Skip to content

VSI Micro-Introducer Kit 7195VRecalls & reported problems

Vascular Solutions, Inc. · VSI Micro-Introducer Kits are intended for use in percutaneous introduction of up to a 0.018 inch or 0.038 inch guidewire or catheter into the vascular system following a small gauge needle stick. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M2067195V0), received between Nov 7, 2018 and Nov 29, 2018 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 29, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    COPY OF MDR AND LETTER RECEIVED OCTOBER 30, 2018 FROM HOSPITAL SITE. PATIENT WAS TO HAVE HEART CATH. ACCESS TO RIGHT FEMORAL ARTERY. 4F MICRO-PUNCTURE INTRODUCER INSERTED ON WIRE, SCALPEL WAS USED TO WIDEN INSERTION SITE, PROXIMAL HALF MICROPUNCTURE SHEATH REMOVED FROM WIRE NOTING DISTAL PORTION OF SHEATH REMAINED IN THE FEMORAL ARTERY. WE WERE UNABLE TO RETRIEVE DISTAL PORTION OF SHEATH. VASCULAR SURGERY CONSULTED AND TOOK PATIENT TO THE OR TO RETRIEVE SHEATH. PATIENT REMAINED STABLE THROUGHOUT THE INCIDENT AND PATIENT ADMITTED TO HOSPITAL.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8115188 ↗
  • MalfunctionNov 7, 2018Leak/Splash; Leak/Splash

    MICROPUNCTURE SHEATH WAS INSERTED INTO RIGHT ARM FISTULA. CONTRAST WAS INJECTED INTO MICROPUNCTURE SHEATH. CONTRAST SQUIRTED OUT OF HOLE IN SHEATH, WHERE THE TUBE MEETS THE HUB.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8046873 ↗

Other Introducer, Catheter devices

All Introducer, Catheter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page