Skip to content

AxHA® AX143014Recalls & reported problems

INNO Holdings, Inc. · HA PEEK, Tantalum, Ti 6Al 4V ELI · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M711AX1430140), received between Mar 8, 2021 and Mar 8, 2021 — classified as 1 injury.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Abdominal Distention1
  • Pain1
  • Vertebral Fracture1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 8, 2021Adverse Event Without Identified Device or Use Problem

    ON (B)(6), 2021. INNOVASIS RECEIVED A PHONE CALL FROM DISTRIBUTOR (B)(4). INFORMING THAT DR. (B)(6) USED AN AX CAGE SEVERAL WEEKS AGO AND THE PATIENT HAS A VERTEBRAL BODY FRACTURE ON L5. IT WAS ALSO NOTED THAT NOTHING WENT WRONG WITH CAGE OR SYSTEM ITSELF. TIMELINE (B)(6) 2020 PATIENT SURGERY FOR BACK PAIN, L4 5-L4 5S1, LACK OF LORDOSIS. USED 2 CAGES WITH HIGH DEGREE OF LORDOSIS 14MM 14 CAGE AT L5S1, 16MM 20 AT L4-L5. PATIENT DEVELOPED OGILVIE SYNDROME BLOATING GI ISSUES POST OP ILEUS, REQUIRED MEDICAL TREATMENT 1 WEEK HOSPITALIZATION WHICH WENT WELL . CT SCAN FOR ILEUS SHOWS A LITTLE SETTLING OF THE IMPLANT. NOTE ILEUS NOT SPECIFIC TO IMPLANT DEVICE; RELATED TO ANY ABDOMINAL PROCEDURE. (B)(6) 2020, 2 WEEKS AFTER SURGERY SEVERE BACK PAIN, CT SCAN SHOWS A LOT MORE SETTLING SUBSIDENCE. (B)(6) 2020 REVISION SURGERY PERFORMED, POSTERIOR FIXATION. ONCE PEDICLE SCREWS WERE IMPLANTED THE SETTLING STOPPED. THE IMPLANT HAS MAXED OUT ON SUBSIDENCE, CAGE ALL THE WAY INTO THE VERTEBRAL BODY. REVISION HELPED PATIENT'S BACK PAIN AND SHE IS NOW DOING VERY WELL (B)(6) 2021.

    Patient problems: Abdominal Distention; Pain; Vertebral Fracture

    MAUDE report 11436174 ↗

Other Intervertebral Fusion Device With Integrated Fixation, Lumbar devices

All Intervertebral Fusion Device With Integrated Fixation, Lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page