Mallinckrodt 125039Recalls & reported problems
Covidien LP · Endobronchial Suction Catheters with Color Coded Connectors; Endobronchial Tube; Endobronchial Tube Accessories,Left,(2) Right Angle Swivel Connectors; Carlens Y Adapter · Class 2
All models of Mallinckrodt (73)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 30884522008758), received between May 30, 2017 and Jan 18, 2018 — classified as 1 injury, 15 malfunctions.
The most frequently reported specific device problem is “Deflation Problem” (14 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Deflation Problem14
- Human-Device Interface Problem1
- Inflation Problem1
Patient problems
- No Patient Involvement15
- Extubate1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 18, 2018Deflation Problem; Deflation Problem
ACCORDING TO THE REPORTER, DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WOULD NOT DEFLATE COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7203738 ↗InjuryOct 11, 2017Human-Device Interface Problem; Human-Device Interface Problem
MEDTRONIC RECEIVED A COMMUNICATION REGARDING A BRONCHOSCOPE GETTING STUCK IN BRONCHO-CATH TUBE. IT WAS ALLEGED THAT THE BRONCHO-CATH TUBE MALFUNCTIONED AND CUSTOMER HAD TO REPLACE THE PATIENT TUBE. IT WAS REPORTED THAT THEY USED A NON-MEDTRONIC BRONCHOSCOPE. IT WAS INDICATED THAT THERE WAS NO FURTHER PATIENT INJURY.
Patient problems: Extubate
MAUDE report 6942718 ↗MalfunctionOct 5, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT DURING THE PRE-TEST THE CUFF COULD NOT BE DEFLATED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 6915776 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827891 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827860 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827771 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827692 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827561 ↗MalfunctionAug 29, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A COMMUNICATION THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THE CUSTOMER INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED TO THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6827535 ↗MalfunctionJul 26, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT THE CUFF ON THE ENDOBRONCHIAL TUBE INFLATED, BUT WOULD NOT DEFLATE. THE CUFF NEEDED TO BE SQUEEZED TO DEFLATE. THE CUFF WAS TESTED PRIOR TO INTUBATION. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6746036 ↗MalfunctionJul 26, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THAT THE CUFF ON THE ENDOBRONCHIAL TUBE INFLATED, BUT WOULD NOT DEFLATE. THE CUFF NEEDED TO BE SQUEEZED TO DEFLATE. THE CUFF WAS TESTED PRIOR TO INTUBATION. THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6746031 ↗MalfunctionJun 28, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT WHERE IT WAS ALLEGED THAT DURING THE CUFF TEST, THE AIR FROM THE TRACHEAL BALLOON WAS NOT DEFLATED COMPLETELY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6675523 ↗MalfunctionJun 23, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THE CUFF ON THE TRACHEOSTOMY TUBE WOULD NOT DEFLATE. CUSTOMER INDICATED THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6665503 ↗MalfunctionJun 23, 2017Inflation Problem; Inflation Problem
MEDTRONIC RECEIVED A REPORT THAT THE ENDOBRONCHIAL TUBE'S TRACHEAL CUFF DIDN'T INFLATE COMPLETELY. THIS WAS IDENTIFIED DURING PRE-TESTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 6664019 ↗MalfunctionMay 30, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THE CUFF ON THE ENDOBRONCHIAL TUBE WOULD NOT DEFLATE. CUSTOMER INDICATED THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6598330 ↗MalfunctionMay 30, 2017Deflation Problem; Deflation Problem
MEDTRONIC RECEIVED A REPORT THE CUFF ON THE ENDOBRONCHIAL TUBE WOULD NOT DEFLATE. CUSTOMER INDICATED THERE WAS NO PATIENT INVOLVEMENT WITH THIS EVENT.
Patient problems: No Patient Involvement
MAUDE report 6598292 ↗
Other TUBE, BRONCHIAL (W/WO CONNECTOR) devices
- Shiley 125037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)1 death209 reports
- Shiley 125035Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)114 reports
- Shiley 126035Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)50 reports
- Mallinckrodt 125037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)49 reports
- Mallinckrodt 126037Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)47 reports
- Shiley 125039Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)40 reports
- Shiley 126039Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)39 reports
- Mallinckrodt 125035Covidien LP · TUBE, BRONCHIAL (W/WO CONNECTOR)34 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page