GLADIATOR® BIPOLAR GLBP3650Recalls & reported problems
Microport Orthopedics Inc. · GLADIATOR® BIPOLAR 50MM OD X 36MM ID · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M684GLBP36501), received between Sep 7, 2022 and Sep 7, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “No Fail-Safe Mechanism” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- No Fail-Safe Mechanism1
Patient problems
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 7, 2022No Fail-Safe Mechanism; No Fail-Safe Mechanism
ALLEGEDLY, PATIENT HAD 36MM LONG LINEAGE HEAD DISSOCIATE FROM SIZE 50MM SHELL. UPON INSPECTION, NEITHER THE SHELL LOCK RING NOR THE HEAD SHOWED ANY EVIDENCE OF LOCK RING FAILURE. SPOKE WITH MORGAN BELL REGARDING.
Patient problems: Insufficient Information
MAUDE report 15374471 ↗
Other Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented devices
- STAGEONE 431209Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 recalls0 reports
- SELF CENTERING 1035-45-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented84 reports
- LFIT 6260-9-036Howmedica Osteonics Corp. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented80 reports
- ACTIS 101011060Depuy (Ireland) · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented66 reports
- RINGLOC BI-POLAR HIP SYSTEM 11-165228Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall1 death12 reports
- SELF CENTERING 1035-44-000Depuy Orthopaedics, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented60 reports
- TANDEM 71324046Smith & Nephew, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 recall9 reports
- TAPERLOC COMPLETE XR 123 MICRO FEMORAL 51-145170Biomet Orthopedics, LLC · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented59 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page