Infant Flow SiPAPRecalls & reported problems across all models
Carefusion Corporation · 7 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 109
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls
- 0
We found 7 models of Infant Flow SiPAP by Carefusion Corporation in the FDA's device registry (GUDID). Together they are named in 109 adverse event reports in the FDA's MAUDE database, received between Dec 28, 2018 and Apr 8, 2024 — classified as 20 injuries, 89 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (57 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported Infant Flow SiPAP problems
Device problems
- Mechanical Problem57
- Pressure Problem12
- Leak/Splash10
- Display or Visual Feedback Problem8
- Erratic or Intermittent Display3
- Gas/Air Leak3
- Infusion or Flow Problem3
- No Display/Image3
Patient problems
- No Clinical Signs, Symptoms or Conditions70
- No Consequences Or Impact To Patient13
- No Information11
- No Patient Involvement11
- Low Oxygen Saturation2
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
- Pressure Sore/Ulcer1
Infant Flow SiPAP models
7 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Infant Flow SiPAP 27416-001Ventilator, Continuous, Facility Use53 reports
- Infant Flow SiPAP 675-CFG-005Ventilator, Continuous, Facility Use29 reports
- Infant Flow SiPAP 677-002Ventilator, Continuous, Facility Use21 reports
- Infant Flow SiPAP 7772020LPVentilator, Continuous, Facility Use3 reports
- Infant Flow SiPAP 27415-001Ventilator, Continuous, Facility Use1 report
- Infant Flow SiPAP 777040XSVentilator, Continuous, Facility Use1 report
- Infant Flow SiPAP 777000SVentilator, Continuous, Facility Use1 report
Recent Infant Flow SiPAP adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2024No Device Output; No Device Output
IT WAS REPORTED TO VYAIRE MEDICAL THE SIPAP IS IN BIPHASIC MODE, THE BREATHS WILL NOT DELIVER NO MATTER WHAT RATE IS SET. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19059454 ↗MalfunctionApr 4, 2024No Pressure; No Pressure
IT WAS REPORTED TO VYAIRE MEDICAL THE VENT IS NOT PRESSURIZING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19047222 ↗MalfunctionApr 4, 2024No Display/Image; No Display/Image
IT WAS REPORTED TO VYAIRE MEDICAL PROBLEM WITH THE SIPAP VENTILATOR IN WHICH THE END USERS ARE REPORTING THAT WHEN THEY POWERED ON THE DEVICE, THE SCREEN 'WENT BLANK.'. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19041507 ↗MalfunctionMar 21, 2024Device Alarm System; Device Alarm System
IT WAS REPORTED TO VYAIRE MEDICAL THAT PRE-USE CHECK WAS BEING RAN, THEY WERE ABLE TO DO THE O2 CALIBRATION AND IT PASSED. AFTER THAT THE UNIT WAS "ALARMING" BUT NOT SURE WHY OR IF THERE WAS AN ERROR CODE ETC. FOUND THAT WE HAVE NO HISTORY OF A PM. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18949223 ↗MalfunctionMar 4, 2024No Pressure; No Pressure
IT WAS REPORTED TO VYAIRE MEDICAL THE SIPAP WAS NOT PRESSURIZING AND THAT ALL OF THE GAS WAS COMING FROM THE EXHAUST. FURTHERMORE, THERE WAS NO PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18826493 ↗MalfunctionJan 25, 2024Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED TO VYAIRE MEDICAL THAT LEAK IN UNIT FROM EXHALATION PORT OCCURRED ON THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18584642 ↗MalfunctionNov 9, 2023Gas/Air Leak; Improper Flow or Infusion; Gas/Air Leak; Improper Flow or Infusion
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR HAS LOUD GAS FLOW CONSTANTLY FROM THE EXHAUST. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18102567 ↗InjuryNov 6, 2023Mechanical Problem; No Pressure; Mechanical Problem; No Pressure
IT WAS REPORTED TO VYAIRE MEDICAL THE INFANT FLOW SIPAP HAS NOT HOLDING PRESSURE DURING PATIENT USE. FURTHERMORE, THE PATIENT WAS SWAPPED IN A DIFFERENT VENTILATOR AND THERE WAS NO PATIENT HARM ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18074685 ↗MalfunctionOct 20, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR IS NOT WORKING CORRECTLY AND MAKES NOISE. FURTHERMORE, THERE WAS NO INFORMATION REGARDING PATIENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17976520 ↗MalfunctionOct 3, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT SIPAP VENTILATOR HAD AN EXCESSIVE LEAK FROM THE EXHAUST. THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17865033 ↗MalfunctionAug 23, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR SCREEN IS STILL FLICKERING. THE CUSTOMER CONFIRMED THAT THERE IS NO PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17608769 ↗MalfunctionAug 3, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR HAS EXCESS LEAK WHEN HIGH PRESSURE GAS SOURCES ARE CONNECTED. AS OF THIS TIME THERE IS NO INFORMATION ABOUT PATIENT INVOLVEMENT ASSOCIATED WITH THIS REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17455428 ↗MalfunctionJul 26, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE SIPAP VENTILATOR HAS A LARGE BLEED COMING FROM THE BACK WHEN MEDICAL GASES ARE CONNECTED. AT THIS TIME, THERE WAS NO INFORMATION OF PATIENT INVOLVEMENT REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17401692 ↗MalfunctionJul 21, 2023Mechanical Problem; Mechanical Problem
IT WAS REPORTED TO VYAIRE MEDICAL THAT THE BLENDER OF THE INFANT SIPAP VENTILATOR BLEEDS EXCESSIVELY. THERE IS NO INFORMATION OF PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT AS OF THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17374906 ↗MalfunctionJul 17, 2023Leak/Splash; Leak/Splash
IT WAS REPORTED TO VYAIRE MEDICAL THAT LEAK INTERNALLY OCCURRED ON THE SIPAP VENTILATOR. THE CUSTOMER REPORTED THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17331032 ↗
Other Carefusion Corporation products
All Carefusion Corporation devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page