Ambu aView 2 AdvanceRecalls & reported problems across all models
Ambu A/S · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Ambu aView 2 Advance model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 2
We found 2 models of Ambu aView 2 Advance by Ambu A/S in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Apr 4, 2024 and Jun 12, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (2 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Ambu aView 2 Advance recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 5, 2023Open, Classified
Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.
Z-0714-2024 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
- Jul 21, 2023Open, Classified
Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames
Z-2628-2023 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Ambu aView 2 Advance problems
Device problems
- Battery Problem2
- Failure to Power Up2
- Insufficient Device Problem Information2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Ambu aView 2 Advance models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Ambu aView™ 2 Advance 405011000ACCESSORIES, PHOTOGRAPHIC, FOR ENDOSCOPE (EXCLUDE LIGHT SOURCES)2 recalls4 reports
- Ambu aView™ 2 Advance 405011000US2Bronchoscope (flexible or rigid)1 recall0 reports
Recent Ambu aView 2 Advance adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2024Battery Problem; Insufficient Device Problem Information; Battery Problem; Insufficient Device Problem Information
ACCORDING TO THE INFORMATION FROM THE CUSTOMER, THE HOUSING OF THE MONITOR HAS BULGED. THEY REPORTED THAT THE BATTERY HAS INFLATED AND BURST THE MONITOR. PICTURES WERE DELIVERED FOR THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19522066 ↗MalfunctionJun 12, 2024Battery Problem; Insufficient Device Problem Information; Battery Problem; Insufficient Device Problem Information
ACCORDING TO THE INFORMATION FROM THE CUSTOMER, THE HOUSING OF THE MONITOR HAS BULGED. THEY REPORTED THAT THE BATTERY HAS INFLATED AND BURST THE MONITOR. PICTURES WERE DELIVERED FOR THE INVESTIGATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19518071 ↗MalfunctionApr 4, 2024Failure to Power Up; Failure to Power Up
THE USER REPORTED THAT THE MONITOR WAS NOT RESPONDING AND HAVING ISSUE BOOTING UP WHILST TRYING TO INTUBATE A PATIENT. ANOTHER MONITOR WAS USED AND THE PATIENT WAS INTUBATED SUCCESSFULLY. THE USER HAS REPORTED THAT THE MONITOR HAS WORKED AGAIN AFTER A HARD RESET HOWEVER THEY HAVE NOTICED BOOTING AND SHUTTING DOWN ISSUES WITH THIS MONITOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19048038 ↗MalfunctionApr 4, 2024Failure to Power Up; Failure to Power Up
THE USER REPORTED THAT THE MONITOR WAS NOT RESPONDING AND HAVING ISSUE BOOTING UP WHILST TRYING TO INTUBATE A PATIENT. ANOTHER MONITOR WAS USED AND THE PATIENT WAS INTUBATED SUCCESSFULLY. THE USER HAS REPORTED THAT THE MONITOR HAS WORKED AGAIN AFTER A HARD RESET HOWEVER THEY HAVE NOTICED BOOTING AND SHUTTING DOWN ISSUES WITH THIS MONITOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19039919 ↗
Other Ambu A/S products
All Ambu A/S devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.