Skip to content

SafeSheath® II SSL9Recalls & reported problems

Oscor Inc. · Hemostatic Tear-away Introducer System with Infusion Side Port, 9F · Class 2

All models of SafeSheath II (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30891492002358), received between May 24, 2023 and Oct 20, 2025 — classified as 1 death, 1 injury.

4 other recalls cover devices cleared under the same FDA submission (K122084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • Cardiac Arrest1
  • Low Blood Pressure/ Hypotension1
  • Perforation of Vessels1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE PHYSICIAN INADVERTENTLY PERFORATED THE LEAD THROUGH THE PATIENT'S HEART. THE PATIENT'S BLOOD PRESSURE DROPPED SIGNIFICANTLY AFTER THE PERFORATION. CARDIOPULMONARY RESUSCITATION WAS PERFORMED, AND A HOSPITAL CODE BLUE' WAS INITIATED. THE PATIENT WAS SENT TO THE OPERATING ROOM FOR STERNOTOMY IN ORDER TO CLOSE THE PERFORATION. THE LEAD WAS NOT IMPLANTED. PROCEDURE ABORTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. PRODUCT NOT RETURNING. MEDTRONIC CUSTOMER REPORTED THERE IS NO COMPLAINT CONCERNING THE OSCOR PRODUCT ON THIS EVENT. NON-COMPLAINT JUSTIFICATION AND RATIONALE: THERE WAS NO ALLEGATION AGAINST THE OSCOR INTRODUCER. THE MDT REP STATED THAT THE RV LEAD HAD PERFORATED THE PATIENT'S HEART TISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. NOT REPORTABLE; INFORMATION/ANALYSIS DOES NOT INDICATE A PRODUCT MALFUNCTION. THERE IS NO INDICATION OF A FAILURE OF THIS MARKETED DEVICE TO MEET CUSTOMER OR USER EXPECTATIONS FOR QUALITY OR TO MEET PERFORMANCE SPECIFICATIONS; THUS, THE ASSOCIATED LINE-ITEM TASKS WILL BE PROCESSED AS A NON-COMPLAINT. (B)(6). (B)(6).

    Patient problems: Cardiac Arrest; Low Blood Pressure/ Hypotension

    MAUDE report 23336413 ↗
  • DeathMay 24, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE THERE WERE DIFFICULTIES IN USING THE FIRST INTRODUCER (13CM) DUE TO PATIENT ANATOMY. A DIFFERENT LONGER INTRODUCER WAS USED (23CM) AND ABLE TO PASS THE SUBCLAVIAN ANGLE. THE IMPLANT PROCEDURE WAS COMPLETED SUCCESSFULLY. AFTER THE IMPLANT THE PATIENT HAD SYMPTOMS AND WAS TAKEN TO CARDIAC SURGERY WHERE A SUBCLAVIAN VEIN LACERATION WAS DISCOVERED. THE LACERATION WAS STITCHED UP. THE PATIENT DIED FOLLOWING THE PROCEDURE. THE DEVICE WAS DISCARDED. PATIENT DEATH WAS ON (B)(6) 2023. OTHER EVENT RELATED TO THE ISSUE REPORTED: 1035166-2023-00047 ADDRESSES THE 13CM INTRODUCER (SAME PATIENT).

    Patient problems: Perforation of Vessels

    MAUDE report 16993443 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122084. One submission often covers many models, so check the product and lot details in each record.

  • Z-2678-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
  • Z-2680-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
  • Z-2679-2026 · Jun 10, 2026 · Open, Classified
    Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
  • Z-1467-2021 · Feb 22, 2021 · Terminated
    AngioDynamics Safe Sheath Ultralite - Vascular Introducer/Catheter Introducer Catalogue #: VI10 - Product Usage: For the introduction of various types of pacing leads and catheter

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page