Schick 100008633Recalls & reported problems
Dentsply International Inc. · SCHICK33 S2 STARTER KIT/AE INTERFACE 6FT · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI D0031000086330), received between Feb 10, 2026 and Mar 30, 2026 — classified as 18 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Poor Quality Image18
Patient problems
- No Clinical Signs, Symptoms or Conditions18
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24730136 ↗MalfunctionMar 23, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT. THIS EVENT REQUIRES MEDICAL INTERVENTION TO RETAKE THE X-RAY IMAGE. THIS ALSO EXPOSES THE PATIENT TO ADDITIONAL RADIATION DUE TO THE NEED TO RETAKE THE X-RAY IMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660936 ↗MalfunctionMar 16, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24607407 ↗MalfunctionMar 16, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, THEY ALLEGE EXPERIENCING INTERMITTENT CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606968 ↗MalfunctionMar 16, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606935 ↗MalfunctionMar 16, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606816 ↗MalfunctionMar 16, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE THERE WAS A DEVICE RESPONSE ISSUE WHERE THE X-RAY IMAGE WAS NOT CAPTURED AFTER RADIATION HAS OCCURRED (NO DEVICE OUTPUT). NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606687 ↗MalfunctionFeb 24, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24431632 ↗MalfunctionFeb 24, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24431131 ↗MalfunctionFeb 24, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE A PREVIOUS PATIENT¿S X-RAY IMAGE WAS TRANSFERRED INSTEAD OF THE CURRENT PATIENT. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24429480 ↗MalfunctionFeb 24, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24429265 ↗MalfunctionFeb 17, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24367382 ↗MalfunctionFeb 12, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING CONNECTIVITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN OR AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24327106 ↗MalfunctionFeb 12, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE THE DEVICE SELF TRIGGERED, CAUSING AN X-RAY IMAGE TO BE TAKEN WITHOUT PROMPTING THE DEVICE TO TAKE AN IMAGE. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24327034 ↗MalfunctionFeb 12, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE THERE WAS AN X-RAY IMAGE TRANSMISSION ISSUE WHERE THE X-RAY IMAGE WAS NOT CAPTURED AFTER RADIATION HAS OCCURRED (TRANSMISSION FAILURE). NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24326266 ↗MalfunctionFeb 12, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM, SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24325950 ↗MalfunctionFeb 11, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A PART OF AN INTRAORAL SENSOR SYSTEM,SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6', THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24314463 ↗MalfunctionFeb 10, 2026Poor Quality Image
WHILE THE CUSTOMER WAS USING A SCHICK33 S2 STARTER KIT/3.0 INTERFACE 6' SENSOR SYSTEM, THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24309171 ↗
Other Solid state x-ray imager (flat panel/digital imager) devices
- SEDECAL SM-40HF-B-D-CSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)24 recalls1 report
- SEDECAL SM-40HF-B-D-VIRSedecal · Solid State X-Ray Imager (Flat Panel/Digital Imager)12 recalls1 report
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)7 recalls0 reports
- DigitalDiagnost C90 712034Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)3 recalls90 reports
- DigitalDiagnost C90 712035Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)4 recalls17 reports
- DigitalDiagnost 712031Philips Medical Systems DMC GmbH · Solid state x-ray imager (flat panel/digital imager)2 recalls65 reports
- NA GM85Samsung Electronics Co., Ltd. · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls4 reports
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page