ARROW recallZ-2378-2026 · Arrow International, LLC · initiated Apr 21, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Arrow International, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: [email protected] On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
Recall details
- Recall number
- Z-2378-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 21, 2026
- Posted by the FDA
- Jun 13, 2026
- Recalling firm
- Arrow International, LLC
- Quantity
- 3,481
- Product code
- LJS
- FDA submissions
- K042126, K112896, K113277, K122545, K130876, K153423
- Distribution
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Product description
Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR x 55CM w Biopatch/DLX-45541-HPKB; PICC/DELTA KIT: 2L 5.5 FR X 50 CM/ASK-45052-MC; AGBA PI PICC KIT: 1L 4.5FR 40CM/CDC-44041-HPK1A; AGBA PI PICC KIT 2-LUMEN: 5.5 FR X 40CM/CDC-44052-HPK1A; AGBA PICC/DELTA FG: 2-L 5.5 FR X 40 CM/CDC-44052-VPS2; PI AGBA PICC KIT: 3-L 6 FR X 50 CM/CDC-45063-HPK1A; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 55 CM/CDC-45541-VPS2; AGBA PICC/DELTA FG: 2-L 5.5 FR X 55 CM/CDC-45552-VPS2; PICC/VPS 1L 4.5 FR/ASK-45541-CHS; PI AGBA PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UOI2; PICC KIT: 2L 5.5FR 55CM AGBA VPS/ASK-45552-CHS; PI AGBA PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-UOI2; PICC KIT: 3L 6FR 55CM AGBA/ASK-45563-CHS; PI JACC KIT: 3L 6 FR X 25 CM CG+/ASK-42563-LTAC; PI JACC KIT: 1-L 4.5 FR X 25 CM CG+/CDC-42541-JX1A; PI JACC KIT: 2-L 5.5 FR X 25 CM CG+/CDC-42552-JX1A; PI JACC KIT: 3-L 6 FR X 25 CM CG+/CDC-42563-JX1A; PI JACC KIT: 2-L 5.5 FR X 30 CM CG+/CDC-43052-JX1A; PI JACC KIT: 3-LUMEN: 6 FR X 30 CM CG+/CDC-43063-JX1A; PICC KIT: 2-L 4 FR X 55 CM/CDC-05542-PK1A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
REF(Material)/UDI-DI/Lot(Batch): DLX-35563-HPKC/10801902193275/33F25D0520; ASK-44552-ABKB/10801902140798, 10801902216677/33X25K0002, 33S24A0529; ASK-45552-MHBV/10801902193275/33F25K0372, 33F25M0199, 33F26A1534, 33F24L0348; DLX-45541-HPKB/10801902193275, 10801902208849/33F25G0393, 33F25L0055, 33F26A1570, 33F24D0604; ASK-45052-MC/10801902216660/33S24A1242; CDC-44041-HPK1A/10801902193275/33F24K0430; CDC-44052-HPK1A/10801902193275/33F25D0241; CDC-44052-VPS2/10801902193275/33F25E0514; CDC-45063-HPK1A/10801902193275/33F25D0242; CDC-45541-VPS2/50801902120781, 10801902193275/33S24A0211, 33F25C0699; CDC-45552-VPS2/10801902193275/33F25E0704; ASK-45541-CHS/10801902193275/33F26A1422, 33F24L0897; ASK-45541-UOI2/10801902193275/33F24L0991; ASK-45552-CHS/10801902193275/33F26A1395, 33F25G0260; ASK-45552-UOI2/10801902193275/33F24L0431; ASK-45563-CHS/10801902193275/33F25A0199; ASK-42563-LTAC/10801902203271/33F24F0852; CDC-42541-JX1A/10801902193275/33F26A1469, 33F24J0135; CDC-42552-JX1A/10801902193275/33F25D0561, 33F24L0285; CDC-42563-JX1A/10801902193275, 10801902170863/33F25C0702, 33F25D0562, 33F25G0311, 33F25J0437, 33F24D0507; CDC-43052-JX1A/10801902193275/33F25M0062, 33F24J0120; CDC-43063-JX1A/10801902193275, 10801902188332/33F25G0003, 33F26B0805, 33F25D0240; CDC-05542-PK1A/10801902193275, 50801902116715/33F25H0436, 33F25H0576, 33F24D0325
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- ARROW IPN917372Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls1 report
- ARROW IPN924180Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days2 recalls0 reports
- ARROW IPN928638Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days2 recalls0 reports
- ARROW IPN036456Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days2 recalls0 reports
- ARROW IPN037306Teleflex Incorporated · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall25 reports
- ARROW IPN038495Teleflex Incorporated · INTRODUCER, CATHETER1 recall1 report
- ARROW IPN928635Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall1 report
- ARROW IPN919539Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- ARROW IPN922919Teleflex Incorporated · Catheter, intravascular, therapeutic, short-term less than 30 days1 recall0 reports
Product lines
Other devices under the same FDA submission
71 devices cleared under K042126, K112896, K113277, K122545, K130876, K153423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARROW IPN924181Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days2 recalls0 reports
- ARROW IPN033424Teleflex Incorporated · Catheter,intravascular,therapeutic,long-term greater than 30 days70 reports
- ARROW IPN924163Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall8 reports
- ARROW IPN924156Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall6 reports
- ARROW IPN924157Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall5 reports
- ARROW IPN924159Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall4 reports
- ARROW IPN924178Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall2 reports
- ARROW IPN056680Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall1 report
- ARROW IPN919553Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- ARROW IPN919554Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- ARROW IPN924162Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
- ARROW IPN928624Teleflex Incorporated · Catheter, intravascular, therapeutic, long-term greater than 30 days1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98793).
- Z-2363-2026 · Open, ClassifiedPARACENTESIS KIT, REF: AK-00376, ASK-00376-VCU
- Z-2364-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-SUH-S; ACCESS TRAY/ASK-04001-VM1; STIMUCATH SFTY KIT: 19G CTH, 17Gx4CM NDL/AB-19604-KS; STIMUCATH SFTY KIT: 19G CTH, 17Gx8CM NDL/AB-19608-KS;…
- Z-2365-2026 · Open, ClassifiedDescription/REF: BIOPSY SYSTEM COMPREHENSIVE KIT/ASK-09458-AH, ASK-09458-COC, ASK-09458-EUH, ASK-09458-SD, ASK-09458-VU
- Z-2366-2026 · Open, ClassifiedDescription/REF: PACING PSI KIT: 6 FR/AI-06210-KS, PACING PSI KIT: 5 FR 2-L/AI-07155-KS, PACING PSI KIT: 5 FR 2L/ASK-07155-UPM
- Z-2367-2026 · Open, ClassifiedDescription/REF: AGBA PICC G4 Stylet: 3L 6FR x 55CM BP/DLX-45563-VPSB; AGBA PICC G4 Stylet: 3L 6FR x 55CM TCG/DLX-45563-VPSC; ARROW(R) VPS(R) ACCESS KIT/ASK-04001-UCHD1; ARROW(R) V…
- Z-2368-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1 3/4 IN/TJ-04020-1; RA CATH KIT: 20 GA X 1 1/2IN/NA-04220-S1A; ARTERIAL LINE KIT/AK-04510-S; ARTERIAL CATH KIT: 18 GA X 6IN (16 CM)/ASK-04018…
- Z-2369-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4IN ANTIMICROBIAL/ASK-09701-UMH; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU4; PSI/CVC KIT: 9 FR/7 FR X 6IN (16CM)/ASK-09903-AMC1; PSI KIT: 9 …
- Z-2370-2026 · Open, ClassifiedDescription/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/ASK-45052-NS AGBA PICC/DELTA KIT: 1L 4.5 FR X 55 CM/ASK-45541-MHBV PICC KIT…
- Z-2371-2026 · Open, ClassifiedDescription/REF: EPIDURAL CATHETER KIT: 19 GA X 90 CM/ASK-05502-VM1; EPIDURAL CATHETERIZATION KIT: 19 GA/AK-05501; EPIDURAL CATHETERIZATION KIT/AK-05502; EPIDURAL CATHETERIZATION K…
- Z-2372-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2L 14 FR X 20 CM AGB CDC-25142-X1A
- Z-2373-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 8 FR X 11 CM/AK-11802-F; HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-22122; HEMODIALYSIS KIT 2L: 14 FR X 15CM AGB/AK-22142-CDC; HEMODIALYS…
- Z-2374-2026 · Open, ClassifiedDescription/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-1…
- Z-2376-2026 · Open, ClassifiedDescription/REF: CVC/PSI KIT: 2 LUMEN 8 FR X 6 IN/ASK-42802-PLH; CVC KIT: 3-LUMEN 12 FR X 16 CM/AK-12123-F; CVC KIT: 3-LUMEN 12 FR X 20 CM/AK-15123-F; 3L CVC KIT: 7 FR X 16 CM/ASK-…
- Z-2377-2026 · Open, ClassifiedDescription/REF: ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM6 RA CATH KIT: 20 GA X 1-3/4IN/NA-04020-X1ASP ACCESS TRAY/ASK-04001-MS RA CATH KIT: 20 GA X 8 CM/ASK-00820-UM1 VESSE…
- Z-2379-2026 · Open, ClassifiedDescription/REF: PI JACC KIT: 2L 5.5 FR X 15 CM CG+/ASK-41552-LTAC; PI JACC KIT: 3L 6 FR X 20 CM CG+/ASK-42063-LTAC; AGBA PI JACC KIT: 1-L 4.5 FR X 15 CM/CDC-41541-JX1A; AGBA PI JA…
- Z-2380-2026 · Open, ClassifiedDescription/REF: PERITONEAL LAVAGE CATH KIT:8FR/AK-09000
- Z-2381-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 16 CM/AK-12122-F
- Z-2382-2026 · Open, ClassifiedDescription/REF: DRAINAGE KIT/AK-01600, DRAINAGE KIT/AK-01601, DRAINAGE KIT/ASK-01601-MM, THORACENTESIS KIT/AK-01000-T, THORACENTESIS KIT/AK-01000, THORACENTESIS KIT/ASK-01000-CHP,…
- Z-2383-2026 · Open, ClassifiedDescription/REF: PACING/PSI KIT: 5 FR 2L/ASK-07155-MGU
- Z-2384-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/ASK-04001-JHH; ACCESS TRAY/ASK-04001-LMDTG; PI CVC KIT: 3 LUMEN 7 FR X 6 IN (16 CM)/ASK-12703-PSPH1; 3L CVC KIT: 7 FR X 16 CM/ASK-12703-WMC2; MULTILUME…
- Z-2385-2026 · Open, ClassifiedDescription/REF: ACCESS TRAY/GU-04020; PSI KIT: 8.5 FR/AK-09803-CDC; PSI KIT: 9 FR/AK-09903-CDC; PSI KIT: 8.5 FR X 4 IN (10 CM)/ASK-09810-PCMH1; PSI KIT: 9 FR X 10 CM/ASK-09903-KH;…
- Z-2387-2026 · Open, ClassifiedDescription/REF: PI MIDLINE 2L: 5FR X 20CM/CDC-32052-MPKC
- Z-2388-2026 · Open, ClassifiedDescription/REF: HEMODIALYSIS KIT: 2-LUMEN 12 FR X 20 CM/AK-15122-F, AHDC KIT: 3L 12 FR X 16 CM/ASK-12123-UOC1, AHDC KIT: 3-L 12 FR X 16 CM/ASK-12123-UPM, AHDC KIT: 3L 12 FR X 20 C…
- Z-2389-2026 · Open, ClassifiedDescription/REF: SPINAL ANESTHESIA KIT/ASA-24090-SB; SPINAL ANESTHESIA KIT/ASA-24103-SB; SPINAL ANESTHESIA KIT/ASA-25090-SB; SPINAL ANESTHESIA KIT: 24 GA X 9 CM/ASK-24090-SA; SPINA…
- Z-2390-2026 · Open, ClassifiedDescription/REF: SAC KIT: 18 GA X 12CM W BIO/SAC-01218-B; SAC KIT: 18 GA X 12CM W CHG/SAC-01218-C; SAC KIT: 18 GA X 16CM W BIO/SAC-01618-B
- Z-2391-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-3/4IN/AK-04020-SPC; RA CATH KIT: 20 GA X 5 IN/ASK-04020-MW; ARTERIAL LINE KIT: SL 20 GA X 5 IN/ASK-04510-MM5; CATH SL 20 GA X 5 CM HYDROPHIL…
- Z-2393-2026 · Open, ClassifiedDescription/REF: PSI KIT: 7 FR X 4 IN/ASK-09701-UP; PSI KIT: 8.5 FR X 4 1/8 IN (10 CM)/ASK-09803-OSU5; PSI KIT: 8.5 FR/ASK-09807-UC; PSI KIT: 9 FR X 4 1/8IN (10 CM)/ASK-09903-HH2; …
- Z-2394-2026 · Open, ClassifiedDescription/REF: MIDLINE CATHETER KIT: 3 FR X 20 CM/CDC-02031-MK1A; AM/AT PI MIDLINE 1L: 4.5FR X 15CM/CDC-41541-MPK1A; AM/AT PI MIDLINE 2L: 5.5FR X 15CM/CDC-41552-MPK1A; AGBA PI MI…
- Z-2396-2026 · Open, ClassifiedDescription/REF: CVC KIT: 16 GA X 8IN/AK-04210, CVC KIT: 1-L 16 GA X 8IN (20 CM)/AK-04300, CVC KIT: 14 GA X 6 IN/CK-04711
- Z-2397-2026 · Open, ClassifiedDescription/REF: RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-1A; RA CATH KIT: 20 GA X 1-1/2IN/NA-04220-X1A; ARTERIAL CATHETER KIT: 18 GA X 16 CM/ASK-04018-HCA; ARTERIAL CATHETER KIT: 18 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page