Skip to content

Schick 100008640Recalls & reported problems

Dentsply International Inc. · SCHICK33 S0 STARTER KIT/AE INTERFACE 9FT · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI D0031000086400), received between Feb 10, 2026 and Feb 10, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Poor Quality Image” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Poor Quality Image1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 10, 2026Poor Quality Image

    WHILE THE CUSTOMER WAS USING A SCHICK 33 SENSOR INTRAORAL X-RAY UNIT SENSOR SYSTEM, THEY ALLEGE EXPERIENCING IMAGE QUALITY ISSUES WHEN AN X-RAY IMAGE WAS TAKEN AND AN ADDITIONAL IMAGE WAS REQUIRED. NO INJURY WAS REPORTED FROM THE ALLEGED EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24306377 ↗

Other Solid state x-ray imager (flat panel/digital imager) devices

All Solid state x-ray imager (flat panel/digital imager) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page