VersaOne UNVCA5STFRecalls & reported problems
Covidien LP · Universal Fixation Cannula · Class 2
All models of VersaOne (47)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is a Covidien VersaOne universal fixation cannula, a 5mm laparoscopic tube used to hold instruments and maintain access during keyhole surgery. The 76 reports mostly describe the cannula's membrane or seal splitting, cracking or detaching, leaking gas or fluid, and being hard to insert or remove; a few mention parts breaking off, and in two cases a fragment became a foreign body in the patient. Reports mention 4 injuries and no deaths, with most other cases noting no effect on the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 76
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 76 adverse event reports that list this exact device (DI 30884521215423), received between Mar 31, 2017 and Aug 6, 2026 — classified as 4 injuries, 72 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (23 reports).
6 other recalls cover devices cleared under the same FDA submission (K112349); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Split, Cut or Torn23
- Difficult to Insert17
- Difficult to Remove17
- Leak/Splash16
- Detachment of Device or Device Component14
- Break13
- Packaging Problem11
- Device Packaging Compromised4
Patient problems
- No Clinical Signs, Symptoms or Conditions33
- No Known Impact Or Consequence To Patient22
- No Consequences Or Impact To Patient12
- No Patient Involvement4
- Foreign Body In Patient2
- No Information2
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 76 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Leak/Splash; Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50151353 ↗MalfunctionFeb 6, 2026Break
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, AT THE BEGINNING OF THE PROCEDURE, THE CANNULA WAS FOUND TO HAVE A RUPTURED MEMBRANE. THE DEVICE WAS URGENTLY REPLACED BY ANOTHER PRODUCT TO CONTINUE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24284532 ↗MalfunctionOct 27, 2025Contamination
CRUMBLY SUBSTANCE INSIDE PACKAGING. PATIENT NOT AFFECTED, CAUGHT BEFORE USE ON A PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23389933 ↗MalfunctionMay 26, 2025Break; Material Twisted/Bent; Break; Material Twisted/Bent
IT WAS REPORTED THAT THE UNVCA5STF UNIVERSAL 5MM STD FIX CANNULA WAS PULLED IN PREPARATION FOR A CASE BUT WAS NOT USED BECAUSE THE CANNULA WAS SMASHED AND BENT. ANOTHER CANNULA WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22083697 ↗MalfunctionJan 8, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A LAPAROSCOPIC FUNDOPLICATION PROCEDURE, A SEAL WAS DAMAGED ON THE CANNULA AT THE BEGINNING OF THE SURGERY. THE TECHNICIAN PROVIDED ANOTHER CANNULA TO RESOLVE THE ISSUE AND COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21103250 ↗MalfunctionDec 6, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC SLEEVE, AT THE START OF THE SURGERY, THE CARBON DIOXIDE (CO2) BEGAN TO LEAK. WHEN THE DEVICE WAS OPENED, THE LUER-LOCK VALVE WAS TAPPED. THE DEVICE WAS REPLACED WITH NEW 5MM CANNULA TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20865689 ↗MalfunctionSep 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC SLEEVE, THE SEAL OF THE THREE DEVICES DISENGAGED. A NEW DEVICE OF THE SAME KIND WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20249923 ↗MalfunctionJun 28, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC DIAGNOSIS, WHEN ENTERING THE FORCEPS AT THE BEGINNING OF THE SURGERY TO MAKE THE DI AGNOSIS, THE TROCAR WAS PLACED ON THE ABDOMINAL WALL, THE SEAL DAMAGED AND WAS LEAKING. ANOTHER TROCAR WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19635360 ↗MalfunctionMay 17, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A LAPAROSCOPIC GASTRIC BYPASS, WHEN INSERTING THE TROCARS THROUGH THE ABDOMINAL WALL, THE SEALS OF THREE TROCARS WERE BROKEN. CARBON DIOXIDE (CO2) LEAKED. ALL THREE TROCARS WERE REPLACED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19343688 ↗MalfunctionMar 19, 2024
PERFORMING LAPAROSCOPIC HYSTERECTOMY WITH INSUFFLATION AND PRESSURE COULD NOT BE MAINTAINED. STRYKER REPS WERE CALLED FOR TROUBLESHOOTING. PROBLEM DID NOT RESOLVE. INSUFFLATION TUBING WAS REPLACED AND PROBLEM WITH MAINTAINING PRESSURE PERSISTED. THE STRYKER REPS SWAPPED OUT INSUFFLATION UNIT. TROCAR BEING USED WAS THE SUSPECTED ISSUE. TROCAR REMOVED FROM THE FIELD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18928954 ↗MalfunctionFeb 23, 2024Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC NEPHRECTOMY, WHEN INSERTING A NEEDLE, THE SEAL OF THE CANNULA RUPTURED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18771073 ↗MalfunctionFeb 22, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, WHEN A GRASPER CLAMP WAS INSERTED MIDWAY THROUGH THE PORT, THE MEMBRANE OR SEAL RUPTURED, RESULTING IN GAS LEAKAGE. THE DEVICE WAS REPLACED BY THE SAME LOT TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18759616 ↗MalfunctionFeb 7, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE CANNULA MEMBRANE RUPTURED, PREVENTING IT FROM CREATING A CAVITY AND ALLOWING CARBON DIOXIDE (CO2) TO ESCAPE. TO RESOLVE THE ISSUE, ANOTHER DEVICE WAS USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18663605 ↗MalfunctionFeb 7, 2024Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, BEFORE USING ON THE PATIENT, THE MEMBRANE WAS BROKEN AND LEAKING. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18661736 ↗MalfunctionJan 30, 2024Leak/Splash; Detachment of Device or Device Component; Insufficient Device Problem Information; Material Split, Cut or Torn; Leak/Splash; Detachment of Device or Device Component; Insufficient Device Problem Information; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A LAPAROSCOPIC GASTRIC BYPASS, THREE TROCARS HAVE FAILED TO FILTER CO2. THE SEALS WERE DAMAGED AND DISENGAGED. THE DEVICES LEAK AIR OR GAS. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18604029 ↗MalfunctionDec 18, 2023Detachment of Device or Device Component; Material Split, Cut or Torn; Detachment of Device or Device Component; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC BYPASS PROCEDURE, WHEN INTRODUCING A LAPAROSCOPIC FORCEPS, THE 5MM THE MEMBRANE OF THE CANNULA RUPTURED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18351434 ↗MalfunctionDec 15, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC BYPASS PROCEDURE, WHEN INTRODUCING A LAPAROSCOPIC FORCEPS, THE TWO 12MM CANNULA MEMBRANE (SEAL) RUPTURED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18338562 ↗MalfunctionDec 14, 2023Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, AT THE BEGINNING OF THE PROCEDURE, THE CANNULA MEMBRANE RUPTURED AND AN AIR LEAK. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18327598 ↗MalfunctionDec 14, 2023Leak/Splash; Material Split, Cut or Torn; Leak/Splash; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, MONOPOLAR HOOK WAS INTRODUCED THROUGH THE CANNULA AND THE MEMBRANE OF THE CANNULA BY THE PHYSICIAN AT THE START OF THE PROCEDURE, BUT THE CANNULA MEMBRANE RUPTURED AND AIR OR GAS LEAKED FROM THE SEAL. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18327334 ↗MalfunctionDec 12, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC ENDOMETRIOSIS PROCEDURE, THE MEMBRANE WAS RUPTURED WHEN A CLAMP WAS INSERTED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18314608 ↗MalfunctionNov 21, 2023Break; Break
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC BYPASS, WHEN THE CLAMP WAS INSERTED, THE CANNULA MEMBRANE RUPTURED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187793 ↗MalfunctionNov 20, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC BYPASS PROCEDURE, WHEN A LAPAROSCOPIC CLAMP WAS INSERTED, THE CANULA MEMBRANE (SEAL) RUPTURED. A NEW PRODUCT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18174410 ↗MalfunctionNov 20, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC BYPASS PROCEDURE, WHEN A CLAMP WAS INSERTED, THE MEMBRANE RUPTURED (SEAL DAMAGED). ANOTHER PRODUCT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18174362 ↗MalfunctionNov 20, 2023Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC BYPASS, WHEN INTRODUCING THE FORCEPS, THE CANNULA RUPTURED. A NEW DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18171251 ↗MalfunctionNov 20, 2023Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC SLEEVE, WHEN A CLAMP WAS INSERTED, THE TROCAR SEAL WAS DAMAGED. THE SURGEON USED ANOTHER DEVICE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18170445 ↗
FDA clearance and classification
Cleared through Special 510(k) K112349 on Sep 1, 2011 (decision: Substantially Equivalent). Submitted by Covidien, Formerly US Surgical A Divison of Tyco H as “Versaport V2 Bladeless Optiical Trocar And/Or Versaport Bladeless Optical Trocar”. The FDA review took 16 days. FDA record ↗
The FDA regulates product code GCJ (laparoscope, general & plastic surgery) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112349. One submission often covers many models, so check the product and lot details in each record.
- Z-1471-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Fixation Cannula - 5mm Short Product Code: ONBFCA5SH The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic, ge
- Z-1469-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Product Code: ONB5STF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic,
- Z-1470-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Trocar With (2) Fixation Cannulae - 5mm Product Code: ONB5STF2C The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gyne
- Z-1472-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Fixation Cannula - 5mm Product Code: ONBFCA5ST The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynecologic, general,
- Z-1467-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Long Product Code: ONB5LGF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gyneco
- Z-1468-2013 · May 6, 2013 · TerminatedVersaport Bladeless Optical Trocar With Fixation Cannula - 5mm Short Product Code: ONB5SHF The Versaport" Bladeless Optical 5mm trocars are intended for use in a variety of gynec
Common questions about the VersaOne UNVCA5STF
Has the VersaOne UNVCA5STF been recalled?
No FDA recall record in our data names this exact device identifier (DI 30884521215423), based on the FDA recall file dated Nov 27, 2026. However, 6 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the VersaOne UNVCA5STF?
The FDA's MAUDE database lists 76 adverse event reports for this device, received between Mar 31, 2017 and Aug 6, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (23), “Difficult to Insert” (17) and “Difficult to Remove” (17); the most common patient problems are “Foreign Body In Patient” (2) and “Device Embedded In Tissue or Plaque” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaOne UNVCA5STF?
None of the reports are classified as deaths, 4 as injuries and 72 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VersaOne UNVCA5STF FDA approved?
It was cleared through Special 510(k) K112349 on Sep 1, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the VersaOne UNVCA5STF, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand VersaOne, model UNVCA5STF). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Laparoscope, General & Plastic Surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,913 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,857 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,068 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall945 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery850 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery770 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall694 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery569 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.