AquaBeam Handpiece HP2000 (UDI B614HP2000X1)Recalls & reported problems
PROCEPT BioRobotics Corporation · Class 2
All models of AquaBeam Handpiece (5)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI B614HP2000X1), received between Dec 22, 2023 and Oct 11, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Visual Prompts will not Clear” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Visual Prompts will not Clear2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 11, 2024Visual Prompts will not Clear; Visual Prompts will not Clear
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT THE AQUABEAM ROBOTIC SYSTEM GENERATED AN "E22 - MOTORPACK ERROR" WHEN CONNECTING THE AQUABEAM HANDPIECE TO THE AQUABEAM MOTORPACK. THE E22 ERROR WAS CLEARED. DURING NOZZLE REGISTRATION THE AQUABEAM ROBOTIC SYSTEM GENERATED ANOTHER "E22 - MOTORPACK ERROR". THE ERROR WAS AGAIN CLEARED. DURING THE FIRST TREATMENT PASS, THE AQUABEAM ROBOTIC SYSTEM GENERATED MULTIPLE "E22 - MOTORPACK ERROR", "E23 - MOTORPACK ERROR", AND "E31 - MOTORPACK ERROR". AFTER TROUBLESHOOTING, THE FIRST TREATMENT PASS WAS ABLE TO BE COMPLETED. DURING THE SECOND TREATMENT PASS, THE AQUABEAM ROBOTIC SYSTEM GENERATED MULTIPLE "E22 - MOTORPACK ERROR", "E23 - MOTORPACK ERROR", AND "E31 - MOTORPACK ERROR". AS A RESULT, THE TREATING SURGEON DECIDED TO USE A BIPOLAR LOOP TO FINISH THE TREATMENT. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20433942 ↗MalfunctionDec 22, 2023Visual Prompts will not Clear; Visual Prompts will not Clear
A MALE PATIENT UNDERWENT AN AQUABLATION PROCEDURE FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS CORPORATION (PROCEPT) BECAME AWARE THAT DURING JET ALIGNMENT, THE AQUABEAM ROBOTIC SYSTEM GENERATED AN "E22 - MOTORPACK ERROR" MESSAGE. A SECOND AQUABEAM HANDPIECE WAS USED WHICH RESOLVED THE ISSUE AND THE AQUABLATION PROCEDURE WAS SUCCESSFULLY COMPLETED. THE REPORTED EVENT CAUSED A SURGICAL DELAY OF OVER 20 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18384528 ↗
Other Fluid jet removal system devices
- AquaBeam Robotic System AB2000PROCEPT BioRobotics Corporation · Fluid jet removal system21 deaths827 reports
- HYDROS Robotic System HY1000PROCEPT BioRobotics Corporation · Fluid jet removal system5 deaths102 reports
- AquaBeam Robotic System AB2000PROCEPT BioRobotics Corporation · Fluid jet removal system5 deaths101 reports
- AquaBeam Robotic System AB2000CPROCEPT BioRobotics Corporation · Fluid jet removal system16 reports
- AquaBeam Handpiece HP2000PROCEPT BioRobotics Corporation · Fluid jet removal system12 reports
- HYDROS Handpiece HH1000PROCEPT BioRobotics Corporation · Fluid jet removal system12 reports
- AquaBeam System 210101PROCEPT BioRobotics Corporation · Fluid Jet Removal System7 reports
- AquaBeam Handpiece HP2000PROCEPT BioRobotics Corporation · Fluid jet removal system5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page