AquaBeam Handpiece HP2000 (UDI B614HP20001)Recalls & reported problems
PROCEPT BioRobotics Corporation · Class 2
All models of AquaBeam Handpiece (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI B614HP20001), received between Dec 20, 2021 and Jul 12, 2023 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Display or Visual Feedback Problem3
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Visual Prompts will not Clear1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Urinary Incontinence1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 12, 2023Display or Visual Feedback Problem; Display or Visual Feedback Problem
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT DURING THE AQUABLATION PROCEDURE, THE AQUABEAM SCOPE WAS UNABLE TO PROVIDE A CLEAR VISUALIZATION. MULTIPLE TROUBLESHOOTING STEPS WERE MADE TO ADDRESS THE ISSUE WITHOUT AVAIL. AS A RESULT, THE SURGEON DECIDED TO ABORT THE AQUABLATION PROCEDURE AND CONVERTED TO TRANSURETHRAL RESECTION OF THE PROSTATE (TURP) SURGERY. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17309931 ↗MalfunctionApr 6, 2023Display or Visual Feedback Problem; Visual Prompts will not Clear; Display or Visual Feedback Problem; Visual Prompts will not Clear
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT DURING THE SECOND TREATMENT PASS AN "E22 - MOTORPACK ERROR" WAS GENERATED BY THE AQUABEAM ROBOTIC SYSTEM. MULTIPLE TROUBLESHOOTING STEPS WERE TAKEN IN AN EFFORT TO RESOLVE THE ISSUE, WHICH WAS MOMENTANEOUSLY RESOLVED BUT THE ERROR PERSISTED. AS A RESULT, THE AQUABLATION PROCEDURE WAS ABORTED DUE TO POSITIVE RESULTS WITH THE AMOUNT OF RESECTION OF THE BLADDER TISSUE PREVIOUSLY PERFORMED. UPON FURTHER INSPECTION, IT WAS THEN OBSERVED THAT THE LITHOTOMY DRAPE COVERING THE MOTORPACK WAS BUNCHED UP, CAUSING THE MOTORPACK TO DISENGAGE UPON OPERATION. NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT WERE REPORTED DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16694777 ↗InjuryJul 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT THE PATIENT WAS INCONTINENT POST AQUABLATION PROCEDURE (PER MANUFACTURER'S INSTRUCTIONS FOR USE INCONTINENCE OR OVERACTIVE BLADDER ARE IS A PERIOPERATIVE RISK OF THE AQUABLATION PROCEDURE). NO MALFUNCTION OF THE AQUABEAM ROBOTIC SYSTEM WAS REPORTED. ADDITIONAL INFORMATION REGARDING CURRENT PATIENT'S CONDITION IS BEING REQUESTED.
Patient problems: Urinary Incontinence
MAUDE report 15036928 ↗MalfunctionMar 3, 2022Display or Visual Feedback Problem; Display or Visual Feedback Problem
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT DURING THE AQUABLATION PROCEDURE, AND WHILE PRIMING THE AQUABEAM HANDPIECE, AN "E05 - CONSOLE ERROR" MESSAGE WAS GENERATED BY THE AQUABEAM ROBOTIC SYSTEM. AFTER MULTIPLE TROUBLESHOOTING STEPS WERE PERFORMED, THE PROCEDURE WAS ABLE TO BE CONTINUED THROUGH SUCCESSFUL COMPLETION. THE REPORTED EVENT CAUSED A PROCEDURAL DELAY OF OVER 20 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13660207 ↗MalfunctionDec 20, 2021Connection Problem; Connection Problem
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION BECAME AWARE THAT DURING THE AQUABLATION DIFFICULTY WAS ENCOUNTERED ENGAGING THE AQUABEAM HANDPIECE PUMP CARTRIDGE TO THE AQUABEAM CONSOLE. SEVERAL TROUBLESHOOTING STEPS WERE PERFORMED WITHOUT RESOLVING THE ISSUE. THE AQUABEAM HANDPIECE WAS REPLACED WITH A NEW HANDPIECE UNIT AND THE AQUABLATION PROCEDURE WAS ABLE TO BE SUCCESSFULLY COMPLETED. THE REPORTED EVENT CAUSED A PROCEDURAL DELAY OF OVER 20 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13049199 ↗
Other Fluid jet removal system devices
- AquaBeam Robotic System AB2000PROCEPT BioRobotics Corporation · Fluid jet removal system21 deaths827 reports
- HYDROS Robotic System HY1000PROCEPT BioRobotics Corporation · Fluid jet removal system5 deaths102 reports
- AquaBeam Robotic System AB2000PROCEPT BioRobotics Corporation · Fluid jet removal system5 deaths101 reports
- AquaBeam Robotic System AB2000CPROCEPT BioRobotics Corporation · Fluid jet removal system16 reports
- AquaBeam Handpiece HP2000PROCEPT BioRobotics Corporation · Fluid jet removal system12 reports
- HYDROS Handpiece HH1000PROCEPT BioRobotics Corporation · Fluid jet removal system12 reports
- AquaBeam System 210101PROCEPT BioRobotics Corporation · Fluid Jet Removal System7 reports
- AquaBeam Handpiece HP2000PROCEPT BioRobotics Corporation · Fluid jet removal system2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.