Skip to content

SLR BIPOLAR 3110284200Recalls & reported problems

Microport Orthopedics Inc. · BIPOLAR SHELL 42mm OD 28mm ID · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M68431102842001), received between Nov 7, 2022 and Nov 7, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 7, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY, TROUBLE OCCURRENCE DURING BEDSIDE REHABILITATION, AN ABNORMALITY WAS NOTICED, AND CUP DISLOCATION WAS CONFIRMED BY RADIOGRAPHY. DURING THE HEALING PROCESS, THE NECK WAS ALSO DISLOCATED, AND THE PATIENT UNDERWENT REVISION. DOCTOR'S FINDING: THE PATIENT ORIGINALLY HAD A STANDING ACETABULUM AND PROBLEM WAS PROGRESSING, AND WHILE THE DISLOCATION WAS EXPECTED, THE FACT THAT THE NECK WAS DISLOCATED WAS NOT. THE PATIENT REQUESTED AN INVESTIGATION. ACCORDING TO THE X-RAYS, THE NECK AND THE STEM WERE DISLOCATED, THIS SHOULD NOT HAVE HAPPENED, SINCE THESE DEVICES MUST BE FIXED, SO IT IS PRESUMED THAT THESE COMPONENTS WERE DISSOCIATED FROM EACH OTHER. JAPAN (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 15743668 ↗

Other Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices

All Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page