Skip to content

Ambu aView 2 Advance recallZ-0714-2024 · Ambu Inc. · initiated Dec 5, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ambu Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Instructions for Use have been updated to add the following warning: Use only M4 screws with the length of 14 16 mm when mounting Ambu aView 2 Advance on a VESA interface. Using longer screw lengths will penetrate the lithium-ion battery and result in a fire hazard and battery leakage which can cause severe burns, smoke inhalation and skin irritation. Using shorter screw lengths could result in unsecure device fastening.

FDA root cause category: Device Design

What the company is doing

On December 5, 2023, the firm began notifying their customers via Urgent Field Safety Notice (Correction) letters. Customers were provided with an insert of Instructions for Use to include with their Instructions for Use booklet that highlights the new warning. Should you have additional questions, please contact Sanjay Parikh ([email protected]) or Tammy Feyerherm ([email protected]).

Recall details

Recall number
Z-0714-2024
Status
Open, Classified
Initiated by the company
Dec 5, 2023
Posted by the FDA
Jan 11, 2024
Recalling firm
Ambu Inc.
Quantity
11845 distributed US
Product code
EOQ
FDA submissions
K223299
Distribution
Domestic distribution nationwide.

Product description

Ambu aView 2 A Monitor, Catalog numbers 405011000 and 405011000US2

Code information (models, lots, serial numbers)

Catalog No. 405011000 - UDI-DI 5707480145089 Catalog No. 405011000US2 - UDI-DI 5707480156757 All products distributed since launch (May 2020)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates