AFFIXUSHIP FRACTURE NAIL - LEFTRecalls & reported problems across all models
Biomet Orthopedics, LLC · 27 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls
- 24
We found 27 models of AFFIXUSHIP FRACTURE NAIL - LEFT by Biomet Orthopedics, LLC in the FDA's device registry (GUDID). Together they are named in 7 adverse event reports in the FDA's MAUDE database, received between Jul 6, 2022 and Oct 30, 2025 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Fracture” (6 reports).
24 FDA recall records name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
AFFIXUSHIP FRACTURE NAIL - LEFT recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0447-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0446-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0445-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0444-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0443-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0442-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0441-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0440-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0439-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0438-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0437-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0436-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0435-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0434-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0433-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0432-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0431-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0430-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0429-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0428-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0427-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0426-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0425-2018 · Root cause: Process change control · Recall details · FDA record ↗
- Aug 7, 2017Terminated
During the rework process for the April 17, 2017, field action, it was detected that the set screw and sleeve assembly which are designed to have an interference fit, were no longer in the installed location. This could prevent the product assembly with the mating screw and targeting jig.
Z-0424-2018 · Root cause: Process change control · Recall details · FDA record ↗
Most reported AFFIXUSHIP FRACTURE NAIL - LEFT problems
Device problems
- Fracture6
- Device Slipped1
Patient problems
- Failure of Implant6
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
- Swelling/ Edema1
AFFIXUSHIP FRACTURE NAIL - LEFT models
27 models, most reported first. Each model has its own page with its reports, recalls and device details.
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-11-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-09-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-11-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8144-11-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-280ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-320ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-340ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-400ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-420ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-260ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-280ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-300ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-420ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-440ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-11-460ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-13-460ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 recall0 reports
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-09-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-13-360ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- AFFIXUS®HIP FRACTURE NAIL - LEFT 8146-13-380ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
Recent AFFIXUSHIP FRACTURE NAIL - LEFT adverse event reports
The 7 most recent reports across all models, in the reporter's own words as published by the FDA.
InjuryOct 30, 2025Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT FEMUR FRACTURE SURGERY. SUBSEQUENTLY, THE PATIENT WAS REVISED ON APPROXIMATELY 1 MONTH POST-IMPLANTATION DUE TO THE AFFIXUS NAIL BREAKING AT THE LAG SCREW HOLE AND REQUIRED A CONVERSION TO ARCOS HEMI. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 23422693 ↗InjuryJan 14, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Nonunion/Delayed-union Bone Fracture; Pain; Swelling/ Edema
MAUDE report 21150397 ↗InjuryJun 17, 2024Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 19547283 ↗InjuryDec 7, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 18283791 ↗InjuryOct 3, 2023Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL HIP SURGERY ON AND UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT WAS REVISED APPROXIMATELY TWO (2) MONTHS AGO DUE TO THE IMPLANT FRACTURING. THERE WAS NO REPORT OF A FOREIGN BODY BEING RETAINED BY THE PATIENT.
Patient problems: Failure of Implant
MAUDE report 17861580 ↗InjuryAug 10, 2022Fracture; Fracture
ON AN UNKNOWN DATE A PATIENT HAD AN INITIAL SURGERY. ON ANOTHER UNKNOWN DATE THE PATIENT CAME IN FOR A REVISION DUE TO AN IMPLANTED NAIL FRACTURE.
Patient problems: Failure of Implant
MAUDE report 15199681 ↗MalfunctionJul 6, 2022Device Slipped; Device Slipped
IT WAS REPORTED THAT THE NAIL DID NOT PROPERLY ASSEMBLE TO THE JIG. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ATTEMPTS HAVE BEEN MADE. ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14936190 ↗
Other Biomet Orthopedics, LLC products
- AFFIXUSHIP FRACTURE NAIL - RIGHT39 models · 6 reports · 37 recalls
- Compress Device71 models · 99 reports · 22 recalls
- Arcos Modular Revision Hip System71 models · 191 reports · 19 recalls
- Comprehensive Reverse Shoulder57 models · 895 reports · 3 recalls
- Vanguard Knee System147 models · 306 reports · 13 recalls
- STAGEONE17 models · 0 reports · 18 recalls
- StageOne Select15 models · 2 reports · 15 recalls
- Comprehensive Shoulder System68 models · 191 reports · 11 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page