SafeSheath® Ultra SU9Recalls & reported problems
Oscor Inc. · Hemostatic Tear-away Introducer System with Infusion Side Port, 9F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 30891492002471), received between Feb 3, 2017 and Feb 3, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Backflow” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K122084); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Backflow2
Patient problems
- Blood Loss2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2017Backflow; Backflow
THE CUSTOMER REPORTED THE INTRODUCER/DILATOR WAS INSERTED INTO AXILLARY VEIN OVER THE INCLUDED GUIDEWIRE IN STANDARD FASHION. THE DILATOR WAS THEN REMOVED WHILE KEEPING THE GUIDEWIRE IN PLACE. THERE WAS SIGNIFICANT BLEED-BACK THROUGH THE HOMEOSTATIC VALVE (PATIENT WITH ELEVATED PULMONARY/VENOUS PRESSURE) WITH ONLY THE GUIDEWIRE IN PLACE THROUGH THE HOMEOSTATIC VALVE. PHYSICIAN FELT THE BLEED-BACK VIA THE HOMEOSTATIC VALVE WAS EXCESSIVE AND POSSIBLY DEFECTIVE. ONE OF THE INTRODUCERS WAS RETURNED AND ONE WAS DISPOSED OF IN THE FIELD. THE VISUAL ANALYSIS DIDN'T FIND ANY ANOMALIES ON THE RETURNED INTRODUCER, HOWEVER THE RETURNED PRODUCT ANALYSIS LAB DOESN'T PERFORM LEAK TESTING ON INTRODUCERS. ADDITIONAL PRODUCT EVENT DETAILS TWO DIFFERENT LOT NUMBER SU9 INTRODUCERS WERE INITIALLY OPENED FOR THE PROCEDURE, SO IT IS UNKNOWN WHICH IS THE AFFECTED LOT. THOUGH, THE INTRODUCER WITH THE ABOVE DESCRIBED PROBLEM WAS REMOVED FROM THE BODY INTACT AND PLACED IN A RETURNED PRODUCT BAG. THE AFFECTED INTRODUCER WAS REPLACED WITH ANOTHER MODEL SU9 INTRODUCER. THE CUSTOMER REPORTED THAT THEY DID NOT RECEIVE THIS DEVICE FOR ANALYSIS.
Patient problems: Blood Loss
MAUDE report 6301804 ↗MalfunctionFeb 3, 2017Backflow; Backflow
THE CUSTOMER REPORTED THE INTRODUCER/DILATOR WAS INSERTED INTO AXILLARY VEIN OVER THE INCLUDED GUIDEWIRE IN STANDARD FASHION. THE DILATOR WAS THEN REMOVED WHILE KEEPING THE GUIDEWIRE IN PLACE. THERE WAS SIGNIFICANT BLEED-BACK THROUGH THE HOMEOSTATIC VALVE (PATIENT WITH ELEVATED PULMONARY/VENOUS PRESSURE) WITH ONLY THE GUIDEWIRE IN PLACE THROUGH THE HOMEOSTATIC VALVE. PHYSICIAN FELT THE BLEED-BACK VIA THE HOMEOSTATIC VALVE WAS EXCESSIVE AND POSSIBLY DEFECTIVE. ONE OF THE INTRODUCERS WAS RETURNED AND ONE WAS DISPOSED OF IN THE FIELD. THE VISUAL ANALYSIS DIDN'T FIND ANY ANOMALIES ON THE RETURNED INTRODUCER, HOWEVER THE RETURNED PRODUCT ANALYSIS LAB DOESN'T PERFORM LEAK TESTING ON INTRODUCERS. ADDITIONAL PRODUCT EVENT DETAILS: TWO DIFFERENT LOT NUMBER SU9 INTRODUCERS WERE INITIALLY OPENED FOR THE PROCEDURE, SO IT IS UNKNOWN WHICH IS THE AFFECTED LOT. THOUGH, THE INTRODUCER WITH THE ABOVE DESCRIBED PROBLEM WAS REMOVED FROM THE BODY INTACT AND PLACED IN A RETURNED PRODUCT BAG. THE AFFECTED INTRODUCER WAS REPLACED WITH ANOTHER MODEL SU9 INTRODUCER LOT NUMBER DP02107.
Patient problems: Blood Loss
MAUDE report 6299660 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122084. One submission often covers many models, so check the product and lot details in each record.
- Z-2678-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
- Z-2680-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
- Z-2679-2026 · Jun 10, 2026 · Open, ClassifiedAbiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
- Z-1467-2021 · Feb 22, 2021 · TerminatedAngioDynamics Safe Sheath Ultralite - Vascular Introducer/Catheter Introducer Catalogue #: VI10 - Product Usage: For the introduction of various types of pacing leads and catheter
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page