Skip to content

AquaBeam Robotic System AB2000CRecalls & reported problems

PROCEPT BioRobotics Corporation · Class 2

Adverse event reports
16
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI B614AB2000C1), received between Jul 10, 2023 and Mar 20, 2024 — classified as 15 injuries, 1 malfunction.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (1 report).

1 other recall cover devices cleared under the same FDA submission (K202961); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem15
  • Visual Prompts will not Clear1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding4
  • Sexual Dysfunction4
  • Urinary Retention3
  • Urinary Tract Infection3
  • Dysuria1
  • Hematuria1
  • No Clinical Signs, Symptoms or Conditions1
  • Sepsis1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION PROCEDURE, THE PATIENT IS EXPERIENCING EJACULATORY DYSFUNCTION (PER THE MANUFACTURER'S INSTRUCTIONS FOR USE, SEXUAL DYSFUNCTION, INCLUDING EJACULATORY AND ERECTILE DYSFUNCTION ARE POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE). THE TREATING PHYSICIAN IS GOING TO OBSERVE THE PATIENT FOR THIS ISSUE AND HAS ASSESSED THE EVENT AS RELATED TO THE PROCEDURE. THE EVENT WAS REPORTED AS MILD AND NON-SERIOUS, WITH NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM REPORTED.

    Patient problems: Sexual Dysfunction

    MAUDE report 18942457 ↗
  • InjuryDec 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT WAS TAKEN BACK TO THE OPERATING ROOM DUE TO BLEEDING, WHICH WAS ADDRESSED WITH CLOT EVACUATION. THE ISSUE IS REPORTED TO HAVE BEEN RESOLVED WITHOUT SEQUELAE. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 18384545 ↗
  • InjuryNov 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD PRESENTED WITH URINARY TRACT INFECTION WHICH WAS TREATED WITH ANTIBIOTICS. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Urinary Tract Infection

    MAUDE report 18085369 ↗
  • InjuryNov 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED URINARY RETENTION WHICH WAS TREATED WITH CATHETERIZATION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE UNRELATED TO THE PROCEDURE OR DEVICE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Urinary Retention

    MAUDE report 18083513 ↗
  • InjuryNov 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED A URINARY TRACT INFECTION WHICH WAS TREATED WITH ANTIBIOTICS. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Urinary Tract Infection

    MAUDE report 18083470 ↗
  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED URINARY RETENTION WHICH WAS TREATED WITH CATHETERIZATION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Urinary Retention

    MAUDE report 17952954 ↗
  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED A URINARY TRACT INFECTION WHICH WAS TREATED WITH MEDICATION AND CATHETERIZATION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Urinary Tract Infection

    MAUDE report 17952408 ↗
  • InjuryOct 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED DYSURIA WHICH WAS TREATED WITH MEDICATION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Dysuria

    MAUDE report 17951927 ↗
  • InjuryOct 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT EXPERIENCED EJACULATORY DYSFUNCTION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Sexual Dysfunction

    MAUDE report 17938931 ↗
  • InjuryOct 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD PRESENTED WITH EJACULATORY DYSFUNCTION (PER MANUFACTURER INSTRUCTIONS FOR USE, SEXUAL DYSFUNCTION, INCLUDING EJACULATORY AND ERECTILE DYSFUNCTION ARE POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE). IT WAS REPORTED THAT THE TREATING PHYSICIAN WAS TO MONITOR THE PATIENT'S CONDITION. THE PHYSICIAN DETERMINED THAT THE EJACULATORY DYSFUNCTION WAS RELATED TO THE PROCEDURE AND UNLIKELY RELATED TO THE DEVICE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Sexual Dysfunction

    MAUDE report 17934955 ↗
  • InjuryOct 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD PRESENTED WITH BLEEDING AND ACUTE URINARY RETENTION (PER MANUFACTURER'S INSTRUCTIONS FOR USE, URINARY RETENTION IS A PERIOPERATIVE RISK OF THE AQUABLATION PROCEDURE). IT WAS REPORTED THAT THE PATIENT WAS TREATED WITH MEDICATION AND CATHETERIZATION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE LIKELY RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Urinary Retention

    MAUDE report 17934949 ↗
  • InjuryOct 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE PATIENT HAD PRESENTED WITH EJACULATORY DYSFUNCTION (PER MANUFACTURER INSTRUCTIONS FOR USE, SEXUAL DYSFUNCTION, INCLUDING EJACULATORY AND ERECTILE DYSFUNCTION ARE POTENTIAL PERIOPERATIVE RISKS OF THE AQUABLATION PROCEDURE) AND HEMATURIA. IT WAS REPORTED THAT HEMATURIA WAS TREATED WITH MEDICATION AND CATHETERIZATION, WHILE THE TREATING PHYSICIAN WAS TO MONITOR THE PATIENT'S CONDITION FOR EJACULATORY DYSFUNCTION. THE PHYSICIAN DETERMINED THAT THE EJACULATORY DYSFUNCTION WAS RELATED TO THE PROCEDURE AND UNLIKELY RELATED TO THE DEVICE, AND ASSESSED THE HEMATURIA AS RELATED TO THE PROCEDURE AND NOT RELATED TO THE DEVICE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Hematuria; Sexual Dysfunction

    MAUDE report 17934923 ↗
  • InjuryAug 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT TWO MONTHS POST-AQUABLATION TREATMENT, THE PATIENT PRESENTED WITH SEPSIS. THE PATIENT WAS GIVEN ANTIBIOTICS AND A CATHETER WAS PLACED, WHICH RESOLVED THE REPORTED ISSUE WITHIN 2 DAYS FROM THE INITIAL OCCURRENCE. HOWEVER, THE ROOT CAUSE OF THIS EVENT REMAINS UNIDENTIFIED. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE LIKELY RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Sepsis

    MAUDE report 17436203 ↗
  • InjuryJul 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION TREATMENT, THE PATIENT REQUIRED A BLOOD TRANSFUSION. THE TREATING SURGEON CONFIRMED THAT THE AQUABEAM ROBOTIC SYSTEM FUNCTIONED AS INTENDED, AND ASSESSED THE EVENT TO BE LIKELY RELATED TO THE PROCEDURE. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17398160 ↗
  • MalfunctionJul 10, 2023Visual Prompts will not Clear; Visual Prompts will not Clear

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE DURING THE AQUABLATION PROCEDURE, MULTIPLE "E208 - MOTORPACK ERROR" MESSAGES WERE GENERATED BY THE AQUABEAM ROBOTIC SYSTEM. MULTIPLE TROUBLESHOOTING STEPS WERE PERFORMED TO RESOLVE THE ISSUE WITHOUT SUCCESS. HOWEVER, ONCE THE POWER REACHED 80%, THE OCCURRENCE OF ERROR CODES CEASED AND THE PROCEDURE WAS CONTINUED THROUGH SUCCESSFUL COMPLETION. THE REPORTED EVENT CAUSED A SURGICAL DELAY OF OVER 20 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17288561 ↗
  • InjuryJul 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT POST-AQUABLATION PROCEDURE THE PATIENT WAS TAKEN BACK TO THE OPERATING ROOM (OR) FOR CLOT EVACUATION AND FULGURATION. THE PATIENT'S HEMOGLOBIN (HB) LEVEL WAS 130 G/DL, WITH NO BLEEDING VESSELS, AND WAS DISCHARGED HOME THE NEXT DAY. NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 17288558 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K202961. One submission often covers many models, so check the product and lot details in each record.

  • Z-2330-2021 · Jul 9, 2021 · Open, Classified
    AQUABEAM Handpiece, REF: HP2000, a sterile single-use component of the AQUABEAM Robotic System.

Other Fluid jet removal system devices

All Fluid jet removal system devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.