DRX Plus 3543C DetectorRecalls & reported problems
Carestream Health, Inc. · AY-PACKED-3543C DETECTOR-DRX PLUS · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 60889974000238), received between Aug 11, 2021 and Aug 11, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Image Display Error/Artifact” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Image Display Error/Artifact1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2021Image Display Error/Artifact; Image Display Error/Artifact
ON (B)(6) 2021, CARESTREAM HEALTH (CSH) WAS INFORMED OF AN INCIDENT RELATED TO THE DRX PLUS 3543C DETECTOR WHICH OCCURRED AT (B)(6) HOSPITAL ((B)(6)). PER THE REPORT, AFTER A SURGICAL PROCEDURE WAS COMPLETED AND THE PATIENT WAS STITCHED UP, PER HOSPITAL PROTOCOL, THE CUSTOMER PERFORMED A COUNT OF SURGICAL INSTRUMENTS AND DETERMINED THE COUNT TO BE OFF. THE CUSTOMER THEN PROCEEDED AS FOLLOWS: TOOK 1ST IMAGE OF ONE (1) FORCEPS USING A DRX REVOLUTION SYSTEM AND THE DRX PLUS 3543C DETECTOR (SN: (B)(4)). TOOK 2ND AND 3RD IMAGES OF PATIENT ABDOMEN WHICH SHOWED A FAINT SIGNAL, RETAINED FROM THE INITIAL FORCEPS IMAGE, WHICH THE CUSTOMER BELIEVED TO BE THE MISSING INSTRUMENT. AS A RESULT, THE PATIENT WAS RE-OPENED IN ERROR AS THE FORCEPS WERE NOT LEFT IN THE PATIENT. THERE WAS NO INJURY OR ADVERSE EFFECT TO THE PATIENT DUE TO THIS INCIDENT. HOWEVER, DUE TO THE CUSTOMER ERROR IN VIEWING THE IMAGES, THE PATIENT WAS RE-OPENED AFTER BEING STITCHED UP.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12310935 ↗
Other Solid state x-ray imager (flat panel/digital imager) devices
- SEDECAL SM-40HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)24 recalls1 report
- SEDECAL SM-40HF-B-D-VIRSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)12 recalls1 report
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)7 recalls0 reports
- DigitalDiagnost C90 712034Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)3 recalls90 reports
- DigitalDiagnost C90 712035Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)4 recalls17 reports
- DigitalDiagnost 712031Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)2 recalls65 reports
- NA GM85Samsung Electronics Co., Ltd. · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls4 reports
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page