Mallinckrodt 125035Recalls & reported problems
Covidien LP · Endobronchial Suction Catheters with Color Coded Connectors; Endobronchial Tube; Endobronchial Tube Accessories,Left,(2) Right Angle Swivel Connectors; Carlens Y Adapter · Class 2
All models of Mallinckrodt (73)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Covidien endobronchial tube, a breathing tube used in the airway during anesthesia or ventilation. Most of the 34 reports describe the tube's cuff failing to inflate or deflate properly, sometimes leaving it unable to hold air, plus a few cases of partial blockage or air leaking. Reports mention 3 injuries and no deaths. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 30884522008734), received between Mar 24, 2017 and Oct 21, 2019 — classified as 3 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Deflation Problem25
- Inflation Problem3
- Use of Device Problem2
- Gas/Air Leak1
- Insufficient Device Problem Information1
- Mechanical Problem1
- Partial Blockage1
Patient problems
- No Patient Involvement29
- Extubate3
- No Known Impact Or Consequence To Patient2
- No Information1
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE UNIT BALLOON WAS MISSED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9216471 ↗MalfunctionOct 14, 2019Mechanical Problem; Mechanical Problem
ACCORDING TO THE REPORTER, THE UNIT HAD POOR CUFF AND ONLY ONE SIDE BULGING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9189337 ↗MalfunctionOct 14, 2019Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE HAD NO TRACHEAL BALLOONING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9186922 ↗MalfunctionOct 14, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE UNIT AIR DID NOT LEFT OUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9186079 ↗MalfunctionJun 28, 2019Inflation Problem; Inflation Problem
ACCORDING TO THE REPORTER, THE DEVICE CUFF WOULD NOT INFLATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8742880 ↗MalfunctionApr 5, 2018Partial Blockage; Partial Blockage
ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE WAS NOT DEFLATED PROPERLY DURING THE CUFF TEST. IT REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7401267 ↗MalfunctionFeb 23, 2018Deflation Problem; Deflation Problem
ACCORDING TO THE REPORTER, PRIOR TO USE AN A PATIENT, THE UNIT'S CUFF COULD NOT BE DEFLATED. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7292073 ↗MalfunctionFeb 15, 2018Deflation Problem; Deflation Problem
ACCORDING TO THE REPORTER, DURING CUFF CHECK PRIOR TO USE ON A PATIENT, DEFLATION FAILURE WAS CONFIRMED. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7273567 ↗MalfunctionDec 7, 2017Deflation Problem; Deflation Problem
MANUFACTURER RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE TRACHEAL CUFF WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7097007 ↗MalfunctionNov 22, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT PRIOR TO USE, DURING INSPECTION, THE CUFF OF AN ENDOBRONCHIAL TUBE COULD NOT BE DEFLATED COMPLETELY. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. THE SAMPLE IS EXPECTED TO BE RETURNED.
Patient problems: No Patient Involvement
MAUDE report 7054103 ↗MalfunctionNov 22, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION REGARDING A CUSTOMER THAT COULD TAKE OUT THE AIR FROM THE BALLOON. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7053518 ↗MalfunctionNov 13, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION REGARDING AN ENDOBRONCHIAL TUBE IN WHICH THE TRACHEAL CUFF WOULD NOT DEFLATE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7026470 ↗MalfunctionNov 13, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION REGARDING AN ENDOBRONCHIAL TUBE IN WHICH THE TRACHEAL CUFF WOULD NOT DEFLATE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 7026463 ↗MalfunctionNov 13, 2017Deflation Problem; Deflation Problem
RECEIVED A COMMUNICATION REGARDING AN ENDOBRONCHIAL TUBE IN WHICH THE TRACHEAL CUFF WOULD NOT DEFLATE. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT. RECEIVED NOTIFICATION OF A TRACHEAL CUFF THAT DID NOT DEFLATE COMPLETELY DURING CUFF TESTING. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 7026455 ↗MalfunctionOct 5, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT DURING THE PRE-TEST THE CUFF WOULD NOT DEFLATE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 6915773 ↗MalfunctionOct 5, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT DURING THE PRETEST THE CUFF COULD NO BE DEFLATED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 6915576 ↗MalfunctionOct 5, 2017Inflation Problem; Inflation Problem
MEDTRONIC RECIEVED A REPORT THAT DURING THE CUFF TEST, AIR LEAKED FROM THE AIR INLET.
Patient problems: No Patient Involvement
MAUDE report 6914908 ↗MalfunctionOct 4, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT DURING THE PRE-TEST THE TUBE'S CUFF WOULD NOT DEFLATE. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 6914474 ↗InjurySep 13, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THE CUFF ON THE TRACHEOSTOMY TUBE WOULD NOT DEFLATE. THE TUBE WAS REMOVED AND REPLACED. THE CUSTOMER INDICATED THERE WAS NO PATIENT INJURY WITH THIS EVENT.
Patient problems: Extubate
MAUDE report 6862523 ↗MalfunctionSep 12, 2017Inflation Problem; Inflation Problem
MEDTRONIC RECEIVED A REPORT THAT DURING A CUFF CHECK A LEAK WAS FOUND. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 6859026 ↗MalfunctionAug 24, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON DID NOT DEFLATE COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6819387 ↗MalfunctionAug 24, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON DID NOT DEFLATE COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6819383 ↗MalfunctionJul 26, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT WHILE IN USE ON A PATIENT, AFTER 3 HOURS, THE CUFF WOULD NOT DEFLATE. THE CUFF WAS PRE-TESTED PRIOR TO USE. RECANNULATION OF A REPLACEMENT TUBE WAS NOT REQUIRED. THERE WAS NO PATIENT HARM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6745998 ↗InjuryJul 21, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED A COMMUNICATION THAT THE BLUE TUBE (BRONCHIAL) AND THE TRANSPARENT TUBE (TRACHEAL) WERE FOUND ADHERED ALTERNATIVELY.
Patient problems: Extubate
MAUDE report 6734829 ↗MalfunctionJul 17, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT PRIOR TO USE, DURING INSPECTION, THE CUFF OF AN ENDOBRONCHIAL TUBE COULD NOT BE DEFLATED COMPLETELY. THE CUSTOMER REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. THE SAMPLE IS EXPECTED TO BE RETURNED FOR EVALUATION.
Patient problems: No Patient Involvement
MAUDE report 6717263 ↗
Other TUBE, BRONCHIAL (W/WO CONNECTOR) devices
- Shiley 125037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)1 death209 reports
- Shiley 125035Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)114 reports
- Shiley 126035Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)50 reports
- Mallinckrodt 125037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)49 reports
- Mallinckrodt 126037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)47 reports
- Shiley 125039Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)40 reports
- Shiley 126039Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)39 reports
- Shiley 125032Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)34 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page