VSI Micro-Introducer Kit 7272XRecalls & reported problems
Vascular Solutions, Inc. · VSI Micro-Introducer Kits are intended for use in percutaneous introduction of up to a 0.018 inch or 0.038 inch guidewire or catheter into the vascular system following a small gauge needle stick. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M2067272X0), received between Jul 3, 2018 and Jul 3, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2018Detachment Of Device Component; Detachment Of Device Component
THE BLUE INNER DILATOR BENT, AND TODAY THEY GOT THE WIRE STUCK AND THEN IT BROKE OFF INSIDE THE PATIENT. PHYSICIAN DID RETRIEVE THE WIRE. PER PHYSICIAN THE WIRE LOOPED IN THE GRAFT, BECAUSE OF THE LOOP/KINK THE WIRE WOULDN'T PULL OUT. HE PUSHED FORWARD AND THEN PULLED BACK HARD AND IT FRACTURED. HE SNARED WITH A SNARE DEVICE. IT FRACTURED IN THE STIFF PART OF THE WIRE. IT WAS STUCK IN A CLOTTED GRAFT AND SAID HE PULLED PRETTY HARD AND HAD NO PROBLEM IF HE HAD TO SNARE IT. NO HARM OR DELAY IN TREATMENT TO THE PATIENT. THE CASE WAS FINISHED. ADDITIONAL INFORMATION REC'D 05JUN18: HE WAS A DECLOT PATIENT, IN OPERATING ROOM. WE USED THE MICRO THEN USED THE BLUE DILATOR AND WIRE IN PRE-ACCESSED SITE TO PUT IN 7FR SHEATH. THEN GOT A 2ND ACCESS USING THE ACCESS KIT. NEEDLE WIRE, BLUE DILATOR AND WHITE DILATOR. THEN SWITCHED TO ANOTHER 7FR SHEATH. THE WIRE GOT STUCK, AND THE BLUE DILATOR BENT. THE TIP OF THE BLUE BROKE OFF, AND THE WIRE BROKE OFF, BOTH INSIDE OF THE PATIENT. PHYSICIAN THEN SUCCESSFULLY SNARED THE WIRE, WHICH HAD THE TIP OF THE BROKEN BLUE DILATOR ON IT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7659262 ↗
Other Introducer, Catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page