2008 Series Hemodialysis Machines recallZ-0829-2018 · Fresenius Medical Care Renal Therapies Group, LLC · initiated Oct 5, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Medical Care Renal Therapies Group, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.

FDA root cause category: Software Design Change

What the company is doing

Fresenius Medical Care sent an Urgent Medical Device Correction letter dated April 24, 2017 to affected customers. the letter identified the affected product, problem and actions to be taken. For questions contact Technical Services at 1-800-227-2572.

Recall details

Recall number
Z-0829-2018
Status
Terminated
Initiated by the company
Oct 5, 2017
Posted by the FDA
Feb 20, 2018
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Quantity
15 machines
Product code
KDI
FDA submissions
K080964, K113427, K120017, K120505, K121341
Distribution
US Nationwide Distribution

Product description

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

Code information (models, lots, serial numbers)

2008T machines: 190766 2008T Hemodialysis System w/Bibag 190895 2008T GEN 2 Bibag without CDX  190766 = 00840861100910 (+Serial Number) 190895 = 00840861100927 (+Serial Number)ux=&Su