Infinity KBL191Recalls & reported problems

Conmed Corporation · Infinity Anteromedial Guide Left 9/10mm · Class 1

All models of Infinity (12)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30845854902305), received between Sep 1, 2020 and Jun 4, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation6

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE KBL191, INFINITY ANTEROMEDIAL GUIDE LEFT 9/10 MM WAS BEING USED ON (B)(6) 2026 DURING AN ACL RECONSTRUCTION AND ¿THE OFFSET GUIDE FROM THE LOANER SET SNAPPED IN HALF TODAY. 'YEAH THE SPADE PART POPPED RIGHT OFF WHILE DRILLING THE PIN' FROM THE SURGEON.¿ PER FURTHER ASSESSMENT IT WAS REPORTED "THE DEVICE BROKE IN THE JOINT WHILE DRILLING THE GUIDE PIN. THE BROKEN PARTS OF THE DEVICE WERE RETRIEVED." THERE WAS NO IMPACT OR INJURY TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25390266 ↗
  • MalfunctionOct 25, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE KBL191, INFINITY ANTEROMEDIAL GUIDE LEFT 9/10 MM WAS BEING USED ON (B)(6) 2024 DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION AND ¿TIP OF FEMORAL GUIDE BROKE OFF INSTRUMENT¿. PER FURTHER ASSESSMENT IT WAS FOUND "A FRAGMENT DID FALL INTO SURGICAL SITE. IT WAS RETRIEVED.". THERE WAS NO IMPACT OR INJURY FOR THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE UNKNOWN DEVICE. THERE WAS NO DELAY TO THE PROCEDURE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20535131 ↗
  • MalfunctionMar 8, 2024Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE KBL191, INFINITYTM ANTEROMEDIAL GUIDE L 9/10MM, WAS BEING USED DURING A LCA RECONSTRUCTION PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED, ¿WHEN RETRACTING THE INSTRUMENT FROM THE JOINT (AFTER PERFORMING FEMORAL DRILL WITH KGP035) IT DISASSEMBLED. THE DISTAL PART REMAINED IN ARTICULATION AND WAS REMOVED WITHOUT DIFFICULTY.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. FURTHER ASSESSMENT QUESTIONING FOUND THAT THE DEVICE FELL INTO THE PATIENT AND WAS RETRIEVED. THE PROCEDURE WAS COMPLETED AS PLANNED WITHOUT ANY DELAY. THE DEVICE, KGP035, 3.5MM SPADE TIP GUIDE PIN, WAS USED DURING THE PROCEDURE AS A CONCOMITANT DEVICE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18864960 ↗
  • MalfunctionJul 26, 2023Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, THAT THE KBL191, INFINITYTM ANTEROMEDIAL GUIDE L 9/10MM WAS BEING USED DURING AN ACL PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED ¿DR. WAS ABOUT TO DRILL THE 3.5MM SPADE TIP PIN (KGP035) WHILE USING THE 9/10MM FEMORAL OFFSET GUIDE (KBL191). AS HE WENT TO HOOK THE BACK OF THE FEMORAL CONDYLE, THE GUIDE TIP BROKE OFF INSIDE THE PATIENT. DR. WAS ABLE TO RETRIEVE ALL THE BROKEN PIECES AND THE PROCEDURE WAS FINISHED WITHOUT INJURY TO THE PATIENT BY DRILLING THE PIN FREEHAND.¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE. THERE WAS ALSO A 5-MINUTE DELAY REPORTED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION TO THE PATIENT. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17395578 ↗
  • MalfunctionApr 28, 2023Material Fragmentation; Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THE CUSTOMER THAT THE KBL191, INFINITYTM ANTEROMEDIAL GUIDE L 9/10MM, WAS BEING USED DURING A BTB PROCEDURE ON THE APPROXIMATE DATE OF (B)(6) 2023 WHEN IT WAS REPORTED, ¿DR. MITHRA WHILE PERFORMING BTB CASE, THE BELOW MENTIONED PARTICULAR INSTRUMENT HAS BROKEN IN THE MID OF THE SURGERY. THEN WITHOUT THIS INSTRUMENT ONLY, DOCTOR HAS COMPLETED THE SURGERY.¿. AFTER FURTHER ASSESSMENT IT WAS REPORTED, THAT THE DEVICE DID FRAGMENT AND ALL PIECES WERE RETRIEVED WITH GENERAL INSTRUMENTATION. THE PROCEDURE WAS COMPLETED AND THERE WAS A 20 MINUTE DELAY. THERE WAS NO REPORT OF INJURY TO THE PATIENT, PROLONGED HOSPITALIZATION, OR MEDICAL INTERVENTION. THE PATIENT WAS REPORTED AS "FINE". THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16835147 ↗
  • MalfunctionSep 1, 2020Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE KBL191 WAS BEING USED DURING A ACL RECONSTRUCTION ON 6AUG20 WHEN THE DEVICE WAS "INSERTED GUIDE PER RECOMMENDATIONS AND THE SHAFT SNAPPED IN HALF. FRAGMENT WAS LODGED IN JOINT BUT EASILY REMOVED WITH GRASPER. DELAY OF 1-MIN TO REMOVE FRAGMENT." THERE WAS NO REPORT OF INJURY TO THE PATIENT. THERE WAS NO MEDICAL INTERVENTION OR HOSPITALIZATION CAUSED BY THE EVENT. THE PATIENT WAS REPORTED AS NOW RECOVERING. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10479809 ↗

FDA clearance and classification

The FDA regulates product code LXH (orthopedic manual surgical instrument) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Infinity KBL191

Has the Infinity KBL191 been recalled?

No FDA recall record in our data names this exact device identifier (DI 30845854902305), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Infinity KBL191?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Sep 1, 2020 and Jun 4, 2026. In those reports, the most common specific device problem is “Material Fragmentation” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Infinity KBL191?

None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Infinity KBL191 FDA approved?

Devices of this type (product code LXH) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Conmed Corporation.

Who makes the Infinity KBL191, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand Infinity, model KBL191). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.