jun 13, 2017 · Terminated
    BD Needle 22GA 1-1/2in SafetyGlide(TM) Non Sterile, Bulk, REF 301701 The BD SafetyGlide(TM) needle is intended to be used for general purpose injection, aspiration of fluids from
  • Z-2748-2017 · Sep 16, 2016 · Terminated
    BD SafetyGlide" Injection Needle Catalog #305917 The SafetyGlide " needle is intended to be used for general purpose injection, aspiration of fluids from parts of the body below th

Other Needle, hypodermic, single lumen devices

All Needle, hypodermic, single lumen devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page