Skip to content

TRANSCEND® 26000027Recalls & reported problems

Microport Orthopedics Inc. · COCR TRANSCEND® FEMORAL HEAD 36MM SLT TAPER LONG NECK · Class 3 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M684260000271), received between May 20, 2019 and Oct 3, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Corroded” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Corroded2
  • Device Dislodged or Dislocated1
  • Insufficient Device Problem Information1

Patient problems

  • Metal Related Pathology3
  • Pain2
  • No Information1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 3, 2024Corroded; Corroded

    ALLEGEDLY, PATIENT UNDERWENT REVISION SURGERY ON RIGHT HIP DUE TO PAIN FROM UNDERLYING MECHANICAL ASSISTED CREVICE CORROSION FROM MODULAR, COBALT CHROME RIGHT HIP REPLACEMENT WITH AN ELEVATED COBALT LEVELS AND EVIDENCE OF PSEUDOTUMOR ON MRI AND METALLOSIS. THE PRESENCE OF A BLACK RESIDUE PREDOMINANTLY AT THE NECK BODY JUNCTION THAN THE HEAD NECK JUNCTION WAS NOTICED. NON MPO PRODUCTS: ZIMMER BIOMET RB 50 MM ACETABULAR COMPONENT, 50/36 POLYLINER, 25 MM ACETABULAR SCREW.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 20361185 ↗
  • InjuryMay 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, METALLOSIS AND ASEPTIC CUP LOOSENING.

    Patient problems: Metal Related Pathology

    MAUDE report 19243564 ↗
  • InjuryApr 26, 2022Corroded

    ALLEGEDLY, PATIENT UNDERWENT REVISION SURGERY ON RIGHT HIP DUE TO CORROSION AT THE NECK-STEM JUNCTION, PAIN AND ELEVATED COBALT AND CHROMIUM LEVELS. DR. (B)(6) REMOVED AND REPLACED THE WRIGHT PROFEMUR COCR MODULAR NECK, WRIGHT ACETABULAR HEAD, AND POLYETHYLENE LINER OF PLAINTIFF'S ARTIFICIAL HIP. PRODUCTS NOT REVISED: PARTID: PHA00242 PROFEMUR® Z FEMORAL STEM S 6 CEMENT LESS/ LOT NUMBER: 0511353146/ QTY:1.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 14212833 ↗
  • InjuryMay 20, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY THE PATIENT UNDERWENT RIGHT THR ON (B)(6) 2019, REVISED ON (B)(6) 2019 DUE TO CUP MAL-POSITIONING. THE PATIENT DISLOCATED THREE TIMES FOLLOWING PRIMARY SURGERY. DURING REVISION THE SAME CUP WAS REPOSITIONED, FLAT LINER REPLACED WITH HOODED LINER, LONGER HEAD WAS USED AND A SCREW ADDED. BOTH PROCEDURES WERE PERFORMED THROUGH POSTERIOR APPROACH.

    Patient problems: No Information

    MAUDE report 8623390 ↗

Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices

All Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.