Skip to content

Ambu aView 2 AdvanceRecalls & reported problems across all models

Ambu A/S · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Ambu aView 2 Advance model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls
2

We found 2 models of Ambu aView 2 Advance by Ambu A/S in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Apr 4, 2024 and Jun 12, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Battery Problem” (2 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Ambu aView 2 Advance recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Dec 5, 2023Open, Classified

    Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

    Z-0714-2024 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  2. Jul 21, 2023Open, Classified

    Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

    Z-2628-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported Ambu aView 2 Advance problems

Device problems

  • Battery Problem2
  • Failure to Power Up2
  • Insufficient Device Problem Information2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Ambu aView 2 Advance models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Ambu aView 2 Advance adverse event reports

The 4 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Ambu A/S products

All Ambu A/S devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page