ARROW IPN037306Recalls & reported problems
Teleflex Incorporated · Pressure Injectable Arrowg+ard Blue Advance(R) One-Lumen PICC pre-loaded with Arrow(R) VPS Precision(TM) Stylet · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Teleflex Arrow PICC is a long-term intravenous catheter. The 25 reports mostly describe the catheter or dilator bending, breaking, fraying, or leaking; some also mention allergic reactions. There were 6 injuries and no deaths. One open recall began in April 2026 because kits included lidocaine, bupivacaine, and saline covered by a supplier recall following quality issues and limited reports of ineffectiveness.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 25
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 25 adverse event reports that list this exact device (DI 50801902120781), received between Jan 29, 2021 and Oct 8, 2025 — classified as 6 injuries, 19 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
1 FDA recall record name this device's identifier, 1 still open.
8 other recalls cover devices cleared under the same FDA submission (K153487); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 21, 2026Open, Classified
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
Z-2378-2026 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Break4
- Deformation Due to Compressive Stress4
- Material Frayed4
- Fluid/Blood Leak3
- Biocompatibility2
- Material Separation2
- Adverse Event Without Identified Device or Use Problem1
- Component Misassembled1
Patient problems
- No Clinical Signs, Symptoms or Conditions17
- Hypersensitivity/Allergic reaction5
- Convulsion/Seizure2
- Insufficient Information2
- Nausea2
- Device Embedded In Tissue or Plaque1
- Loss of consciousness1
Recent adverse event reports
The 25 most recent of 25 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT "PICC LEAKING NEAR WHERE SLIDE CLAMP WAS DISENGAGED PICC REMOVED PER MD OTHER CENTRAL LINE WILL BE ORDERED FOR PATIENT." ANOTHER CATHETER WAS PLACED. THERE WAS NO PATIENT HARM OR INJURY. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23245251 ↗MalfunctionMay 20, 2024Material Frayed; Material Frayed
IT WAS REPORTED "CUSTOMER REPORTED THAT TIP OF DILATOR FRAYED UPON TRYING TO INSERT. CUSTOMER HAD DROP A SINGLE STERILE DILATOR TO COMPLETE INSERTION." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353096 ↗MalfunctionMay 20, 2024Material Frayed; Material Frayed
IT WAS REPORTED "CUSTOMER REPORTED THAT TIP OF DILATOR FRAYED UPON TRYING TO INSERT. CUSTOMER HAD DROP A SINGLE STERILE DILATOR TO COMPLETE INSERTION." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353094 ↗MalfunctionMay 20, 2024Material Frayed; Material Frayed
IT WAS REPORTED "CUSTOMER REPORTED THAT TIP OF DILATOR FRAYED UPON TRYING TO INSERT. CUSTOMER HAD DROP A SINGLE STERILE DILATOR TO COMPLETE INSERTION." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19353068 ↗MalfunctionMay 17, 2024Material Frayed; Material Frayed
IT WAS REPORTED "CUSTOMER REPORTED THAT TIP OF DILATOR FRAYED UPON TRYING TO INSERT. CUSTOMER HAD DROP A SINGLE STERILE DILATOR TO COMPLETE INSERTION." NO PATIENT HARM OR INJURY. NO MEDICAL INTERVENTION REQUIRED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS "FINE".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19340473 ↗MalfunctionOct 25, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED "KINKED PICC WHILE IN-DWELLING". THE CATHETER WAS PLACED ON (B)(6) 2023 IN THE RIGHT BRACHIAL. THE CATHETER WAS NOT FLUSHING AND HAD SLUGGISH OR ABSENT BLOOD RETURN. THE PICC DID NOT HAVE TO BE REMOVED. THE PICC WAS PULLED OUT 2 CM AND XRAY CONFIRMED IT WAS PLACED CENTRAL. NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18003611 ↗InjurySep 11, 2023Biocompatibility; Biocompatibility
IT WAS REPORTED "PATIENT HAD AN AGBA PICC INSERTED AND WITHIN SECONDS HAD AN ANAPHYLACTIC REACTION. PATIENT'S THROAT CLOSED CAUSING AN EMERGENCY. THE PICC CATHETER WAS REMOVED IMMEDIATELY." THE PATIENT REQUIRED ANAPHYLAXIS MEDICATIONS IN LINE WITH ACLS REGIMEN. A NEW VASCULAR ACCESS WAS PLACED. THE PATIENT'S CURRENT CONDITION IS REPORTED AS FINE.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 17718758 ↗MalfunctionAug 29, 2023Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED "WIRE KINKED OUT OF BOX". THERE WAS NO PATIENT INVOLVEMENT WITH THE DEVICE. A NEW KIT WAS OPENED AND A LINE PLACED WITHOUT FURTHER ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17646340 ↗MalfunctionJun 28, 2023Deformation Due to Compressive Stress; Physical Resistance/Sticking; Deformation Due to Compressive Stress; Physical Resistance/Sticking
IT WAS REPORTED "WIRE HAS EXTRA PAINT ON THE PROXIMAL END AND MADE IT SO THE DILATOR COULD NOT BE SUCCESSFULLY THREADED OVER THE WIRE - DEFECT IS VISUALLY AVAILABLE".
Patient problems: Insufficient Information
MAUDE report 17216227 ↗InjuryMar 27, 2023Biocompatibility; Adverse Event Without Identified Device or Use Problem; Biocompatibility; Adverse Event Without Identified Device or Use Problem
MEDWATCH FORM RECEIVED REPORTS: "PICC LINE INSERTION. CASE PROCEEDED NORMALLY UNTIL LINE WAS IN SUPERIOR VENA CAVA. WHILE POSITIONING THE LAST FEW CM, PATIENT STATED HE WAS FEELING NAUSEOUS. PATIENT FELT THE NEED TO URINATE. PATIENT THEN BEGAN HAVING SEIZURE LIKE MOVEMENTS. PATIENT BECAME UNRESPONSIVE. PROCEDURE ABORTED. CODE BLUE CALLED. DISCOVERED TO BE AN ALLERGIC REACTION TO THE CHLOREHEXIDINE." ADDITIONAL INFORMATION RECEIVED FROM THE SALES REP 25-APR-2023 INDICATES THE PATIENT HAS A COMPLEX MEDICAL HISTORY WITH SEVERAL ALLERGIES. IT WAS ALSO REPORTED THAT THE CODE BLUE WAS CALLED WHEN THE PATIENT'S HEART RATE CHANGED, AND THE PATIENT BECAME UNRESPONSIVE BUT NEVER LOST PULSE. THEY NEEDED TO REMOVE THE CATHETER AND TREAT THE PATIENT WITH LEVOPHED, EPI, BENADRYL AND OXYGEN. SHORTLY AFTER REMOVAL OF THE LINE AND TREATMENT, THE ALLERGY-LIKE SYMPTOMS RESOLVED. THE PATIENT WAS TRANSFERRED TO THE ICU FOR JUST UNDER 24 HOURS FOR OBSERVATION. THERE WAS NO INJURY TO THE PATIENT AND THE CURRENT CONDITION OF THE PATIENT WAS REPORTED AS "FINE". THE PATIENT WAS TO FOLLOW UP WITH AN ALLERGIST TO CONFIRM THE CHG ALLERGY; HOWEVER, IT WAS REPORTED THAT THE PATIENT CANCELLED THEIR LAST ALLERGY T
Patient problems: Convulsion/Seizure; Hypersensitivity/Allergic reaction; Loss of consciousness; Nausea
MAUDE report 16620205 ↗MalfunctionMar 27, 2023Component Misassembled; Component Misassembled
UPON OPENING THE KIT, THE NURSE FOUND A DOUBLE LUMEN PICC INCLUDED INSTEAD OF A SINGLE LUMEN PICC. ANOTHER KIT HAD TO BE OPENED. NO PATIENT INVOLVEMENT WITH THE DEVICE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16620145 ↗InjuryMar 7, 2023Use of Device Problem
PICC LINE INSERTION. CASE PROCEEDED NORMALLY UNTIL LINE WAS IN SUPERIOR VENA CAVA. WHILE POSITIONING THE LAST FEW CM, PATIENT STATED HE WAS FEELING NAUSEOUS. PATIENT FELT THE NEED TO URINATE. PATIENT THEN BEGAN HAVING SEIZURE LIKE MOVEMENTS. PATIENT BECAME UNRESPONSIVE. PROCEDURE ABORTED. CODE BLUE CALLED. DISCOVERED TO BE AN ALLERGIC REACTION TO THE CHLOREHEXIDINE.
Patient problems: Convulsion/Seizure; Hypersensitivity/Allergic reaction; Nausea
MAUDE report 16494835 ↗MalfunctionSep 15, 2022Break; Break
CUSTOMER COMPLAINT REPORTS "PEEL AWAY SHEATH HUB CRACKED AND BROKE OFF AS CLINICIAN TRIED TO BREAK AND PEEL AWAY". NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15428318 ↗MalfunctionSep 15, 2022Break; Break
CUSTOMER COMPLAINT REPORTS "HUB CRACKED OFF ON PEEL AWAY SHEATH AS CLINICIAN TRIED TO BREAK AND PEEL AWAY SHEATH FROM PATIENT". NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15428317 ↗MalfunctionSep 15, 2022Break; Break
CUSTOMER COMPLAINT REPORTS "HUB CRACKED OFF ON PEEL AWAY SHEATH AS CLINICIAN TRIED TO BREAK AND PEEL AWAY SHEATH FROM PATIENT". NO PATIENT HARM WAS REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15428144 ↗MalfunctionJun 8, 2022Break; Break
IT WAS REPORTED THE PEEL-AWAY SHEATH WING BROKE OFF WHEN THEY WENT TO PEEL THE SHEATH. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14633756 ↗InjuryMar 12, 2022Material Separation; Material Separation
THE COMPLAINT IS REPORTED AS: PART OF THE PEELAWAY SHEATH BROKE OFF IN THE PATIENT'S SUBCUTANEOUS TISSUE UPON REMOVAL. THE PIECE OF SHEATH HAS REMAINED IN THE PATIENT'S ARM SINCE THE INSERTION ON 28-JANUARY. REMOVAL OF THE SHEATH PIECE WAS TO TAKE PLACE YESTERDAY OR TODAY 18-FEBRUARY BY CUT DOWN OF THE SUBCUTANEOUS TISSUES AT THE INSERTION SITE. THE PIECE WAS NOT ACTUALLY IN THE VESSEL. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13742509 ↗MalfunctionMar 1, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
CUSTOMER REPORTED PICC LINE WAS PLACED 19-JAN AND HAD INTERMITTENT ISSUES WITH FLUSHING. TPA (TISSUE PLASMINOGEN ACTIVATOR) WAS GIVEN ON 03-FEB, BUT ON 04-FEB THE NURSE COULD NOT FLUSH THE CATHETER UNTIL HE REPOSITIONED IT AND STRAIGHTENED OUT A "SUBQ KINK". ONCE THE KINK WAS STRAIGHTENED OUT, THE CATHETER FLUSHED WITH EASE. IT WAS ALSO REPORTED THE PICC WAS LEAKING FROM THE LUER HUB/TUBING CONNECTION (CAPTURED IN MDR# 9680794-2022-00118) AND MAY HAVE BEEN SUBJECTED TO A LOT OF PRESSURE BY NURSING STAFF ATTEMPTING TO FLUSH IT. NO PATIENT HARM REPORTED. THE PICC LINE WAS REPLACED WITH A NEW PICC. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13638908 ↗MalfunctionMar 1, 2022Fluid/Blood Leak; Fluid/Blood Leak
CUSTOMER REPORTED PICC THAT WAS LEAKING FROM THE LUER HUB/TUBING CONNECTION. IT WAS REPORTED THE PICC ALSO HAD INTERMITTENT ISSUES WITH FLUSHING AND "SUBQ KINK" NOTED (CAPTURED IN MDR # 9680794-2022-00164). THE PICC LINE WAS REPLACED WITH A NEW PICC. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13638749 ↗MalfunctionJan 4, 2022Material Separation; Material Separation
NURSE REPORTED THE CATHETER WAS FOUND WITH HUB SNAPPED OFF CLEAN. CATHETER WAS THROWN AWAY AND REPLACED. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13151638 ↗MalfunctionDec 20, 2021Fluid/Blood Leak; Fluid/Blood Leak
WHEN INFUSING, DEVICE IS LEAKING WHERE THE LUMEN MEETS THE PINK LUER LOCK. THE DEVICE WAS REMOVED AND A NEW ONE PLACED. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13047119 ↗MalfunctionDec 20, 2021Fluid/Blood Leak; Fluid/Blood Leak
WHEN INFUSING, DEVICE IS LEAKING WHERE THE LUMEN MEETS THE PINK LUER LOCK. THE DEVICE WAS REMOVED AND A NEW ONE PLACED. NO PATIENT HARM REPORTED. THE PATIENT'S CONDITION IS REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13046972 ↗InjuryJul 22, 2021Patient-Device Incompatibility; Patient-Device Incompatibility
THE COMPLAINT IS REPORTED AS: GETTING READY TO PUT DRESSING ON. BBE (BLUE BULLSEYE) OBTAINED. FLUSHED PICC, AFTER FLUSH PT COMPLAINED OF NAUSEA. RN ATTEMPTED TO FINISH CASE. GOT UNSTERILE/JIGGLED WIRES TO MAKE SURE THE CASE PRINTED. PT BECAME DIAPHORETIC AND VITAL SIGNS WERE ABNORMAL. PICC REMOVED AND CHEST XRAY PERFORMED. TACHYCARDIC, PT COMPLAINED OF BLURRED VISION. PHYSICIAN SUSPECTED AIR EMBOLISM. PT INDICATED THROAT FELT SWOLLEN. IT WAS REPORTED THERE WAS NO AIR EMBOLISM AND CHG (CHLORHEXIDINE) WAS ADDED AS AN ALLERGY FOR THE PATIENT. THE PATIENT WAS TREATED WITH BENADRYL AND STEROIDS. THE PATIENT HAS BEEN DISCHARGED FROM THE HOSPITAL AND IS DOING FINE.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 12213079 ↗InjuryApr 20, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
THE COMPLAINT IS REPORTED AS: "PICC INSERTION WENT FINE INITIALLY. STYLET MOVED FINE WHEN GOING TO TRIM. IN SVC AND PICC WAS GOING INTO SVC AND AS SOON AS WENT TO CAJ PT STARTED FROTHING AT THE MOUTH, SPITTING, NAUSEA, RIGID, PTS WIFE WAS AT BEDSIDE AND HAD INSERTER REMOVE CATHETER. BLOOD SUGAR WAS 62 AND NOT ABLE TO GET BLOOD PSI. PERFORMED CPR AND INTUBATED. TRANSFERRED TO SICU." PATIENT'S CONDITION REPORTED AS CRITICAL AT THE TIME OF THIS REPORT.
Patient problems: Hypersensitivity/Allergic reaction
MAUDE report 11696241 ↗MalfunctionJan 29, 2021Retraction Problem
WHEN TRIMMING PICC FOR PLACEMENT, I RETRACTED THE VPS WIRE. IT BEGAN TO GATHER IN CENTER OF THE PICC DURING RETRACTION IN AN ACCORDION LIKE FASHION. UPON INSPECTION, IT WAS NOTED TO HAVE TWO FLATTENED CRIMPS IN PICC. NEW PICC OBTAINED FOR PROCEDURE, NO INJURY TO PATIENT, BUT POSSIBLY A MANUFACTURING ISSUE. FDA SAFETY REPORT ID # (B)(4).
Patient problems: Insufficient Information
MAUDE report 11261178 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K153487. One submission often covers many models, so check the product and lot details in each record.
- Z-0001-2024 · Aug 10, 2023 · Open, ClassifiedArrow ErgoPack, Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC pre-loaded with Arrow VPS Precision Stylet, REF CDC-45552-VPS2; Catheter: Percutaneous Implanted Long-Ter
- Z-1807-2019 · Apr 23, 2019 · TerminatedArrowg+ard Blue Advance PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 44063 VPS2
- Z-1808-2019 · Apr 23, 2019 · TerminatedArrowg+ard Blue Advance" PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 45041 VPS2
- Z-1810-2019 · Apr 23, 2019 · TerminatedArrowg+ard Blue Advance PICC pre loaded with Arrow VPS" Precision Stylet Product Code: CDC 45063 VPS2
- Z-1365-2019 · Dec 13, 2018 · TerminatedArrow PICC powered by Arrow VPS Precision Stylet Product Code:CDC-45052-VPS2 Product Usage: The Arrow Pressure Injectable PICC is indicated for short-term or longterm peripheral a
- Z-1366-2019 · Dec 13, 2018 · TerminatedArrow PICC powered by Arrow VPS Precision Stylet Product Code:CDC-44052-VPS2 Product Usage: The Arrow Pressure Injectable PICC is indicated for short-term or longterm peripheral a
- Z-3200-2018 · Jul 17, 2018 · Terminated5.5 Fr x 40 cm double-lumen PICC Kit, Product Code CDC-44052-VPS2
- Z-3199-2018 · Jul 17, 2018 · Terminated4.5 Fr x 40 cm single-lumen PICC Kit, Product Code CDC-44041-VPS2, 5.5 Fr double-lumen PICC Kit, Product Code CDC-44052-VPS2
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.